  1. Slide 1: From Monitoring to Controlling with Vaporized H2O2 Sensors

    Slide 1: From Monitoring to Controlling with Vaporized H2O2 Sensors

    From Monitoring to Controlling with Vaporized H2O2 Sensors – Why, How & Case Study Webinar 18 June 2019 Vaisala & Cleamix © Vaisala

  2. Slide 2: Meet the Presenters

    Slide 2: Meet the Presenters

    Meet the Presenters Sanna Lehtinen Product Manager at Vaisala with 20 years of experience in life science applications and wide product management experience from leading international high tech companies. Piritta Maunu Life Science Regulatory and Industry Expert at Vaisala with over 15 years of exp

  3. Slide 3: Meet the Presenters

    Slide 3: Meet the Presenters

    Meet the Presenters Cleamix: Hanna-Kaisa Kyyrönen Hanna-Kaisa Kyyrönen has over 10 years of experience in decontamination and sterilization processes having worked in different sales and product management positions in health technology. Cleamix: Panu Wilska With over 25 years of international work

  4. Slide 4: Agenda

    Slide 4: Agenda

    Agenda  Basics of Bio-Decontamination with Vaporized H2O2  Vaisala’s patented PEROXCAP® Technology  Inline Measurements During Bio-Decontamination - From Monitoring to Controlling  Cleamix Technology Highlights  Variables That Influence Hydrogen Peroxide Vapor Concentration  How Measurements C

  5. Slide 5: Basics of Bio-Decontamination with Vaporized H2O2

    Slide 5: Basics of Bio-Decontamination with Vaporized H2O2

    Basics of Bio- Decontamination with Vaporized H2O2 © Vaisala

  6. Slide 6: Why H2O2 for Bio-Decontamination

    Slide 6: Why H2O2 for Bio-Decontamination

    Why H2O2 for Bio-Decontamination  Easy to use  Destroys all biological contaminants  Works in low temperature processes  Processes can be validated  Compatible with a wide variety of materials  Environmentally friendly process  Leaves no real residues – only water vapor and oxygen From Monito

  7. Slide 7: Example Bio-Decontamination Cycle

    Slide 7: Example Bio-Decontamination Cycle

    Example Bio-Decontamination Cycle Typical non-condensing bio-decontamination process in isolators From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 7

  8. Slide 8: Vaisala's Patented PEROXCAP Technology

    Slide 8: Vaisala's Patented PEROXCAP Technology

    Vaisala’s Patented PEROXCAP® Technology © Vaisala

  9. Slide 9: HPP270 series

    Slide 9: HPP270 series

    HPP270 series Patented PEROXCAP® measurement technology Based on capacitive thin-film polymer sensor (HUMICAP®) Measurements:  Hydrogen peroxide vapor, up to 2000 ppm, not for safety (<1ppm)  Humidity, 0-100% – Relative Saturation (%RS) – Relative Humidity (%RH)  Temperature, 5-50ºC  Digital and

  10. Slide 10: The Difference Between Relative Humidity (%RH) and Relative Saturation (%RS)

    Slide 10: The Difference Between Relative Humidity (%RH) and Relative Saturation (%RS)

    The Difference Between Relative Humidity (%RH) and Relative Saturation (%RS) © Vaisala

  11. Slide 11: Relative Saturation vs. Relative Humidity

    Slide 11: Relative Saturation vs. Relative Humidity

    Relative Saturation vs. Relative Humidity Only H2O vapor present H2O and H2O2 vapor present From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 11

  12. Slide 12: Relative Saturation %

    Slide 12: Relative Saturation %

    RELATIVE SATURATION % Condensation 100% 0% RS% value is the only parameter for controlling condensation when H2O2 vapor is present. © Vaisala 12

  13. Slide 13: Customer Case / FDA Audit

    Slide 13: Customer Case / FDA Audit

    Customer Case / FDA Audit From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 13

  14. Slide 14: Relative Humidity vs. Relative Saturation

    Slide 14: Relative Humidity vs. Relative Saturation

    Relative Humidity vs. Relative Saturation CONDENSATION 100% %RH %RS 100% H2O2 H2O H2O 0% 0% From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 14

  15. Slide 15: Solution For All Bio-Decontamination Processes; Dry, Wet or Micro-condensing

    Slide 15: Solution For All Bio-Decontamination Processes; Dry, Wet or Micro-condensing

    Solution For All Bio-Decontamination Processes; Dry, Wet or Micro-condensing  Relative saturation (RS) value is needed to be able to control a condensation phenomenon. Process Humidity Condensing environment type values Dry < 100 % Avoided Micro- ~ 100 % Being at a sub-micron level is condensing in

  16. Slide 16: Inline Measurements During Bio-Decontamination

    Slide 16: Inline Measurements During Bio-Decontamination

    Inline Measurements During Bio- Decontamination - From Monitoring to Controlling © Vaisala

  17. Slide 17: Indicators vs. Inline Measurements

    Slide 17: Indicators vs. Inline Measurements

    Indicators vs. Inline Measurements Chemical Biological indicators Enzyme indicators Measurement sensor; indicators (CI) (BI) (EI) Vaisala HPP272 What the product looks like? PROS Inexpensive Quantitative results Quantitative results Continuous, stable and Easy to use Instantaneous reaction repeatabl

  18. Slide 18: Why Inline Measurements?

    Slide 18: Why Inline Measurements?

    Why Inline Measurements?  Provides continuous measurement data  Guarantees that a process works as planned  May decrease a number of biological, chemical or enzymatic indicators  Monitoring processes also between validations enables you to see changes in the environment (°C, %RH)  Enables good

  19. Slide 19: Why from Monitoring to Controlling?

    Slide 19: Why from Monitoring to Controlling?

    Why from Monitoring to Controlling? Without Controlling Systems; Target = 400 ppm With Controlling Systems; Target = 400 ppm 600 600 H2O2 Concentration (ppm) H2O2 Concentration (ppm) 500 500 400 400 300 300 200 200 100 100 0 0 Time Time  From H2O2 liquid injection rate to real in-line controlling w

  20. Slide 20: Measurement Used for Process Controlling Should

    Slide 20: Measurement Used for Process Controlling Should

    Measurement Used for Process Controlling Should  Be repeatable  Be stable  Withstand condensing environments  Be accurate  Not absorb, desorb nor catalyze H2O2 vapor  Have digital signal  Have long calibration intervals  Have fast response time From Monitoring to Controlling with Vaporized H

  21. Slide 21: PEROXCAP Technology Supports Controlling Activities

    Slide 21: PEROXCAP Technology Supports Controlling Activities

    PEROXCAP Technology Supports Controlling Activities © Vaisala

  22. Slide 22: PEROXCAP H2O2 Measurement Repeatability

    Slide 22: PEROXCAP H2O2 Measurement Repeatability

    PEROXCAP® H2O2, Measurement Repeatability HPP272 at 500 ppm & 40 %RS 15 Difference to average [H2O2] error 10 5 0 (ppm) -5 -10 -15 0 1 2 3 4 5 6 7 8 9 10 Cycle number P2230776 P2230779 P2230780 P2230782 P2230784 P2430885 P2430886 P2430887 P2430888 P2430889 From Monitoring to Controlling with Vaporiz

  23. Slide 23: PEROXCAP %RS Measurement Repeatability

    Slide 23: PEROXCAP %RS Measurement Repeatability

    PEROXCAP® %RS, Measurement Repeatability HPP272 at 500 ppm & 40 %RS 0.6 Difference to average [RS] error 0.4 0.2 0 (%RS) -0.2 -0.4 -0.6 0 1 2 3 4 5 6 7 8 9 10 Cycle number P2230776 P2230779 P2230780 P2230782 P2230784 P2430885 P2430886 P2430887 P2430888 P2430889 From Monitoring to Controlling with Va

  24. Slide 24: PEROXCAP HPP272 Long Term Stability, H2O2 ppm

    Slide 24: PEROXCAP HPP272 Long Term Stability, H2O2 ppm

    PEROXCAP®, HPP272, Long Term Stability, H2O2 ppm H2O2 measurement drift after different exposure times 130 110 218h at 90 460 ppm 70 = e.g. 436 d [H2O2] (ppm) 50 times at 30 460 ppm 10 for 30 minutes ‐10 0 50 100 150 200 250 300 350 400 450 500 ‐30 → drift only ~ +10 ppm ‐50 ref [H2O2] (ppm) 4h 16 h

  25. Slide 25: Reliable Measurements Also Under Condensing Environments

    Slide 25: Reliable Measurements Also Under Condensing Environments

    Reliable Measurements Also Under Condensing Environments HPP272 stability after 200 times at H2O2 cycles (400-500 ppm, RS~100%RS) 0.5 0.45 0.4 Measurement error [%RH / %RS] 0.35 0.3 0.25 0.2 0.15 0.1 0.05 0 0 10 20 30 40 50 60 70 80 90 100 Relative humidity [%RH] RH RS From Monitoring to Controlling

  26. Slide 26: Accurate H2O2 and Humidity Measurements

    Slide 26: Accurate H2O2 and Humidity Measurements

    Accurate H2O2 and Humidity Measurements  Traceable factory calibration  Traceable to the international system of units (SI) through national metrology institutes  Quality of probe materials  Limited absorption, desorption and catalyzing of H2O2  Materials selected according to the rigorous and

  27. Slide 27: Cleamix Technology Highlights

    Slide 27: Cleamix Technology Highlights

    Cleamix Technology Highlights © Vaisala

  28. Slide 28: Cleamix VCS Series: A New Method for Generating Vaporized H2O2

    Slide 28: Cleamix VCS Series: A New Method for Generating Vaporized H2O2

    Cleamix VCS Series: A New Method for Generating Vaporized H2O2  A powerful, portable bio-decontamination unit having the size of a briefcase.  Performance experiments conducted in 3 different enclosures: Size of enclosure/ ppm/H2O2 vapor time/min m3 10 > 800 15 36 > 600 30 68 > 500 50  One litre

  29. Slide 29: Variables That Influence Hydrogen Peroxide Vapor Concentration

    Slide 29: Variables That Influence Hydrogen Peroxide Vapor Concentration

    Variables That Influence Hydrogen Peroxide Vapor Concentration © Vaisala

  30. Slide 30: Measurable Variables That Impact H2O2 Vapor Concentration

    Slide 30: Measurable Variables That Impact H2O2 Vapor Concentration

    Measurable Variables That Impact H2O2 Vapor Concentration  Humidity  Temperature From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 30

  31. Slide 31: Other Enclosure Variables That Impact H2O2 Vapor Concentration

    Slide 31: Other Enclosure Variables That Impact H2O2 Vapor Concentration

    Other Enclosure Variables That Impact H2O2 Vapor Concentration  Total surface area of the space or room to be decontaminated  Properties of the surface materials:  Porosity  Tendency to break down hydrogen peroxide molecules  Temperature variations of the surfaces  The cooler the surface than

  32. Slide 32: How Measurements Can Help You Provide Efficient H2O2 Vapor Output?

    Slide 32: How Measurements Can Help You Provide Efficient H2O2 Vapor Output?

    How Measurements Can Help You to Provide Efficient H2O2 Vapor Output? © Vaisala

  33. Slide 33: Sensor to PLC Connection for Control

    Slide 33: Sensor to PLC Connection for Control

    Sensor to PLC Connection for Control  Sensor must be rugged enough to maintain accuracy through multiple processes.  Testing determined that available sensing technologies were unsatisfactory.  EXCEPT with Vaisala’s HPP272 probe… From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, H

  34. Slide 34: Why We Selected Vaisala's Technology?

    Slide 34: Why We Selected Vaisala's Technology?

    Why We Selected Vaisala's Technology?  The probe provides all the parameters you need:  Hydrogen peroxide vapour  Temperature  Humidity as both relative saturation and relative humidity  Options for dew point and vapor pressure  The HPP270 probe with digital output signal  Analog signals are

  35. Slide 35: Things to Consider When Planning Bio-Decontamination Processes

    Slide 35: Things to Consider When Planning Bio-Decontamination Processes

    Things to Consider When Planning Bio- Decontamination Processes  Evaluate applicable process parameters  H2O2 vapor concentration and exposure time, required to eliminate contamination from surfaces  Consider environmental factors  Relative saturation limit value is the dominant variable  Hydro

  36. Slide 36: Controlling Condensation

    Slide 36: Controlling Condensation

    Controlling Condensation  Real time monitoring of relative saturation is needed to maintain dry conditions with minimum condensation  Challenges:  Optimizing the hydrogen peroxide vapour generation rate to fit small decontamination volumes – Difficulties in controlling the hysteresis  Condensati

  37. Slide 37: Validation of Decontamination Cycles

    Slide 37: Validation of Decontamination Cycles

    Validation of Decontamination Cycles  In order to validate a process, accurate, real time information on cycle parameters is necessary  When a desired H2O2 ppm level is easily attained, the total treatment time can be relatively short and the time needed for aeration is propably short as well  Wh

  38. Slide 38: Case Study: Hospital Room Bio-Decontamination

    Slide 38: Case Study: Hospital Room Bio-Decontamination

    Case Study: Hospital Room Bio- Decontamination © Vaisala

  39. Slide 39: Challenges in Infection Prevention

    Slide 39: Challenges in Infection Prevention

    Challenges in Infection Prevention  Suboptimal cleaning practises  Important pathogens not eliminated from surfaces:  Clostridium Difficile (C.diff)  Methicillin-resistant Staphylococcus aureus (MRSA)  Vancomycin-resistant Enterococcus (VRE)  Multidrug-resistant Acinetobacter  Risk of infecti

  40. Slide 40: No Touch Decontamination Solution

    Slide 40: No Touch Decontamination Solution

    No Touch Decontamination Solution  Automated technology – exceeds human performance  More intensive decontamination  Reliable process monitoring and control From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 40

  41. Slide 41: Hospital Room Decontamination

    Slide 41: Hospital Room Decontamination

    Hospital Room Decontamination REMOVE OPEN REMOVE From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 41

  42. Slide 42: Hospital Room Decontamination

    Slide 42: Hospital Room Decontamination

    Hospital Room Decontamination Cleamix H2O2 vapor generator HPP272 PROBE From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 42

  43. Slide 43: Hospital Room Decontamination

    Slide 43: Hospital Room Decontamination

    Hospital Room Decontamination Cleamix H2O2 vapor generator HPP272 PROBE Oscillating fans From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 43

  44. Slide 44: Hospital Room Decontamination

    Slide 44: Hospital Room Decontamination

    Hospital Room Decontamination BIs and CIs Cleamix H2O2 vapor generator HPP272 PROBE Oscillating fans From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 44

  45. Slide 45: Hospital Room Decontamination

    Slide 45: Hospital Room Decontamination

    Hospital Room Decontamination BIs and CIs Cleamix H2O2 vapor generator HPP272 PROBE Oscillating fans Catalytic converter From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 45

  46. Slide 46: Monitoring of the Decontamination Cycles

    Slide 46: Monitoring of the Decontamination Cycles

    Monitoring of the Decontamination Cycles  Control:  Via cable or WiFi – Touch screen and computer – Tablet – Phone  Accurate  Real time  Documented From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 46

  47. Slide 47: Safety Concerns

    Slide 47: Safety Concerns

    Safety Concerns  Post warning signs around perimeter to restrict access to H2O2 exposed area  In the event you must enter the process area, wear appropriate protective clothing and mask  Use handheld meter to detect any leaks of H2O2  Wait until proper aeration to enter area From Monitoring to C

  48. Slide 48: Conclusion

    Slide 48: Conclusion

    Conclusion  By moving from monitoring to controlling you can  Control the process based on real results  Reduce batch variability  Reduce validation efforts  Repeatable, stable and accurate PEROXCAP technology provides easy way to  Keep the H2O2 ppm value at a desired level  Control the proce

  49. Slide 49: Would you like a Sales Engineer to Contact You?

    Slide 49: Would you like a Sales Engineer to Contact You?

    Would you like a Sales Engineer to Contact You?  Yes, Vaisala Sales Engineer (HPP271, HPP272, PEROXCAP®)  Yes, Cleamix Sales Engineer (VCS-100 vapor generators) From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 49

  50. Slide 50: Q&A Session

    Slide 50: Q&A Session

    Q&A Session From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-06-18 © Vaisala 50

  51. Slide 51: Thank You for Attending!

    Slide 51: Thank You for Attending!

    Thank You for Attending! Vaisala:  HPP270 Series data sheet  Learn more about Bio-decontamination – Blogs, Videos, Application Notes  More information about HPP270 series products Cleamix:  Video of VCS100 From Monitoring to Controlling with Vaporized H2O2 Sensors - Why, How and Case Study 2019-