Covidien — Technical Presentation (April 2015)
31 slides from Covidien's technical presentation, April 2015
Slide 1: Nellcor Bedside SpO2 Patient Monitoring System, PM100N In-Service Presentation
Nellcor™ Bedside SpO2 Patient Monitoring System, PM100N In-Service Presentation COVIDIEN, COVIDIEN with logo, Covidien logo and positive results for life are U.S. and internationally registered trademarks of Covidien AG. Other brands are trademarks of a Covidien company. ©2015 Covidien. – EU-15-5745
Slide 2: Intended use of the PM100N
The Nellcor™ Bedside SpO2 Patient Monitoring System, PM100N, is intended for prescription use only for continuous monitoring of functional arterial oxygen saturation and pulse rate of adult, pediatric and neonatal patients, who are well or poorly perfused, during motion and no motion conditions. It
Slide 3: New Homecare and Sleep Study Modes
New Homecare and Sleep Study Modes Nellcor™ Bedside SpO2 Patient Monitoring System, PM100N is the only bedside monitor with standard, homecare, and sleep study modes. • Standard Mode: Advanced functionality to monitor in-hospital patients with SpO2, pulse rate, and SatSeconds alarm management. • Hom
Slide 4: Rear Panel
Rear Panel 1. Nurse Call Port 2. AC Power Connector 3. Battery Cover
Slide 5: Front & Side Panel
Front & Side Panel Audio Pause Silence audible Home Button alarms for set time Return to this main screen Power Button Press and HOLD to turn on and off USB Ports For data download and upgrades SpO2 Connector Knob Insert interface cable here Rotate and press, to control system functions
Slide 6: Display – Right Side
Display – Right Side AC Power Indicator Battery Charge Indicator Interference Indicator Sensor Disconnect Indicator Sensor Off Indicator Sensor Message Indicator
Slide 7: Hands on Review
Hands on Review • Attach Power Cord to device and plug into AC receptacle • Power on device and determine when it is ready for use • Connect cable and sensor to device • Create the various sensor indicator messages – Interference indicator – Sensor disconnect indicator – Sensor off indicator – Senso
Slide 8: Display – Main View
Standard Monitoring Display – Main View Pulse Rate: Battery SpO2: Upper/Lower Alarm Limits Real Time Value Indicator SpO2: Real Time Value Plethysmographic Blip Bar Pulse Rate: Upper/Lower Alarm Limits Waveform (Pleth) Informative Message Area
Slide 9: Display – Bottom
Standard Monitoring Display – Bottom Informative Message Area Patient Mode Alarm Limits Menu Options Menu 9 Covidien Respiratory & Monitoring Solutions | 4/10/15 |
Slide 10: Alarm Limits Menu
Standard Monitoring Alarm Limits Menu To access the Alarm Limits Menu, rotate the knob to highlight the alarm limits display area and press. Alarm Paused Icon Adjust Up/Down SpO2, PR, SatSeconds Alarm Icon Status Alarm Active Alarm Audio Paused Alarm audio deactivated in Service Menu Alarm Audio ico
Slide 11: Quick Access Menus
Standard Monitoring Quick Access Menus For quick access to alarm settings, rotate the knob to highlight one of two areas and press. SpO2 Alarm Settings SatSeconds Settings Pulse Rate Alarm Settings
Slide 12: Hands on Review
Hands on Review • Attach the sensor to a patient – Determine the patient’s real time SpO2 and PR reading – Set parameter alarm limits • SpO2 target range: 85% to 100% • Pulse rate target range: 80 bpm to 110 bpm • Once alarm limits are set – Discuss alarm status • What do you see? • What do you hear
Slide 13: Patient Mode Menu
Standard Monitoring Patient Mode Menu To access the Patient Mode Menu, turn the knob to highlight the patient mode display area and press the knob. Patient Mode Menu
Slide 14: Options Menu
Standard Monitoring Options Menu To access the Options Menu, turn the knob to highlight the Options Icon Press as a shortcut to and press the knob. access the Options Menu Volume Adjustment • Alarm • Key Beep • Pulse Patient Response Mode • Normal Clear Trend Data • Fast Service Menu 1. Power On Set
Slide 15: Hands on Review
Hands on Review • Your next patient is a 1 year old child. – Set the pulse oximeter to the appropriate patient mode. • It is very loud in the patient’s room. – Set the alarm volume to the loudest setting for SpO2 and PR.
Slide 16: SpO2 Waveform Menu
Standard Monitoring SpO2 Waveform Menu To access the SpO2 Waveform Menu, turn the knob to highlight the waveform or trend display area and press the knob. Sweep Speed • 6.25mm/s • 12.5mm/s • 25mm/s Trend Data • 88,000 data events • 96 hours • Collected every 4 seconds
Slide 17: Sweep Speed Overview
Standard Monitoring Sweep Speed Overview Allows the clinician to set the speed to where the SpO2 waveform trace moves across the screen. Sweep Speed = 25.0mm/s (i.e. “Zoom In”) The higher the sweep speed value, the more granular the data appears on the screen. Sweep Speed = 6.25mm/s (i.e. “Zoom Out”
Slide 18: Tabular Trend View
Standard Monitoring Tabular Trend View To access the Tabular Trend View, turn the knob to highlight the waveform or trend display area and press the knob. Select Tabular Trend and press the knob. Date / Scroll Time Interval SpO2 Pulse Monitor Value Rate errors Value
Slide 19: Tabular Trend Scroll
Standard Monitoring Tabular Trend Scroll To scroll through the Tabular Trend View, rotate the knob to the left for historical data and to the right to get back to present time Scroll Interval Options • 1 (every 4 sec) • 5 (every 20 sec) • 100 (~6.5 min) • 500 (~33.5 min) Turn knob to the left for hi
Slide 20: Trend Data Download
Standard Monitoring Trend Data Download To access the Trend Data Download Menu, turn the knob to highlight the Options Menu and press. Mini-USB Port 1. Connect mini-USB port to a PC 2. Select Options 3. Select Trend Data Download 4. Select Start
Slide 21: Graphical Trend View
Standard Monitoring Graphical Trend View To access the Graphical Trend View, turn the knob to highlight the waveform or trend display area and press the knob. Select Graphical Trend and press the knob. Pulse Rate SpO2 Trend Scale Line (blue line) Pulse Rate SpO2 Trend Line Scale (green line)
Slide 22: Hands on Review
Hands on Review • Access tabular trend data for your patient – How have your patient’s SpO2 values changed in the last hour? – How have your patient’s PR values changed in the last hour? • Use the scroll interval options to review your patient’s data from 3 hours ago. • Download your patient’s data
Slide 23: SatSeconds Alarm Management Setting
Standard Monitoring SatSeconds™ Alarm Management Setting SatSeconds Setting SatSeconds Icon
Slide 24: Alarm Limits Menu
Standard Monitoring Alarm Limits Menu To access SatSeconds™ Alarm Limits Menu, turn the knob to highlight the alarm limits display area and press the knob. SatSeconds Value Can be adjusted to off, 10, 25, 50, 100
Slide 25: Hands on Review
Hands on Review • Activate SatSeconds™ Alarm Management – If you want to minimize clinically insignificant alarms that occur with your patient, set SatSeconds to 100. – How would you respond to a SatSeconds Alarm?
Slide 26: Homecare Mode
Homecare Mode To access Homecare Mode, turn the knob to highlight and select the Options Menu. Within menu, select the Change Patient Mode display area and press the knob. Turn knob to highlight Change Patient Mode and press the knob
Slide 27: Homecare Mode
Homecare Mode After selecting Change Patient Mode display area, highlight Switch to Homecare and press the knob. Turn the knob to highlight Switch to Homecare and the press knob.
Slide 28: Homecare Mode
Homecare Mode After selecting Switch to Homecare display area, enter passcode. Entering Passcode: • Turn the knob to highlight the number in passcode and press the knob. • Once number is highlighted in yellow, turn knob to change number then press the knob to select. • Once number is selected, repea
Slide 29: Homecare Mode
Homecare Mode After passcode is entered, a confirmation screen will pop up. Select OK and then you will be in Homecare Mode. Select OK and press the knob Note: The Homecare mode monitoring screen does not appear until the user presses the knob to select OK
Slide 30: Sleep Study Mode
Sleep Study Mode To access Sleep Study Mode, turn the knob to highlight the Options Menu and select the Change Patient Mode display area and press the knob. Turn the knob to highlight Change Patient Mode and press the knob
Slide 31: Sleep Study Mode
Sleep Study Mode After selecting the Change Patient Mode , then highlight Switch to Sleep Study and press the knob. Sleep Study Mode