Nemaura Medical — Technical Presentation (September 2018)

11 slides from Nemaura Medical's technical presentation, Sep 12, 2018

  1. Slide 1: Prospective Single Centre Evaluation of sugarBEAT CGM: Interim Evaluation

    A Prospective Single Centre Evaluation of the Accuracy and safety of the Nemaura Medical sugarBEAT® Non-invasive Continuous Glucose Monitor (CGM) System: An interim Evaluation Better Diagnostics for Life 12th September 2018 NASDAQ: NMRD Author Dr T Rahman, MBBS, PhD Clinical research Manager, Nemaur

  2. Slide 2: Study Objective

    Study Objective A EU-based study to determine the safety and accuracy of the sugarBEAT® CGM to support FDA submission 2

  3. Slide 3: Study Design

    Prospective single arm, single centre study Abbott Freestyle Optimum Neo BGM used as reference for Home Study Portion NEMAURA MEDICAL Accuracy and safety assessed over 7 consecutive Venous blood samples used as reference for in-clinic wear days, consisting of 3 non-consecutive in-clinic portion of s

  4. Slide 4: Study Methods

    Study Methods –– –– –– –– –– –– –– 4

  5. Slide 5: Baseline Characteristics

    Baseline Characteristics Variable Age - Range 27-67 years Age - Mean 52.8 years Age - Median 54.5 years Type 1 (n) 12 Type 2 (n) 13 5

  6. Slide 6: Accuracy vs Glucose Range In-clinic portion 1-point calibration

    Accuracy vs Glucose Range In-clinic portion of study 1-point calibration Glucose range Number of Within ±10% or Within ±15% or Within ±20% or Within ±30% or Within ±40% or Outside±40% (mg/dl) paired points ±10 mg/dl ±15 mg/dl ±20 mg/dl ±30 mg/dl ±40 mg/dl or ±40 mg/dl (MARD) (MARD) (MARD) (MARD) (MA

  7. Slide 7: Accuracy vs Glucose Range In-clinic portion 2-point Calibration

    Accuracy vs Glucose Range In-clinic portion of study 2-point Calibration Glucose range Number of Within ±10% or Within ±15% or Within ±20% or Within ±30% or Within ±40% or Outside±40% or (mg/dl) paired points ±10 mg/dl ±15 mg/dl ±20 mg/dl ±30 mg/dl ±40 mg/dl ±40 mg/dl (MARD) (MARD) (MARD) (MARD) (MA

  8. Slide 8: Interim Comparative MARD for In-Clinic phase

    Interim Comparative MARD for In-Clinic phase as Primary Accuracy Metric DexcomG5* no. of calibrations not known Overall MARD (20%/20mg/dL) 9.00 % Data 94.00 sugarBEAT 1-point calibration Overall MARD (20%/20mg/dL) 8.77 % Data 62.61 2-point calibration Overall MARD (20%/20mg/dL) 7.97 % Data 74.00 1-p

  9. Slide 9: Interim Comparative MARD for home-study phase

    Interim Comparative MARD for home-study phase as Primary Accuracy Metric Device No. of Subjects Paired Data Points with BGM Nominal MARD Reference BGM Senseonics Eversense* 23 829 14.80% Nova Biomedical StatStrip Xpress Dexcom G5* 23 829 16.30% Nova Biomedical StatStrip Xpress Abbott Libre Pro* 23 8

  10. Slide 10: Safety Evaluation

    Safety Evaluation Past Studies Current Studies –– –– –– 10

  11. Slide 11: Conclusions

    Conclusions The interim analysis indicates the 3 day in- clinic portion of the study shows accuracy levels, as measured using overall nominal MARD, as comparable to previously reported data and within the targets for the intended indications. 11