Inspiremd Inc. · NASDAQ:NSPR

Inspiremd — Technical Presentation (October 2013)

25 slides from Inspiremd's technical presentation, Oct 29, 2013

  1. Slide 1: MASTER Trial 12 Month Results

    MASTER Trial 12 Month Results Prof. Dariusz Dudek Hospital University of Krakow, Poland on behalf of the MASTER Trial Investigators

  2. Slide 2: Disclosure Statement of Financial Interest

    Disclosure Statement of Financial Interest • Grant/Research Support • Consulting Fees/Honoraria • Major Stock Shareholder/Equity • Royalty Income • Ownership/Founder • Intellectual Property Rights • Other Financial Benefit • InspireMD • No • No • No • No • No • No Within the past 12 months, I or my

  3. Slide 3: Background

    Background » Impaired myocardial perfusion after PCI in STEMI is common, and results in increased infarct size, heart failure and mortality. » PCI - induced distal embolization of thrombus and/or friable atheromatous debris contributes to impaired myocardial perfusion. » Efforts to improve myocardia

  4. Slide 4: MGuard stent: designed for embolic protection

    MGuard stent: designed for embolic protection » MGuard is a thin - strut BMS wrapped with an expandable MicroNet mesh, designed to trap, exclude and secure the thrombus and friable atheromatous debris to prevent distal embolization during and post procedure. » The mesh is made of a 20  M Polyethyle

  5. Slide 5: MGuard & MGuard Prime Stent

    MGuard & MGuard Prime STENT MGuard MGuard Prime MATERIAL Stainless steel Cobalt chromium STRUT THICKNESS 100  m 80  m CROSSING PROFILE 1.1 - 1.3mm 1.0 - 1.2mm MICRONET MATERIAL PET FIBER DIAMETER 20  M PORE SIZE 150 - 180  M

  6. Slide 6: MASTER Trial design

    MASTER Trial design Primary Endpoint: Complete ST - segment resolution at 60 - 90 min Secondary endpoints: TIMI flow, Myocardial blush grade, MACE ( 30 d, 6 m, 12 m) Substudies: Cardiac MRI at 3 - 5 days ( 2 x 30 patients) Angiographic follow - up at 13 months ( 50 patients) 432 patients with STEMI

  7. Slide 7: Enrollment in participating countries

    Enrollment in participating countries Czech republic (19) Poland ( 166 ) Hungary ( 68 ) Germany ( 45 ) Brazil ( 14 ) South Africa (10) France ( 9 ) Israel (100) Ireland (2)

  8. Slide 8: Main Inclusion / exclusion criteria

    Main Inclusion / exclusion criteria INCLUSION CRITERIA » Symptoms consistent with STEMI within 12 hours of symptom onset » ≥2 mm of ST - segment elevation in ≥2 leads » PCI of a single de novo lesion with RVD ≥3.0 to ≤4.0 mm » Lesion length ≤33 mm EXCLUSION CRITERIA » LBBB, paced rhythm, etc. (unabl

  9. Slide 9: Baseline characteristics

    Baseline characteristics MGUARD (N=217) CONTROL (N=216) Age (years) 60 [52, 68] 58 [51, 67] Male 75.1% 76.9% Hypertension 42.3% 47.4% Hyperlipidemia 27.4% 27.1% Diabetes mellitus 12.0% 18.1% Cigarette smoking 55.3% 46.8% Prior MI 3.7% 8.8% Prior PCI 3.7% 5.6% Symptoms to device, mins 207 [156, 308]

  10. Slide 10: Procedural characteristics

    Procedural characteristics *Out of 217 patients, 27 were treated with MGuard Prime Stone GW et al. J Am Coll Cardiol . 2012;60:1975 - 1984 MGUARD (N=217) CONTROL (N=216) P Aspiration performed 65.9% 67.1% 0.79 Balloon pre - dilatation performed 50.2% 44.9% 0.27 Direct stenting 12.0% 10.6% 0.66 > 1 s

  11. Slide 11: Epicardial coronary flow

    Epicardial coronary flow 1.8 6.5 91.7 5.6 11.6 82.9 0 10 20 30 40 50 60 70 80 90 100 0/1 2 3 Percent (%) TIMI flow MGuard Control stent P=0.006 P= 0.01 Stone GW et al. J Am Coll Cardiol . 2012 ; 60 : 1975 - 1984

  12. Slide 12: ST-segment resolution

    ST - segment resolution Stone GW et al. J Am Coll Cardiol . 2012;60:1975 - 1984 16.7 25.5 57.8 17 38.3 44.7 0 10 20 30 40 50 60 Absent (≤ 30 %) Partial (>30% - <70%) Complete (≥70%) Percent (%) ST resolution MGuard Control stent P= 0.005 P=0.008 PRIMARY ENDPOINT

  13. Slide 13: Distal embolization

    Distal embolization Stone GW et al. J Am Coll Cardiol . 2012 ; 60 : 1975 - 1984 4.1 7.0 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 Distal Embolization Percent (%) MGuard Control stent P= 0.21

  14. Slide 14: Cardiac MRI sub-study

    Cardiac MRI sub - study at 5 days Stone GW et al. J Am Coll Cardiol . 2012;60:1975 - 1984 MGUARD (N=30) CONTROL BMS / DES (N=29) P Total LV myocardial mass, g 141 [117, 163] 147 [118, 174] 0.41 Infarct mass, g 17.1 [10.0, 30.0] 22.3 [15.7, 30.1] 0.27 Infarct mass (% total LV mass) 13.3 [7.9, 25.0] 1

  15. Slide 15: 30-day clinical results

    30 - day clinical results Stone GW et al. J Am Coll Cardiol . 2012;60:1975 - 1984 MGUARD (N=217) CONTROL (N=214) P MACE 4 (1.8%) 5 (2.3%) 0.75 All cause mortality 0 (0.0%) 4 (1.9%) 0.06 Cardiac death 0 (0.0%) 4 (1.9%) 0.06 Reinfarction 3 (1.4%) 2 (0.9%) 1.00 TLR, ischemia - driven 4 (1.8%) 1 (0.5%)

  16. Slide 16: 12 month clinical results

    12 month clinical results NEW RESULTS MGUARD (N=217) CONTROL (N=216) P MACE 9.1% (19) 3.3% (7) 0.02 All cause mortality 1.0% (2) 3.3% (7) 0.09 Cardiac death 0.5% (1) 2.3% (5) 0.10 Reinfarction 1.4% (3) 0.9% (2) 0.66 Death or reinfarction 2.3% (5) 3.7% (8) 0.39 TLR, ischemia - driven 8.6% (18) 0.9% (

  17. Slide 17: Propensity score matching (TLR / TVR)

    Propensity score matching ( TLR / TVR ) HORIZONS - AMI BMS N=443 MASTER MGuard N=191 P MACE 7.5% (33) 9.2% (17) 0.54 TLR, ischemia - driven 6.0% (26) 8.7% (16) 0.26 TVR, ischemia - driven 6.9% (30) 10.3% (19) 0.18 Propensity Score Adjustment on the following covariates (3 - 1 caliper = 0.35 ): Age,

  18. Slide 18: Death Time-to-Event curve

    Death Time - to - Event curve NEW RESULTS Death (%) 0 1 2 3 4 5 Months 0 1 2 3 4 5 6 7 8 9 10 11 12 217 214 209 206 126 216 210 209 207 123 Number at risk: MGuard BMS/DES P= 0.09 HR: 0.28 [ 95 % CI: 0.06 , 1.36 ] 1.0% 3.3% MGuard BMS/DES

  19. Slide 19: Cardiac Death Time-to-Event curve

    Cardiac Death Time - to - Event curve Cardiac Death - 12 months (%) 0 1 2 3 4 5 Time in Months 0 1 2 3 4 5 6 7 8 9 10 11 12 217 214 209 206 126 216 210 210 207 124 Number at risk: MGuard BMS/DES P= 0.099 HR: 0.20 [ 95 % CI: 0.02 , 1.69 ] 0.5 % 2.3% MGuard BMS/DES NEW RESULTS

  20. Slide 20: Death or MI Time-to-Event curve

    Death or MI Time - to - Event curve DEATH/MI (%) 0 1 2 3 4 5 Months 0 1 2 3 4 5 6 7 8 9 10 11 12 217 212 207 205 125 216 209 208 206 123 Number at risk: MGuard BMS/DES 2.3% 3.7 % MGuard BMS/DES P= 0.39 HR: 0.62 [ 95 % CI: 0.20 , 1.89 ] NEW RESULTS

  21. Slide 21: 13 Month angiographic sub-study follow-up

    13 Month angiographic sub - study follow - up » Enrollment into the angiographic sub - study started after the study enrolled 100 patients, and occurred in pre - specified sites. » 48 consecutively enrolled patients randomized to the MGuard were consented to return at 13 months for an invasive angio

  22. Slide 22: 13 Month angiographic sub-study follow-up results

    13 Month angiographic sub - study follow - up ENTIRE SUBSTUDY COHORT (N=36)* SINGLE MGUARD STENT (N=31)** Average 95% CI Average 95% CI Late loss (in stent) 0.99 ± 0.80 [0.73,1.26] 0.88 ± 0.70 [0.63,1.14] Late loss (in segment) 0.82 ± 0.75 [0.57,1.06] 0.72 ± 0.65 [0.48,0.95] Binary Restenosis (in st

  23. Slide 23: Limitations

    Limitations » The MASTER trial was powered for ST - segment resolution, and not for infarct size or clinical events. » Endpoints other than ST - segment resolution should be considered exploratory and hypothesis - generating. » The control arm in MASTER consisted of a mixture of patients treated wit

  24. Slide 24: Conclusions

    Conclusions » Among patients with acute STEMI undergoing emergent PCI, MGuard EPS resulted in superior rates of TIMI3 and complete STR. » A trend towards reduced mortality was present in the MGuard arm at 30 days, which persisted throughout 12 month follow - up. » No significant differences in reinf

  25. Slide 25: MASTER-II IDE study design

    MASTER - II IDE study design BMS or DES MGuard Prime Randomization 1 : 1 STEMI patients for primary PCI 1114 patients at 35 US and 35 non - US sites Infarct size assessed by cardiac MRI In - stent late lumen loss (LLL) at 13m (BMS) Complete ST - segment resolution at 60 - 90 min All - cause death or