  1. Slide 1: OraQuick ADVANCE Rapid HIV-1/2 Antibody Test

    OraQuick ADVANCE® Rapid HIV-1/2 Antibody Test

  2. Slide 2: The Importance of Routine Rapid HIV Testing

    The Importance of Routine Rapid HIV Testing

  3. Slide 3: Approximately 1 in 5 HIV+ people do not know it, causing over 50% of new infections

    While approximately 1 in 5 people who are HIV+ do not know it, they cause over 50% of new infections ~20% Unaware of ~50-70% of New Infection Infections ~80% Aware of ~30-50% of New Infection Infections People living with HIV/AIDS ~1.1 million New infections per year ~56,300 AMA Jour of Ethics, Dec

  4. Slide 4: OraQuick ADVANCE Rapid HIV-1/2 Antibody Test

    OraQuick ADVANCE® Rapid HIV-1/2 Antibody Test • Simple • Rapid HIV-1/2 antibody testing with oral fluid collection – results in 20 minutes • Accurate • Results with >99% sensitivity and specificity across all specimen types • Versatile • Testing platform suitable for both clinical and non-clinical s

  5. Slide 5: Overview of the OraQuick ADVANCE Rapid HIV-1/2 Antibody Test

    Overview of the® OraQuick ADVANCE Rapid HIV-1/2 Antibody Test

  6. Slide 6: Intended Use

    Intended Use ® • The OraQuick ADVANCE Rapid HIV-1/2 Antibody Test is a single-use, qualitative immunoassay to detect antibodies to Human Immunodeficiency Virus Type 1 and 2 (HIV-1/2) in oral fluid, venipuncture whole blood, fingerstick whole blood, and plasma specimens. • For in vitro diagnostic use

  7. Slide 7: Clinical Laboratory Improvement Amendments (CLIA)

    Clinical Laboratory Improvement Amendments (CLIA) CLIA Certificate of Waiver • The Centers for Medicare & Medicaid Services (CMS) regulates all laboratory testing (except research) performed on humans in the U.S. through the Clinical Laboratory Improvement Amendments (CLIA) • The objective of the CL

  8. Slide 8: OraQuick ADVANCE is a CLIA-Waived Test

    OraQuick ADVANCE ® is a CLIA-Waived Test CLIA Certificate of Waiver • OraQuick ADVANCE® Rapid HIV-1/2 Antibody Test is CLIA-Waived for the following samples: • Oral fluid • Fingerstick whole blood • Venipuncture whole blood

  9. Slide 9: OraQuick ADVANCE Rapid HIV-1/2 Antibody Test

    OraQuick ADVANCE ® Rapid HIV-1/2 Antibody Test • Single-use testing device with Reusable Developer Test built-in procedural control Solution Stand Vial Collection • Single-use test developer Loop solution vial Absorbent Packet Flat • Reusable test stand Pad Test Result Device • Disposable single-use

  10. Slide 10: Clinical Features: Operating Principle

    OraQuick ADVANCE ® Clinical Features Operating Principle Anti-Human Antibody Colloidal Gold Human Antibodies Conjugated HIV Antigen to Protein A HIV-1/2 Antibodies Control Test Conjugate pad with colloidal Line Line gold particle bound to protein A. Cellulose Flat Membrane Pad Using a lateral flow p

  11. Slide 11: Clinical Features: Operating Principle

    OraQuick ADVANCE ® Clinical Features Operating Principle Anti-Human Antibody Colloidal Gold Human Antibodies Conjugated HIV Antigen to Protein A HIV-1/2 Antibodies Control Test “T” “C” Line Line Absorbent Pad Colloidal gold particles containing HIV antibodies bind to the HIV antigen “T” line forming

  12. Slide 12: Clinically Proven with >99% Accuracy

    Clinically Proven with >99% Accuracy • What defines accuracy? Sensitivity: The ability to detect a true positive. Specificity: The ability to detect a true negative

  13. Slide 13: Clinically Proven with >99% Accuracy

    Clinically Proven with >99% Accuracy Claim Sensitivity Specificity Claim Sensitivity Specificity HIV-2* 100.0% - Oral Fluid 99.3% 99.8% Fingerstick 99.6% 100.0% Whole Blood Plasma 99.6% 99.9% *HIV-2 (Based on >800 banked serum/plasma specimens. In addition, FSWB and OF tests were done on 3 HIV-2 inf

  14. Slide 14: Tips to Avoid False Positive and False Negative Results

    Tips to Avoid False Positive and False Negative Results • In order to ensure accurate results are obtained, specimen collection is important. The following may cause inaccurate results: • Improper oral fluid specimen collection (over or under swabbing) • Not following proper fingerstick collection p

  15. Slide 15: Test Kit - Available Packaging

    Test Kit - Available Packaging Kit Size 100 Count 25 Count Item No. 1001-0078 1001-0079 Reusable Test Stand 10 5 Specimen Collection Loops 100 25 Subject Information Pamphlets 100 25 Package Insert 1 1 Customer Letter 1 1 Shelf Life 30 Months from Date of Manfacture Storage Requirements 2—27°C (36 —

  16. Slide 16: Additional Materials Required and Available as an Accessory Kit

    Additional Materials Required and Available as an Accessory Kit OraQuick ADVANCE® Kit Controls Positive Controls • Black-capped vial - inactivated human plasma positive for antibodies to HIV-1 • Red-capped vial - inactivated human plasma positive for antibodies to HIV-2 Negative Control • White-capp

  17. Slide 17: Additional Materials Required

    Additional Materials Required • Timer or Watch • Biohazard Waste Container • Disposable, Absorbent Workspace Cover Additional Required Phlebotomy Materials (WB, Plasma): • Disposable Gloves • Sterile Lancet • Phlebotomy materials • Centrifuge • Antiseptic Wipe • Sterile Gauze Pads

  18. Slide 18: Test Preparation

    Test Preparation Rvs —, ee. ee < — QraSure Technologies, Inc. =Or. a Qu IC. Kk me ADVANCE?

  19. Slide 19: Prior to Testing

    Prior to Testing • Remember to observe “Universal Precautions” at all times. • Read product instructions first. • Gather testing materials. • Let test come to operating temperature. • Set up workspace cover and reusable Test Stand on a flat level surface. • Put on disposable gloves if working with b

  20. Slide 20: General Test Preparation

    General Test Preparation • Check expiration date on OraQuick ADVANCE® Pouch. • Open two chambers of Divided Pouch by tearing at the notches. • Leave the Test Device in the Pouch. • Remove the Developer Vial. Gently rock the cap back and forth to remove. • Slide the Vial into the top of one of the sl

  21. Slide 21: General Test Preparation

    General Test Preparation • Remove test device from Pouch. DO NOT touch the flat pad. • Label device with subject information. DO NOT block holes on back of device. NOTE: Test Device must be inserted into vial within 30 minutes of sample introduction. • Make sure an Absorbent Packet is present. If no

  22. Slide 22: Specimen Collection and Test Initiation

    Specimen Collection and Test Initiation

  23. Slide 23: Universal Precautions

    Universal Precautions Handling of Potentially Infectious Human Samples • Before handling any specimens please refer to your facility’s procedures on universal precautions. • Universal guidelines stress that all patients should be assumed to be infectious for blood-borne diseases such as AIDS and hep

  24. Slide 24: Prior to Oral Fluid Specimen Collection

    Prior to Oral Fluid Specimen Collection • Ensure prior to testing that the subject has not had anything to eat, drink (including water) or has chewed gum for at least 15 minutes. • Have the subject wait for at least 30 minutes prior to testing if they have used any oral care products.

  25. Slide 25: Oral Fluid – Specimen Collection

    Oral Fluid – Specimen Collection • Place the flat pad above above the teeth against the outer gum. • Gently swab completely around the outer gums, both upper and lower, one time around, using the Flat Pad. • DO NOT swab the roof of the mouth, the inside of the cheek or the tongue. • NOTE: It is okay

  26. Slide 26: If Device is transported prior to testing

    If Device is transported prior to testing • NOTE: When collecting oral fluid specimens, the Test Device must be inserted into Developer Solution vial within 30 minutes of collection • Blood and plasma insertion within 60 minutes • A test device containing an oral fluid specimen that is not inserted

  27. Slide 27: Oral Fluid – Test Procedure

    Oral Fluid – Test Procedure • Insert Flat Pad of device into the bottom of Developer Vial. • Start timing test. • Pink fluid will travel up Result Window. Fluid disappears as test develops. DO NOT remove device while test is running. • Read results after 20 minutes but not more than 40 minutes. Adeq

  28. Slide 28: Fingerstick – Specimen Collection

    Fingerstick – Specimen Collection • Use an antiseptic wipe; clean finger of person being tested. Dry completely. • Use sterile lancet, puncture skin off center of finger pad. • WIPE first droplet with gauze. Hold the hand downward for new droplet. Gently apply pressure to express if needed. • With n

  29. Slide 29: Fingerstick – Mixing Specimen

    Fingerstick – Mixing Specimen • Insert blood-filled end of Loop into the vial. Be careful not to touch the sides of the vial. • Use Loop to stir sample in Vial. Dispose of used loop in biohazard waste container. • Check Solution to make sure it appears pink in color if using whole blood.

  30. Slide 30: Fingerstick – Test Procedure

    Fingerstick – Test Procedure • Insert Flat Pad of device into the bottom of Developer Vial. • Start timing test. • Pink fluid will travel up Result Window. Fluid disappears as test develops. DO NOT remove device while test is running. • Read results after 20 minutes but not more than 40 minutes. Ade

  31. Slide 31: Whole Blood – Specimen Collection

    Whole Blood – Specimen Collection • Use standard phlebotomy procedures, collect whole blood sample with one of the following collection tubes: • EDTA, Sodium Heparin, or Sodium Citrate. • Mix blood tube by inversion. • With new Specimen Collection Loop, dip loop into test tube. • Visually inspect th

  32. Slide 32: Whole Blood – Mixing Specimen

    Whole Blood – Mixing Specimen • Insert blood-filled end of Loop into the vial. Be careful not to touch the sides of the vial. • Use Loop to stir sample in Vial. Dispose of used loop in biohazard waste container. • Check Solution to make sure it appears pink in color if using whole blood.

  33. Slide 33: Whole Blood – Test Procedure

    Whole Blood – Test Procedure • Insert Flat Pad of device into the bottom of Developer Vial. • Start timing test. • Pink fluid will travel up Result Window. Fluid disappears as test develops. DO NOT remove device while test is running. • Read results after 20 minutes but not more than 40 minutes. Ade

  34. Slide 34: Plasma – Specimen Collection

    Plasma – Specimen Collection • Collect a whole blood sample in a blood collection tube containing an anticoagulant. • Centrifuge the blood to separate cells from plasma. • With new Specimen Collection Loop, dip loop into the plasma specimen. • Visually inspect the Loop to make sure that it is comple

  35. Slide 35: Plasma – Mixing Specimen

    Plasma – Mixing Specimen • Insert plasma-filled end of Loop into the vial. Be careful not to touch the sides of the vial. • Use Loop to stir sample in Vial. Dispose of used loop in biohazard waste container.

  36. Slide 36: Plasma – Test Procedure

    Plasma – Test Procedure • Insert Flat Pad of device into the bottom of Developer Vial. • Start timing test. • Pink fluid will travel up Result Window. Fluid disappears as test develops. DO NOT remove device while test is running. • Read results after 20 minutes but not more than 40 minutes. Adequate

  37. Slide 37: Test Interpretation

    Test Interpretation =OraQuick ADVAN laa HIV-1/72

  38. Slide 38: Test Reading & Interpretation

    Test Reading & Interpretation • Non-reactive result • Reactive result • Invalid

  39. Slide 39: Reading a Non-Reactive Test

    Reading a Non-Reactive Test A test is NON-REACTIVE if: • A reddish-purple line appears next to the triangle labeled “C” and no line appears next to the triangle labeled “T”.

  40. Slide 40: Interpreting a Non-Reactive Test

    Interpreting a Non-Reactive Test A Non-Reactive test result means that HIV-1 and HIV-2 antibodies were not detected in the specimen. The test result is interpreted as NEGATIVE for HIV-1 and HIV-2 antibodies. Follow CDC Guidelines to inform subject of test result and interpretation.

  41. Slide 41: Reading a Reactive Test

    Reading a Reactive Test A test is REACTIVE if: • A reddish-purple line appears next to the triangle labeled “C” and a reddish- purple line appears next to the triangle labeled “T”. Lines may vary in intensity. • NOTE: The test is reactive if a complete reddish-purple line appears next to the “T” tri

  42. Slide 42: Interpreting a Reactive Test

    Interpreting a Reactive Test A Reactive test result means that HIV-1 and/or HIV-2 antibodies have been detected in the specimen. The test result is interpreted as PRELIMINARY POSITIVE for HIV-1 and/or HIV-2 antibodies. Follow CDC Guidelines to inform subject of test result and interpretation.

  43. Slide 43: Reading an Invalid Test

    Reading an Invalid Test a b c d e f g A test is INVALID if: • No reddish-purple lines appears next to the triangle labeled “C” (see picture a and b), or • A red background in the Results Window makes it difficult to read the result after 20 minutes (picture c), or • If any of the lines are NOT insid

  44. Slide 44: Interpreting an Invalid Test

    Interpreting an Invalid Test a b c d e f g An Invalid test result means that there was a problem running the test, either related to the specimen or to the Device. IT CANNOT BE INTERPRETED. Repeat test with a new Pouch and a new oral fluid, fingerstick or venipuncture whole blood, or plasma sample.

  45. Slide 45: General Test Clean Up

    General Test Clean Up • Dispose of the used test materials in a biohazard waste container. • When using gloves, change your gloves between each test to prevent contamination. Throw away the used gloves in a biohazard waste container. • Use a freshly prepared 10% solution of bleach to clean up any sp

  46. Slide 46: Quality Assurance

    Quality Assurance Rvs —, ee. ee < — QraSure Technologies, Inc. =Or. a Qu IC. Kk ADVAN CE®

  47. Slide 47: Quality Control

    Quality Control • Positive and Negative Kit Controls provide: • Quality Control to: • Assure test performance • Provide for user proficiency • Positive Controls • Are calibrated specifically to a very low assay reactivity level (challenge line). • Low assay performance reaffirms assay functionality

  48. Slide 48: Test Kit - Kit Controls

    Test Kit - Kit Controls Positive Controls • Black-capped vial - inactivated human plasma positive for antibodies to HIV-1 • Red-capped vial - inactivated human plasma positive for antibodies to HIV-2 Negative Control • White-capped vial - human plasma negative for antibodies to HIV-1 and HIV-2 • Suf

  49. Slide 49: Test Kit Control Contents

    Test Kit Control Contents #1001-0077 Test Kit Control Positive HIV-1 Control Vial (1) (Black Cap) 0.2 mL Positive HIV-2 Control Vial (1) (Red Cap) 0.2 mL Negative Control Vial (1) (White Cap) 0.2 mL Package Insert 1 Storage Requirements 2—8°C (35—46°F)* Shelf-Life 1 Year from manufacture date or 8 w

  50. Slide 50: Performing Kit Controls

    Performing Kit Controls Run one positive HIV-1 control (+), one positive HIV-2 control (+), and one negative control (-) for: • Each new operator, • Each new lot of test kits, • Each new shipment of test kits, • Test kit storage temperature falls outside 2-27°C; 35-80°F, • Testing area temperature f

  51. Slide 51: Performing Kit Controls

    Performing Kit Controls • Getting Started • Read the OraQuick ADVANCE® Kit Control Package Insert. • Remember to observe “Universal Precautions” at all times. • Follow directions for inspecting and opening pouches. • Follow directions for setting up workspace. • Label devices for Negative, Positive-

  52. Slide 52: Performing Kit Controls

    Performing Kit Controls • Open a Kit Control vial. • Insert round end of a new Specimen Collection Loop into the reagent vial. • Remember to use separate unused Loop for each control NOTE: The Kit Control reagents reagent. are clear to straw-colored. Do not use if the reagent appears cloudy or disco

  53. Slide 53: Performing Kit Controls

    Performing Kit Controls • Immerse Loop into Developer Solution Vial. • DO NOT touch side of vial. • Use the loop to stir contents. • Discard Loop in a Biohazard Waste Container.

  54. Slide 54: Performing Kit Controls

    Performing Kit Controls • Retrieve correct labeled device. • Insert Flat Pad of device into the bottom of Developer Vial. • Start timing test. • Pink fluid will travel up Result Window. Fluid disappears as test develops. DO NOT remove device while test is running. • Read results after 20 minutes but

  55. Slide 55: Performing Kit Controls

    Performing Kit Controls Example of a Example of a Expected Results: Non-Reactive Reactive Result Result (Preliminary • Negative Control will produce a (Negative) Positive) Non-Reactive test result. A line should be present at “C” triangle in result window. • Positive Controls will produce a Reactive

  56. Slide 56: Kit Control Failure

    Kit Control Failure • If test result does not perform as expected • Repeat test using new Test Device, Developer Solution Vial, and Control Specimen. • If test result does not perform a second time • Discontinue testing and contact OraSure Technologies.

  57. Slide 57: OraQuick ADVANCE Interpretation Review

    OraQuick ADVANCE ® Interpretation Review A B C D E F G 1-800-ORASURE • 1-800-672-7873

  58. Slide 58: OraQuick ADVANCE Interpretation Review

    OraQuick ADVANCE ® Interpretation Review Reactive Invalid Non‐Reactive Invalid Invalid Invalid Reactive A B C D E F G 1-800-ORASURE • 1-800-672-7873