Slide 1: pNeuton Ventilator Model A-E Self-Guided Tour
pNeuton Ventilator ® Model A-E Self-Guided Tour The ventilator has not been cleared or approved by the FDA. The ventilator has been authorized by FDA under an Emergency Use Authorization (EUA). The ventilator is authorized only for the duration of the U.S. pNeuton is a registered trademark of the Ai
Slide 2: pNeuton Ventilator Model A-E
pNeuton Ventilator Model A-E ® Manufactured by Ford Motor Company for GE Healthcare 3000 N. Grandview Blvd. Waukesha, WI 53188 USA Assembled in USA Distributed by GE Healthcare. For Customer Service, call GE Healthcare at (800) 345-2700. Do not contact Airon. Direct all inquiries to GE Healthcare. T
Slide 3: pNeuton Ventilator
pNeuton Ventilator ® The pNeuton Model A-E Ventilator provides critical care ventilation and CPAP • Invasive ventilation (ET-Tube) – Controlled Ventilation with PEEP – Intermittent Mandatory Ventilation with PEEP – CPAP • Non-invasive ventilation (Mask) – CPAP JB79540US 3
Slide 4: Objectives
Objectives This program is a self-guided tour of the pNeuton Ventilator Model A-E. At the end of this tour you will be able to: • Describe the main operating features of the ventilator • Identify the controls and connections • Describe the patient breathing circuit • Set-up the ventilator for patien
Slide 5: Features and Controls – Rear Panel: Driving Gas Input
Features and Controls – Rear Panel Driving Gas Input • Connection for 55 psi (3.8 bar) oxygen source gas • Always use a high flow gas source that can provide 55 ± 15 psi pressure at a minimum of 40 L/min JB79540US 5
Slide 6: Features and Controls – Rear Panel: Low Driving Gas Alarm
Features and Controls – Rear Panel Alarm, low driving gas • Audible alarm sound generator • Factory preset alarm WARNING: • Do NOT put tape over the holes, push sharp objects into the alarm or, interfere, in any way, with the alarm. • The alarm is for patient protection and, if, for any reason, you
Slide 7: Features and Controls – Rear Panel: Air Inlet Filter
Features and Controls – Rear Panel Ambient air inlet filter • Ambient air is entrained for internal gas mixing • External housing contains air filter • Air filter replacement is part of the preventative maintenance JB79540US 7
Slide 8: Features and Controls – Front Panel: PEEP/CPAP Control
Features and Controls – Front Panel PEEP/CPAP control • Adjusts the PEEP/CPAP from 0 to 20 cm H2O • Automatic sensitivity • Flow adjusts automatically to meet patient demand up to 140 L/min • Oxygen concentration is maintained at set % Oxygen control JB79540US 8
Slide 9: Features and Controls – Front Panel: Peak Pressure Control
Features and Controls – Front Panel Peak Pressure control • Limits peak airway pressure from 15 to 75 cm H2O for mandatory breaths • Alert with “honking” noise when delivered pressure rapidly exceeds set pressure (e.g. cough) JB79540US 9
Slide 10: Features and Controls – Front Panel: Tidal Volume Control
Features and Controls – Front Panel Tidal Volume control • Sets delivered Tidal Volume for mandatory breaths from 360 to 1,500 ml • Calibrated control for volume delivery • Delivered flow rate equals 600 ml/sec (36 L/min) for all mandatory breaths • Operating range sets inspiratory time of 0.6 to 2.
Slide 11: Features and Controls – Front Panel: Respiratory Rate Control
Features and Controls – Front Panel Respiratory Rate control • Sets mandatory breath Respiratory Rate from 3 to >28 breaths/min • Control adjusts expiratory time with a range of 0.6 to 20 seconds • Calibrated control is accurate for tidal volumes from 500 to 950 ml. Lower volumes will have faster ra
Slide 12: Features and Controls – Front Panel: Visual Alarm Indicator
Features and Controls – Front Panel Visual alarm indicator • Illuminates for an active alarm condition • Alarm warns of patient disconnect after approximately 22 seconds • Alarms for low gas supply • Will turn off when alarm is corrected or Reset/Silence button is pressed JB79540US 12
Slide 13: Features and Controls – Front Panel: Alarm Reset/Silence
Features and Controls – Front Panel Alarm Reset/Silence button • Press button to silence the alarm system for 1 minute • Every time the button is pressed the alarm silence period will reset for 1 minute JB79540US 13
Slide 14: Features and Controls – Front Panel: % Oxygen Control
Features and Controls – Front Panel % Oxygen control • Delivered oxygen choice – 65% – 100% • Sets both mandatory breath and spontaneous breath oxygen concentration JB79540US 14
Slide 15: Features and Controls – Front Panel: Mandatory Breath Control
Features and Controls – Front Panel Mandatory Breath control • Activates ventilation mode • Switch for mandatory breaths – On or Off • Ventilator operates in CPAP mode with switch in “Off” position JB79540US 15
Slide 16: Features and Controls – Front Panel: Pressure Gauge
Features and Controls – Front Panel Pressure gauge • Represents pressure in the patient circuit at the output of the ventilator JB79540US 16
Slide 17: Features and Controls – Front Panel: Patient Connection
Features and Controls – Front Panel Patient Connection • Connection for the large bore hose of the patient circuit JB79540US 17
Slide 18: Features and Controls – Front Panel: Expiratory Valve
Features and Controls – Front Panel Expiratory Valve • Patient circuit small hose connection • Connects to the expiratory valve on the patient circuit JB79540US 18
Slide 19: Patient Breathing Circuit
Patient Breathing Circuit Patient circuit is a single limb system • Connect large hose end to Patient Connection port and small hose end to the Exp Valve port on the front of the ventilator • Patient end connects for clinical intervention – Directly to an endotracheal tube (invasive) – Face mask (no
Slide 20: Patient Breathing Circuit: Use Only Recommended Circuits
Patient Breathing Circuit Use ONLY the recommended patient circuits • Calibration of the PEEP/CPAP and Peak Pressure controls are based upon the recommended patient circuit • Using other manufacturers circuits may cause: – Errors in pressure delivery – High expiratory flow resistance – Interference
Slide 21: Patient Breathing Circuit: Accessories
Patient Breathing Circuit Accessories to optimize patient care • CO2 monitoring – For In-line sampling place the tee between the mask and patient circuit – Or use the nasal prongs under the mask • Use the expiratory bacteria/viral filter provided in the patient circuit kit. This can be added using t
Slide 22: Patient Application: Initial Set-up
Patient Application Initial set-up 1 2 3 Attach the recommended Attach a high pressure oxygen Attach the other end of patient circuit to ventilator hose to the rear of the ventilator the high pressure hose to the oxygen source JB79540US 22
Slide 23: Patient Application: Set Main Controls
Patient Application Set main controls Determine Peak Pressure maximum PEEP/CPAP can be set to minimum Choose % Oxygen Set Tidal Volume Set Respiratory Rate JB79540US 23
Slide 24: Patient Application: Begin Ventilation
Patient Application Begin ventilation Alarm system will 2 Press Alarm Silence 1 Turn “On” oxygen supply immediately activate button, if desired For ventilation, turn 3 Mandatory Breaths control “On” Ventilator will start to provide breaths at the settings on the controls 24 JB79540US
Slide 25: Patient Application: Set PEEP/CPAP
Patient Application Set PEEP/CPAP Adjust the PEEP/CPAP control to the level desired If a test lung or patient is not attached to the circuit, a high flow of gas will occur With Mandatory Breaths control “Off”, set at least 4 cm H2O to optimize performance in CPAP mode 25 JB79540US
Slide 26: Patient Application: Tidal Volume and Respiratory Rate Interchange
Patient Application Tidal Volume and Respiratory Rate interchange • Tidal Volume control is calibrated and sets the inspiratory time • Respiratory Rate control sets the expiratory time and is calibrated for tidal volumes between 500 and 900 ml • Set desired Tidal Volume first, then adjust the rate a
Slide 27: Ventilator Operation: Alarm System Functionality
Ventilator Operation Alarm system functionality • Alarm preset to activate within approx. 22 seconds • Disconnect alarm is looking for a specific minimum pressure – Mandatory Breaths “On”, breath must generate a minimum 15 cm H2O – Mandatory Breaths “Off”, minimum CPAP level of 4 cm H2O is needed •
Slide 28: Ventilator Operation: Peak Pressure Limit
Ventilator Operation Peak Pressure limit • Operates to limit the peak airway pressure during mandatory breaths • For volume ventilation, set the Peak Pressure 10 cm H2O above the pressure needed to deliver the set Tidal Volume • Peak Pressure control is not an alarm – During rapid patient exhalation
Slide 29: Ventilator Operation: % Oxygen Control
Ventilator Operation % Oxygen control • Set % Oxygen control to the desired oxygen level • Both mandatory breaths and spontaneous breaths receive set oxygen level • Oxygen tank time almost doubles with 65% setting • Caution – with rapid, large tidal volumes on 100% oxygen, the actual delivered perce
Slide 30: Clinical Issues
Clinical Issues Let’s look at various clinical environments for ventilator/CPAP application Pediatric ventilation Medication nebulization CO2 monitoring Altitude & Clinical operation JB79540US 30
Slide 31: Clinical Issues: Pediatric Ventilation
Clinical Issues Pediatric ventilation • The tidal volume setting of 360 to 1,500 ml is used on patients from 23 kg (50 pounds) and higher • pNeuton is not approved for neonatal or small children due to mandatory breath flow of 36 L/min • Monitor pediatric patients closely to ensure ventilation synch
Slide 32: Clinical Issues: Providing Nebulized Medications
Clinical Issues Providing nebulized medications • A medication nebulizer may be placed between the patient connection and the patient • Patient will inhale medication during every breath and exhale back out through patient circuit • Expiratory filter may be added for personnel safety JB79540US 32
Slide 33: Clinical Issues: CO2 Monitoring
Clinical Issues CO2 monitoring • Any type of CO2 analyzer may be used • With in-line style analyzers place the sensor tee between the patient circuit connection and the patient JB79540US 33
Slide 34: Clinical Issues: Altitude & Clinical Operation
Clinical Issues Altitude & Clinical operation • Ventilator used for air transports in unpressurized cabins up to 15,000 feet • Delivered Tidal Volume will increase and Respiratory Rate will decrease as altitude increases • Pressure settings will not change as altitude changes • Always use an externa
Slide 35: Patient Safety
Patient Safety pNeuton Model A-E: Patient safety features • Audible and visual alarms are for patient disconnect and gas supply pressure below 30 psi (2 bar) • High Pressure Release – pressure is relieved at the Peak Pressure control setting • Internal Safety Pressure Release – limits circuit pressu
Slide 36: Cleaning and Maintenance
Cleaning and Maintenance • Clean ventilator between patient use applications using a soft cloth and a mild detergent/disinfectant • Do not immerse the ventilator in water • Disposable patient circuit is for single use application and should not be cleaned, disinfected or reused JB79540US 36
Slide 37: Conclusion
Conclusion This concludes the pNeuton Ventilator Model A-E self-guided presentation In this course, you learned about: • The main operating features of the ventilator • The pNeuton Ventilator Model A-E controls and connections • The patient breathing circuit • The ventilator set-up for patient opera
Slide 38: pNeuton Ventilator Model A-E: Disclaimer
pNeuton Ventilator Model A-E: Disclaimer ® The ventilator has not been cleared or approved by the FDA. The ventilator has been authorized by FDA under an Emergency Use Authorization (EUA). The ventilator is authorized only for the duration of the U.S. declared health emergency. The design of the dev