Slide 1: From Monitoring to Controlling with Vaporized H2O2 Sensors
From Monitoring to Controlling with Vaporized H2O2 Sensors – Why, How & Case Study Webinar 18 June 2019 Vaisala & Cleamix © Vaisala

Slide 2: Meet the Presenters
Meet the Presenters Sanna Lehtinen Product Manager at Vaisala with 20 years of experience in life science applications and wide product management experience from leading international high tech companies. Piritta Maunu Life Science Regulatory and Industry Expert at Vaisala with over 15 years of exp

Slide 3: Meet the Presenters
Meet the Presenters Cleamix: Hanna-Kaisa Kyyrönen Hanna-Kaisa Kyyrönen has over 10 years of experience in decontamination and sterilization processes having worked in different sales and product management positions in health technology. Cleamix: Panu Wilska With over 25 years of international work

Slide 4: Agenda
Agenda Basics of Bio-Decontamination with Vaporized H2O2 Vaisala’s patented PEROXCAP® Technology Inline Measurements During Bio-Decontamination - From Monitoring to Controlling Cleamix Technology Highlights Variables That Influence Hydrogen Peroxide Vapor Concentration How Measurements C

Slide 5: Basics of Bio-Decontamination with Vaporized H2O2
Basics of Bio- Decontamination with Vaporized H2O2 © Vaisala

Slide 6: Why H2O2 for Bio-Decontamination
Why H2O2 for Bio-Decontamination Easy to use Destroys all biological contaminants Works in low temperature processes Processes can be validated Compatible with a wide variety of materials Environmentally friendly process Leaves no real residues – only water vapor and oxygen From Monito

Slide 7: Example Bio-Decontamination Cycle
Example Bio-Decontamination Cycle Typical non-condensing bio-decontamination process in isolators From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 7

Slide 8: Vaisala's Patented PEROXCAP Technology
Vaisala’s Patented PEROXCAP® Technology © Vaisala

Slide 9: HPP270 series
HPP270 series Patented PEROXCAP® measurement technology Based on capacitive thin-film polymer sensor (HUMICAP®) Measurements: Hydrogen peroxide vapor, up to 2000 ppm, not for safety (<1ppm) Humidity, 0-100% – Relative Saturation (%RS) – Relative Humidity (%RH) Temperature, 5-50ºC Digital and

Slide 10: The Difference Between Relative Humidity (%RH) and Relative Saturation (%RS)
The Difference Between Relative Humidity (%RH) and Relative Saturation (%RS) © Vaisala

Slide 11: Relative Saturation vs. Relative Humidity
Relative Saturation vs. Relative Humidity Only H2O vapor present H2O and H2O2 vapor present From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 11

Slide 12: Relative Saturation %
RELATIVE SATURATION % Condensation 100% 0% RS% value is the only parameter for controlling condensation when H2O2 vapor is present. © Vaisala 12

Slide 13: Customer Case / FDA Audit
Customer Case / FDA Audit From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 13

Slide 14: Relative Humidity vs. Relative Saturation
Relative Humidity vs. Relative Saturation CONDENSATION 100% %RH %RS 100% H2O2 H2O H2O 0% 0% From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 14

Slide 15: Solution For All Bio-Decontamination Processes; Dry, Wet or Micro-condensing
Solution For All Bio-Decontamination Processes; Dry, Wet or Micro-condensing Relative saturation (RS) value is needed to be able to control a condensation phenomenon. Process Humidity Condensing environment type values Dry < 100 % Avoided Micro- ~ 100 % Being at a sub-micron level is condensing in

Slide 16: Inline Measurements During Bio-Decontamination
Inline Measurements During Bio- Decontamination - From Monitoring to Controlling © Vaisala

Slide 17: Indicators vs. Inline Measurements
Indicators vs. Inline Measurements Chemical Biological indicators Enzyme indicators Measurement sensor; indicators (CI) (BI) (EI) Vaisala HPP272 What the product looks like? PROS Inexpensive Quantitative results Quantitative results Continuous, stable and Easy to use Instantaneous reaction repeatabl

Slide 18: Why Inline Measurements?
Why Inline Measurements? Provides continuous measurement data Guarantees that a process works as planned May decrease a number of biological, chemical or enzymatic indicators Monitoring processes also between validations enables you to see changes in the environment (°C, %RH) Enables good

Slide 19: Why from Monitoring to Controlling?
Why from Monitoring to Controlling? Without Controlling Systems; Target = 400 ppm With Controlling Systems; Target = 400 ppm 600 600 H2O2 Concentration (ppm) H2O2 Concentration (ppm) 500 500 400 400 300 300 200 200 100 100 0 0 Time Time From H2O2 liquid injection rate to real in-line controlling w

Slide 20: Measurement Used for Process Controlling Should
Measurement Used for Process Controlling Should Be repeatable Be stable Withstand condensing environments Be accurate Not absorb, desorb nor catalyze H2O2 vapor Have digital signal Have long calibration intervals Have fast response time From Monitoring to Controlling with Vaporized H

Slide 21: PEROXCAP Technology Supports Controlling Activities
PEROXCAP Technology Supports Controlling Activities © Vaisala

Slide 22: PEROXCAP H2O2 Measurement Repeatability
PEROXCAP® H2O2, Measurement Repeatability HPP272 at 500 ppm & 40 %RS 15 Difference to average [H2O2] error 10 5 0 (ppm) -5 -10 -15 0 1 2 3 4 5 6 7 8 9 10 Cycle number P2230776 P2230779 P2230780 P2230782 P2230784 P2430885 P2430886 P2430887 P2430888 P2430889 From Monitoring to Controlling with Vaporiz

Slide 23: PEROXCAP %RS Measurement Repeatability
PEROXCAP® %RS, Measurement Repeatability HPP272 at 500 ppm & 40 %RS 0.6 Difference to average [RS] error 0.4 0.2 0 (%RS) -0.2 -0.4 -0.6 0 1 2 3 4 5 6 7 8 9 10 Cycle number P2230776 P2230779 P2230780 P2230782 P2230784 P2430885 P2430886 P2430887 P2430888 P2430889 From Monitoring to Controlling with Va

Slide 24: PEROXCAP HPP272 Long Term Stability, H2O2 ppm
PEROXCAP®, HPP272, Long Term Stability, H2O2 ppm H2O2 measurement drift after different exposure times 130 110 218h at 90 460 ppm 70 = e.g. 436 d [H2O2] (ppm) 50 times at 30 460 ppm 10 for 30 minutes ‐10 0 50 100 150 200 250 300 350 400 450 500 ‐30 → drift only ~ +10 ppm ‐50 ref [H2O2] (ppm) 4h 16 h

Slide 25: Reliable Measurements Also Under Condensing Environments
Reliable Measurements Also Under Condensing Environments HPP272 stability after 200 times at H2O2 cycles (400-500 ppm, RS~100%RS) 0.5 0.45 0.4 Measurement error [%RH / %RS] 0.35 0.3 0.25 0.2 0.15 0.1 0.05 0 0 10 20 30 40 50 60 70 80 90 100 Relative humidity [%RH] RH RS From Monitoring to Controlling

Slide 26: Accurate H2O2 and Humidity Measurements
Accurate H2O2 and Humidity Measurements Traceable factory calibration Traceable to the international system of units (SI) through national metrology institutes Quality of probe materials Limited absorption, desorption and catalyzing of H2O2 Materials selected according to the rigorous and

Slide 27: Cleamix Technology Highlights
Cleamix Technology Highlights © Vaisala

Slide 28: Cleamix VCS Series: A New Method for Generating Vaporized H2O2
Cleamix VCS Series: A New Method for Generating Vaporized H2O2 A powerful, portable bio-decontamination unit having the size of a briefcase. Performance experiments conducted in 3 different enclosures: Size of enclosure/ ppm/H2O2 vapor time/min m3 10 > 800 15 36 > 600 30 68 > 500 50 One litre

Slide 29: Variables That Influence Hydrogen Peroxide Vapor Concentration
Variables That Influence Hydrogen Peroxide Vapor Concentration © Vaisala

Slide 30: Measurable Variables That Impact H2O2 Vapor Concentration
Measurable Variables That Impact H2O2 Vapor Concentration Humidity Temperature From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 30

Slide 31: Other Enclosure Variables That Impact H2O2 Vapor Concentration
Other Enclosure Variables That Impact H2O2 Vapor Concentration Total surface area of the space or room to be decontaminated Properties of the surface materials: Porosity Tendency to break down hydrogen peroxide molecules Temperature variations of the surfaces The cooler the surface than

Slide 32: How Measurements Can Help You Provide Efficient H2O2 Vapor Output?
How Measurements Can Help You to Provide Efficient H2O2 Vapor Output? © Vaisala

Slide 33: Sensor to PLC Connection for Control
Sensor to PLC Connection for Control Sensor must be rugged enough to maintain accuracy through multiple processes. Testing determined that available sensing technologies were unsatisfactory. EXCEPT with Vaisala’s HPP272 probe… From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, H

Slide 34: Why We Selected Vaisala's Technology?
Why We Selected Vaisala's Technology? The probe provides all the parameters you need: Hydrogen peroxide vapour Temperature Humidity as both relative saturation and relative humidity Options for dew point and vapor pressure The HPP270 probe with digital output signal Analog signals are

Slide 35: Things to Consider When Planning Bio-Decontamination Processes
Things to Consider When Planning Bio- Decontamination Processes Evaluate applicable process parameters H2O2 vapor concentration and exposure time, required to eliminate contamination from surfaces Consider environmental factors Relative saturation limit value is the dominant variable Hydro

Slide 36: Controlling Condensation
Controlling Condensation Real time monitoring of relative saturation is needed to maintain dry conditions with minimum condensation Challenges: Optimizing the hydrogen peroxide vapour generation rate to fit small decontamination volumes – Difficulties in controlling the hysteresis Condensati

Slide 37: Validation of Decontamination Cycles
Validation of Decontamination Cycles In order to validate a process, accurate, real time information on cycle parameters is necessary When a desired H2O2 ppm level is easily attained, the total treatment time can be relatively short and the time needed for aeration is propably short as well Wh

Slide 38: Case Study: Hospital Room Bio-Decontamination
Case Study: Hospital Room Bio- Decontamination © Vaisala

Slide 39: Challenges in Infection Prevention
Challenges in Infection Prevention Suboptimal cleaning practises Important pathogens not eliminated from surfaces: Clostridium Difficile (C.diff) Methicillin-resistant Staphylococcus aureus (MRSA) Vancomycin-resistant Enterococcus (VRE) Multidrug-resistant Acinetobacter Risk of infecti

Slide 40: No Touch Decontamination Solution
No Touch Decontamination Solution Automated technology – exceeds human performance More intensive decontamination Reliable process monitoring and control From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 40

Slide 41: Hospital Room Decontamination
Hospital Room Decontamination REMOVE OPEN REMOVE From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 41

Slide 42: Hospital Room Decontamination
Hospital Room Decontamination Cleamix H2O2 vapor generator HPP272 PROBE From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 42

Slide 43: Hospital Room Decontamination
Hospital Room Decontamination Cleamix H2O2 vapor generator HPP272 PROBE Oscillating fans From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 43

Slide 44: Hospital Room Decontamination
Hospital Room Decontamination BIs and CIs Cleamix H2O2 vapor generator HPP272 PROBE Oscillating fans From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 44

Slide 45: Hospital Room Decontamination
Hospital Room Decontamination BIs and CIs Cleamix H2O2 vapor generator HPP272 PROBE Oscillating fans Catalytic converter From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 45

Slide 46: Monitoring of the Decontamination Cycles
Monitoring of the Decontamination Cycles Control: Via cable or WiFi – Touch screen and computer – Tablet – Phone Accurate Real time Documented From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 46

Slide 47: Safety Concerns
Safety Concerns Post warning signs around perimeter to restrict access to H2O2 exposed area In the event you must enter the process area, wear appropriate protective clothing and mask Use handheld meter to detect any leaks of H2O2 Wait until proper aeration to enter area From Monitoring to C

Slide 48: Conclusion
Conclusion By moving from monitoring to controlling you can Control the process based on real results Reduce batch variability Reduce validation efforts Repeatable, stable and accurate PEROXCAP technology provides easy way to Keep the H2O2 ppm value at a desired level Control the proce

Slide 49: Would you like a Sales Engineer to Contact You?
Would you like a Sales Engineer to Contact You? Yes, Vaisala Sales Engineer (HPP271, HPP272, PEROXCAP®) Yes, Cleamix Sales Engineer (VCS-100 vapor generators) From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 49

Slide 50: Q&A Session
Q&A Session From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 50

Slide 51: Thank You for Attending!
Thank You for Attending! Vaisala: HPP270 Series data sheet Learn more about Bio-decontamination – Blogs, Videos, Application Notes More information about HPP270 series products Cleamix: Video of VCS100 From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-
Source: Internet Archive copy of vaisala.com
More decks from Vaisala Oyj Class A
Q3 2026 Earnings Presentation17 slides · Oct 27, 2026
Q2 2026 Earnings Presentation17 slides · Jul 21, 2026
Investor Presentation (July 2026)32 slides · July 2026
Q1 2026 Earnings Presentation17 slides · Apr 24, 2026
Q4 2025 Earnings Presentation14 slides · Feb 12, 2026
Q3 2025 Earnings Presentation14 slides · Oct 23, 2025