Waters Corporation · NYSE:WAT
Waters — Technical Presentation (May 2013)
57 slides from Waters's technical presentation, May 2013
Slide 1: Protein and Peptide Therapeutics: Addressing Bioanalytical Demands of Complicated Peptides
Thank you for joining us, the session will begin shortly… Protein and Peptide Therapeutics: Addressing Bioanalytical Demands of Complicated Peptides Tips and Tricks for Troubleshooting and Optimization During Method Development Mary E. Lame Senior Applications Chemist Waters Corporation Milford, MA
Slide 2: Friendly Reminders
Friendly Reminders… Please use text chat functionality to submit your questions today. Erin Chambers, Principal Applications Chemist, Waters Corp. “LIVE” Technical support during today’s event Upon conclusion, follow up information will be available: http://www.waters.com/May1 Recorded v
Slide 3: Mary Lame, Sr. Applications Chemist
Mary Lame, Sr. Applications Chemist B.S. in Chemistry from the Western Connecticut State University (Danbury, CT) M.S. in Chemistry from Central Connecticut State University (New Britain, CT). She began her bioanalytical career in 1997 with a global Pharmaceutical company where she supported R
Slide 4: Outline
Outline Introduction — Basic screening methods — Typical challenges faced Understanding your peptide o pI, HPLC index, hydrophobic/hydrophilic Understanding the challenges: symptoms, examples, and solutions for common problems Conclusions ©2012 Waters Corporation 4
Slide 5: Introduction: Why bioanalysis for peptides?
Introduction Why bioanalysis for peptides? 1. Drug discovery/development activities need to be performed o PK/PD, metabolic fate, bioequivalence, drug monitoring 2. Peptides as biomarkers 3. Signature peptides can be used to quantitate protein drugs and biomarkers in complex matrices, after digest
Slide 6: Why LC-MS/MS?
Why LC-MS/MS? Why an LC-MS/MS based assay? — ELISA assays not practical for discovery, no antibodies available yet — Challenges with ELISA assays o time consuming, expensive to develop o require separate assay for each peptide o limited linear dynamic range o Possible cross reactivity Benefits o
Slide 7: Chromatographic Screening Protocol for Peptides
Chromatographic Screening Protocol for Peptides ACQUITY UPLC® BEH300 C18 2.1 X 50 mm, 1.7 µm Peptide Separation Technology (PST) Column — Columns are QC tested with peptide standards — 300Å PST column gave overall best performance (peak shape) for diverse peptides — 2.1 X 50 mm provides adequate t
Slide 8: Single Screening Method: Diverse Peptides
Single Screening Method: Diverse Peptides 1 2 3 5 MRM of 5 Channels ES+ 4 MS Data Peak Points Width Across Analyte MW (seconds) Peak 1. Vasopressin 1084 1.8 15 2. Angiotensin II 1046 2.2 15 % 3. Desmopressin 1069 2.2 18 4. Bivalirudin 2180 2.4 18 5. Enfuvirtide 4492 2.1 16 0 Time 0.20 0.40 0.60 0.80
Slide 9: SPE Screening Protocol for Peptides
SPE Screening Protocol for Peptides: Screening Complementary Sorbents Oasis® WCX Oasis® MAX µElution µElution Protocol Dilute plasma with 4% H3PO4 Condition MeOH/Equilibrate H2O Load Diluted Plasma Wash 1: 5% NH4OH Wash 2: 20% ACN Elution: 1% TFA in 75/25 ACN/H2O Dilute: H2O ©2012 Waters Corporation
Slide 10: SPE Format: Oasis µElution Plates
SPE Format : Oasis® µElution Plates Oasis® µElution plate technology Up to 15X concentration without evaporation — Concentration often necessary to reach LOD’s with peptides Minimizes analyte loss — Minimizes sticking to walls of collection plates — Eliminates problems re-solubilizing after dry-
Slide 11: Final SPE Results after Methods Optimized
Final SPE Results after BNP, Enfuvirtide and Somatostatin Methods Optimized 120 100 % SPE Recovery 80 Screening Protocol 60 Oasis® MAX 40 Screening Protocol Oasis® WCX 20 Modified Protocol 0 Minor, compound specific, modifications for 3 peptides result in excellent recovery for all peptides % Matrix
Slide 12: When the Basic Methods Don't Work: Typical Challenges Faced
When the Basic Methods Don’t Work: Typical Challenges Faced Peptide Loss Poor Sensitivity Solubility Specificity/Selectivity Chromatography Protein Binding Non-specific Binding Low Extraction Recovery Inadequate Retention Chemical Instability ©2012 Waters Corporation 12
Slide 13: Outline
Outline Introduction — Basic screening methods — Typical challenges faced Understanding your peptide o pI, HPLC index, hydrophobic/hydrophilic Understanding the challenges: symptoms, examples, and solutions for common problems Conclusions ©2012 Waters Corporation 13
Slide 14: Understanding Your Peptide
Understanding Your Peptide 1. Obtain the peptide sequence 2. Calculate pI and HPLC index 3. Assess nature by evaluating the residues 4. Look for any potential stability problems ©2012 Waters Corporation 14
Slide 15: Chemical Properties of Therapeutic Peptides
Chemical Properties of Therapeutic Peptides Peptide MW pI # of Residues HPLC Index* Octreotide 1019 9.3 8 40.8 Angiotensin II 1046 7.35 8 38.3 Desmopressin 1069 8.6 9 16.8 Vasopressin 1084 9.1 9 7.6 Goserelin 1270 7.3 10 31.7 Angiotensin I 1296 7.51 10 56.2 Somatostatin 1638 10.4 14 52.6 Neurotensin
Slide 16: Understanding Your Peptide: Step 1
Understanding Your Peptide: Step 1 Teriparatide (1-34 fragment of human parathyroid hormone) example: Obtain sequence 1. ©2012 Waters Corporation 16
Slide 17: Understanding Your Peptide: Step 2
Understanding Your Peptide: Step 2 Teriparatide (1-34 fragment of human parathyroid hormone) example: Calculate pI and HPLC index 2. Teriparatide is basic and sticky ©2012 Waters Corporation 17
Slide 18: Understanding Your Peptide: Step 3
Understanding Your Peptide: Step 3 Teriparatide (1-34 fragment of human parathyroid hormone) example: Assess nature by evaluating the residues 3. Teriparatide sequence: SVSEIQLMHNLGKHLNSMERVEWLRKKLQDVHNF Property Amino Acid Hydrophobic A, F, I, L, M, P, V, W, Y Moderate C, G Hydrophilic D, E, H, K
Slide 19: Understanding Your Peptide: Step 4
Understanding Your Peptide: Step 4 Teriparatide (1-34 fragment of human parathyroid hormone) example: Look for any potential stability problems 4. Teriparatide sequence: SVSEIQLMHNLGKHLNSMERVEWLRKKLQDVHNF Property Amino Acid Hydrophobic A, F, I, L, M, P, V, W, Y Moderate C, G Hydrophilic D, E, H,
Slide 20: How Can I Use This Information?
How Can I Use This Information? pI — Helps to determine solubility — Helps identify best SPE and chromatographic conditions/solvents HPLC Index — Indicates if polar or more hydrophobic o Can guide troubleshooting • Hydrophobic: may lead to non-specific binding, protein binding, carryover, low SP
Slide 21: Outline
Outline Introduction — Basic screening methods — Typical challenges faced Understanding your peptide o pI, HPLC index, hydrophobic/hydrophilic Understanding the challenges: symptoms, examples, and solutions for common problems — Solubility — Poor Sensitivity — Specificity — Chromatography and
Slide 22: Specific Challenges in Bioanalysis of Peptides: Solubility
Specific Challenges in Bioanalysis of Peptides: Solubility When is Solubility Important? – In initial solubilization of powder – During sample preparation o Pre-treatment, wash steps, elution – In mobile phase, at flow rate you are running – In injection solvent Symptoms of inadequate/incomplete
Slide 23: General Guidelines for Solubilizing Powder
General Guidelines for Solubilizing Powder 1. If a peptide is < 5 residues, it will likely dissolve in aqueous solutions unless the sequence is entirely comprised of hydrophobic residues. 2. Peptides containing >25% charged residues and < 25% hydrophobic residues generally dissolve in aqueous soluti
Slide 24: General Guidelines for Solubilizing Powder
General Guidelines for Solubilizing Powder 4. Peptides containing >50% hydrophobic residues may be only slightly soluble or insoluble in aqueous solutions. Hydrophobic peptides are best solubilized in DMSO, DMF, strong acid solutions (TFA, formic, acetic), or isopropanol. For cysteine- containing pe
Slide 25: Teriparatide Solubilization Solvent Assessment
Teriparatide Solubilization Solvent Assessment ) MRM of 4 Channels ES+ 100 0.88 824.32 > 983.78 (Teriparatide 5+) 47318 6.41e5 Area % DMSO 0 0.40 0.50 0.60 0.70 0.80 0.90 1.00 1.10 1.20 1.30 1.40 MRM of 4 Channels ES+ 100 0.86 824.32 > 983.78 (Teriparatide 5+) 47286 6.44e5 Area 1% TFA, 50% ACN % 0 0
Slide 26: Protein Precipitation and Solubility of Large Peptides
Protein Precipitation and Solubility of Large Peptides % Teriparatide Recovery 120 100 80 60 40 20 0 1:1 ACN 2:1 ACN 1:1, 1% FA 1:1, 1% 1:1, 1% AA 1:1, 5% 2:1, 5% in ACN TFA in ACN in ACN NH4OH NH4OH ACN ACN Too much organic causes peptide loss, pH impacts recovery ©2012 Waters Corporation 26
Slide 27: Improving Solubility in Mobile Phase B, Reducing Carryover
Improving Solubility in Mobile Phase B, Reducing Carryover 07-Jun-2012 MRM of 2 Channels ES+ 824.351 > 983.774 (Teriparatide) 100 3.49e5 Area 1.05 9940 % Mobile phase B= 0.1% FA in ACN 0 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 MRM of 2 Channels ES+ 1.00 824.351 > 983.77
Slide 28: Specific Challenges in Bioanalysis of Peptides: Specificity
Specific Challenges in Bioanalysis of Peptides Specificity Possible Causes — Many peptides, high abundance proteins in sample — Interference by similar endogenous compounds — Biomarker Analysis o Difficult to get blank matrix for use in method development/validation • Use of stripped or surrogate
Slide 29: Specificity in Human Plasma: Impact of Pretreatment Prior to SPE
Specificity in Human Plasma: Impact of Pretreatment Prior to SPE Plasma diluted with 10mM TRIS Base 1: MRM of 4 Channels ES+ 4.48 TIC 100 % 2.86 4.59 3.95 5.50 6.30 8.21 7.12 0 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 1: MRM of 4 Channels ES+ 100 Plasma diluted with 1% TFA 5.76 TIC 4.46 % 4.56 8
Slide 30: Lack of Specificity in an MRM assay for Trastuzumab
Lack of Specificity in an MRM assay for Trastuzumab 1 nM Trastuzumab in ~1500 area counts Solvent A human serum digest 1 nM Trastuzumab in serum ~500-700 area counts digest = 2-3X lower! Addressing the problems……… ©2012 Waters Corporation 30
Slide 31: Mixed-mode Cation Exchange SPE Clean-up for Protein Digests
Mixed-mode Cation Exchange SPE Clean-up for Protein Digests Before SPE Increase in Signal Reduction in Background Removal of Digest Reagents After SPE Reduction of Plasma Phospholipids Improved Specificity Digest Concentration ©2012 Waters Corporation 31
Slide 32: MS Specificity: Avoiding Immonium Ion Fragments
MS Specificity: Avoiding Immonium Ion Fragments 02-Mar-2012 1: MRM of 4 Channels ES+ 0.99 867 > 136 (Lantus) 100 9.40e5 1.68 867 -> 136 (tyrosine immonium ion) % 1.58 1.15 1.20 1.30 1.40 1.49 Lack of Specificity 0.56 0 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3
Slide 33: MS Specificity: Use of Higher m/z Precursors and Fragments
MS: Specificity: Use of Higher m/z Precursors and Fragments Human Plasma Extract MRM of 2 Channels ES- 2.48 1031.5 > 1026.8 3.20e4 4th charge state 1031.5 -> 1026.8 % 3.36 0 0.50 1.00 1.50 2.00 2.50 3.00 3.50 MRM of 2 Channels ES- 2.48 825 > 821.3 3.26e4 1.67 1.22 1.63 5th charge state % 1.96 2.49 8
Slide 34: Specific Challenges: Protein Binding
Specific Challenges in Bioanalysis of Peptides Protein Binding Symptoms – Decreases solid phase extraction recovery o Peptide passes through on load step with associated protein – Results in underestimation of peptide concentration Possible Causes – More pronounced in stickier/more hydrophobic p
Slide 35: Specific Challenges: Non-specific Binding (NSB)
Specific Challenges in Bioanalysis of Peptides Non-specific Binding (NSB) Symptoms – Peptide loss – Loss of low end of standard curve – Non-linearity – Irreproducible chromatography – Loss of chromatographic peak – Poor LC or SPE recovery Possible Solution – Requires carrier protein or surfactan
Slide 36: Insulin Analogs: Testing for and Eliminating Non-Specific Binding
Insulin Analogs: Testing for and Eliminating Non-Specific Binding Insulin Glargine solution: <30 min. on benchtop 1: MRM of 4 Channels ES+ 1.01 1011.2 > 1179 (Lantus) 100 469 Area 30% MeOH, 10% acetic acid, 0.05% rat plasma % 0 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.
Slide 37: Carrier Protein Options
Carrier Protein Options MRM of 4 Channels ES+ 1.06 824.25 > 983.795 (Teriparatide 5+) 100 5.57e5 Teriparatide + 40ug/ml Bovine Serum Albumin % 0 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 MRM of 4 Channels ES+ 100 824.25 > 983.795 (Teriparatide 5+) 5.57e5 1.
Slide 38: Specific Challenges: Peptide Loss and Poor Sensitivity
Specific Challenges in Bioanalysis of Peptides Peptide loss — Potential losses during evaporation — Hydrophobic peptides stick to vials/collection plates o Concentration dependent — Addition of organic or acid helps maintain solubility for storage — Non-specific binding — Maintain solubility acros
Slide 39: Specific Challenges: Chromatography
Specific Challenges in Bioanalysis of Peptides: Chromatography Problem: Poor peak shape — Possible Solutions o Increase temperature o Decrease flow rate o Increase stationary phase pore size o Change ligand or column chemistry o Add carrier protein to reduce NSB Problem: Poor Sensitivity — Possi
Slide 40: Improving Chromatography for Amyloid Beta Peptide 1-40: Temperature
Improving Chromatography for Amyloid Beta Peptide 1-40, MW 4330: Temperature 52% increase in area counts at higher temperature, carryover also reduced MRM of 1 Channel ES+ 1.17 100 365 1.32e4 Area 365 area counts 60 C % 0 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50 2.75 3.00 3.25 MRM of 1 Chan
Slide 41: Improving Chromatography for Teriparatide: Column Chemistry
Improving Chromatography for Teriparatide: Column Chemistry 0 MRM of 4 Channels ES+ 1.23 824.32 > 983.78 (Teriparatide 5+) 100 ACQUITY BEH C18 300A % 1.59 1.67 0 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 MRM of 4 Channels ES+ 0.78 824.32 > 983.78 (Teriparat
Slide 42: Improving Chromatography: Ligand and Pore Size
Improving Chromatography: Ligand and Pore Size Enfuvirtide: MW 4492 ACQUITY BEH 1.7 µm C18 130Å ACQUITY BEH 1.7 µm C18 300Å ACQUITY BEH 1.7 µm C4 300Å ©2012 Waters Corporation 42
Slide 43: Chromatography: Use of Carrier Protein to Reduce NSB
Chromatography: Use of Carrier Protein to Reduce NSB Teriparatide 18-May-2012 14:41:24 extracted soln 1ng rep 1 MRM of 2 Channels ES+ 0.95 TIC 100 2.56e5 Pure solution in 30% ACN, 1% TFA % 0 0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00 1.10 1.20 1.30 1.40 1.50 1.60 1.70 1.80 1.90 2.00 2.10 2.20
Slide 44: Poor Peak Shape or No Peak: Column Conditioning
Poor Peak Shape or No Peak: Column Conditioning Insulin Glargine MRM of 10 Channels ES+ 866.8 > 984 (Lantus) 100 2.85e6 After 9 injections of % precipitated plasma 0 0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50 2.75 3.00 3.25 MRM of 10 Channels ES+ 100 866.8 > 984 (Lantus) 2.85e6 2nd injec
Slide 45: Chromatographic Carryover
Chromatographic Carryover Problem: Carryover Possible Causes of carryover — Improper injection solvent — Stuck on column — Stuck to LC tubing, needle etc. Possible Solutions — Stuck on column: increase temperature, decrease flow rate, change mobile phase modifier, chromatographic pore size, CSH
Slide 46: Improving Chromatography for Amyloid Beta Peptide 1-40: Flow Rate
Improving Chromatography for Amyloid Beta Peptide 1-40, MW 4330: Flow Rate 48% increase in area counts at lower flow rate, carryover also reduced MRM of 1 Channel ES+ 100 385 1.32e4 Area 400 µL/min % 0 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 MRM of 1 Channel ES+ 10
Slide 47: Teriparatide Chromatography: Improving Diffusion/Solubility
Teriparatide Chromatography: Improving Diffusion/Solubility Column temp 60C Flow rate 0.4ml/min MRM of 4 Channels ES+ 1.08 824.25 > 983.795 (Teriparatide 5+) 100 22737 2.98e5 Area 15-50% B in 2 min 72% Increase in area counts % 0 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80
Slide 48: Low SPE Recovery
Low SPE Recovery Problem: Low SPE Recovery Possible Causes of Low Recovery – Extraction Device Capacity – Poor Retention – Chemical Stability o Apparent low recovery – Non-specific binding – Protein binding – Not strong enough elution solvent or enough volume ©2012 Waters Corporation 48
Slide 49: Low SPE Recovery
Low SPE Recovery Possible Solutions for Low Recovery – Extraction Device Capacity o Load less sample, increase sorbent bed mass o Change pre-treatment to eliminate interferences – Poor Retention o Use appropriate mixed-mode sorbent and ensure charge through proper pretreatment o Eliminate protein
Slide 50: Troubleshooting Recovery
Troubleshooting Recovery 1ngmL_5050_N4OH_020810_001 MRM of 3 Channels ES- SPE Recovery = 94% 5.69 TIC 4.23e4 5.42 5.44 SPE Recovery = 92% 5.44 3 peptides of the same class, extracted from human CSF 5.45 5.71 SPE Recovery = 64% Elution solvent contains 65% ACN 5.67 and 5% NH4OH 5.88 % 5.40 5.89 5.39
Slide 51: Sample Extraction: Final Method
Sample Extraction: Final Method Oasis® MCX μElution Plate Amyloid β % SPE Peptide Recovery Condition, Equilibrate 1-38 94% Load: 150 μL diluted, pretreated sample 1-40 92% 1-42 90% Wash 1 and 2 Elute: 2 X 25 µL 75/10/15 ACN/conc. NH4OH/water (by volume) Fully solubilizes aβ 1-42 Dilute: 25 µL water
Slide 52: Troubleshooting Recovery
Troubleshooting Recovery Recovery is 94% in human CSF but 40% in human plasma What is the problem? — Most likely to be protein binding What can we do? Modify pretreatment — Dilute with stronger acid — Dilute with base — Denature with guanidine HCl (typically 5M, 9:1 guanidine:plasma) ©2012 W
Slide 53: Amyloid Peptides: Troubleshooting Non-specific Binding in SPE Recovery
Amyloid Peptides: Troubleshooting Non- specific Binding in SPE Recovery Solvent standards used for initial SPE method development % SPE Recovery 100 — SPE recovery = 60% 90 80 70 Amyloid Beta Peptides 60 50 — >4000 MW 40 — > 50% hydrophobic residues 30 20 10 0 Mass Balance Results Solvent Std
Slide 54: Outline
Outline Introduction — Basic screening methods — Typical challenges faced Understanding your peptide o pI, HPLC index, hydrophobic/hydrophilic Understanding the challenges: symptoms, examples, and solutions for common problems Conclusions ©2012 Waters Corporation 54
Slide 55: Conclusions
Conclusions Using predefined guidelines to learn about your peptide can help predict potential problems and can guide you towards solutions. Many problems stem from 5 major causes: protein binding, non-specific binding, solubility, specificity and stability. If you are aware of what the challe
Slide 56: Acknowledgements
Acknowledgements Waters Corporation ICON PLC Erin Chambers Jon Bardsley Debadeep Battacharya Eileen Collins Catalin Doneanu Sally Hannam Kenneth Fountain Liz Thomas Gordon Fujimoto Joanne Mather Paul Rainville Martha Stapels Hua Yang ©2012 Waters Corporation 56
Slide 57: Thank You! Questions?
Thank You! Questions? Landing Page… http://www.waters.com/May1 – Promotional Offers on Products for Peptide Separation and Sample Preparation – Full Webinar Recording of Today’s Session – PDF Slide Deck – Compilation of Literature, White Papers, Brochures, Application Notes General Questions