  1. Slide 1: Codman Hakim Programmable Valve (CHPV)

    Codman Hakim Programmable Valve (CHPV) ® Company Confidential – Not for Distribution 1

  2. Slide 2: CODMAN Neuro Programmable Valve History

    CODMAN Neuro Programmable Valve History 2013 SIPHONGUARD CODMAN ® Device CERTAS PLUS™ 2002 2007 Programmable Valve CODMAN HAKIM 2006 Silicone Programmable Valve BACTISEAL Stator fit housing 1992 1998 Improvement redesign 1999 2003 2007 Second Cam redesign VPV Generation Programmer Company Confidenti

  3. Slide 3: Adjustable Valves

    Adjustable Valves • Valve Mechanism allows the change operating pressure noninvasively • Codman Valve – Codman Programmable Valve – 18 pressure settings Company Confidential – Not for Distribution 3

  4. Slide 4: How does CHPV work?

    4 How does CHPV work ? Spring Spiral Cam Ruby Ball & Seat Base Plate CODMAN HAKIM Programmable Valve (CHPV) Company Confidential – Not for Distribution 4

  5. Slide 5: The CHPV technology

    5 The CHPV technology Codman uses a cantilever spring design in concert with a perfectly matched ruby ball and valve seat. • Maintains a precise constant opening pressure over years of service. • Silicone valve performance degrades over time. • Tight operating pressure tolerance. Company Confidentia

  6. Slide 6: The key features of CHPV

    The key features of CHPV • Eighteen (18) Distinct Pressure Settings • 30 to 200 mm H20 • Precise Pressure Bandwidths • 8 configurations • Available unitized with anti siphon device: SIPHONGUARD Device Company Confidential – Not for Distribution 6 6

  7. Slide 7: The key features of CHPV

    The key features of CHPV • MR conditional up to 3 Tesla • Smallest MRI artifact in the market. Max signal void area = 2439mm² • Minimal magnetic influence (min 80 Gauss): electronic devices such as Iphone and Ipad can be safely used by patients when used as intended • Shunt placement site does not a

  8. Slide 8: The CHPV: The most precise valve in the market

    8 The CHPV: The most precise valve in the market 225 200 • No siphoning effect 175 145 150 • 18 settings that range from 30 -200mmH2O. 115 125 • Enables the physicians to make small and 100 accurate adjustments 80 of 10mmH2O 75 Increments throughout the treatment course 40 50 of their patients. 20 2

  9. Slide 9: SiphonGuard Siphon / Flow Control

    SiphonGuard® Siphon / Flow Control A unique device designed to reduce the risk of CSF overdrainage complications. • Dual-pathway system – Prevents excessive flow while allowing constant drainage • Stable mechanical system – Remains reliably sensitive to differences between normal and excessive flow

  10. Slide 10: Different valve configurations

    10 Different valve configurations CHPV with SG Certas Plus + with SG SG standalone Company Confidential – Not for Distribution 10

  11. Slide 11: SIPHONGUARD: Dual Pathways

    SIPHONGUARD: Dual Pathways • The primary pathway remains open at physiological flow rates ≤40 ml/hr (twice the normal average CSF production rate in adults) • The primary pathway closes at a flow rate of about 100 ml/hr and CSF then flows through (and only through) the secondary pathway, at 10% of t

  12. Slide 12: SiphonGuard - Simplified

    SiphonGuard - Simplified TM Primary Pathway Secondary Pathway Proximal Company Confidential – Not for Distribution 12

  13. Slide 13: X-Ray Technique and X-Ray Overlay Tool

    X-Ray Technique and X-Ray Overlay Tool Valve / Plate Orientation Company Confidential – Not for Distribution 13 13

  14. Slide 14: X-Ray Valve Information

    X-Ray Valve Information • The white marker on the valve (1) indicates the right hand side of the valve – The marker will not be seen on the x-ray if it is positioned on a 30° angle or more; the valve cannot be programmed if the angle is 45° or more • The pressure indicator on the white ring (2) deno

  15. Slide 15: Codman VPV

    Codman VPV • Overview • Tips for use Company Confidential – Not for Distribution 15

  16. Slide 16: The CHPV programming

    16 The CHPV programming • Not painful • Accurate • Adaptability to a variety of clinical situations Company Confidential – Not for Distribution Data Source: Codman Data on File 16

  17. Slide 17: Company Confidential – Not for Distribution

    Company Confidential – Not for Distribution 17

  18. Slide 18: VPV

    VPV • Sensitive Sonar technology • Listens to the valve • Counts the steps • Confirms the adjustment Company Confidential – Not for Distribution 18

  19. Slide 19: A new programmer

    A new programmer Transmitter head A. Start button B. Direction of flow arrow C. Cord D. Central rod (Acoustic device) E. Feet Company Confidential – Not for Distribution 19

  20. Slide 20: Technique tips

    Technique tips • Be aware of noise levels • Transmitter head positioning Programming Zone: Equal to existing programmer Listening Zone: Center coupling rod over cam Company Confidential – Not for Distribution 20

  21. Slide 21: Technique tips

    Technique tips • A pea-sized amount of gel on the central rod • Might also be applied on the patient’s head over the valve mechanism. • Avoid gel contact with programmers’ probes. Programming Zone: Equal to existing programmer Listening Zone: Center coupling rod over cam Company Confidential – Not f

  22. Slide 22: A new programmer

    A new programmer Two modes • Packaged valve • Implanted valve Company Confidential – Not for Distribution 22

  23. Slide 23: A new programmer

    A new programmer Packaged valve mode: allows the adjustment of the valve before the implantation, through the sterile blister. When in this modality, the acoustic sensor is de-activated, and the programmer is equal to model 82-3190R. Implanted valve mode: allows the adjustment of the valve noninvasi

  24. Slide 24: VPV

    VPV • Confirms adjustment – It needs to count the clicks • Compatible with all CHPV • Avoids “false positives” Company Confidential – Not for Distribution 24

  25. Slide 25: VPV system

    VPV system Is the machine in working order? • Previous “Adjustment complete” message, and no system-error messages indicates it is OK • Defect would give a system-error message or would mean you would never get an “adjustment complete” message Company Confidential – Not for Distribution 25

  26. Slide 26: VPV System

    VPV System • Quick Check – centre rod on the transmitter head moves in and out with GENTLE pressure. If not, then it may have been damaged recently. Company Confidential – Not for Distribution 26

  27. Slide 27: VPV performance

    VPV performance • Circumstances that could affect VPV performance – Position of the valve • Close to ear = difficult to get VPV transmitter parallel to valve base • Solution = attempt to keep parallel orientation even if it means only some of the transmitter feet contact the patient • Transmitter he

  28. Slide 28: VPV performance

    VPV performance • Contour of the skull – Esp cranial defects can alter the orientation of the transmitter – Need the transmitter parallel rather than rely 100% on the feet to produce the correct parallel orientation – If the valve is prominent, avoid positioning the head so that the center rod fully

  29. Slide 29: VPV performance

    VPV performance • Skin thickness and scarring – In testing skin thickness was not an issue – Scar tissue can be a problem • Abundance of hair – Better to flatten and part the hair – The closer the rod is to contacting the skin the better – If hair causes air gap between center rod and skin, fill the

  30. Slide 30: VPV performance

    VPV performance • Noises at time of adjustment – If patient coughs, talks, snores, moves or causes a vibration during the adjustment this will be interpreted as an unexpected event – “repeat adjustment” – If noise occurs, then repeat with elimination of noise Company Confidential – Not for Distribut

  31. Slide 31: Note:

    Note: • VPV is more sensitive to transmitter/valve position. Positioning the centre rod directly over the cam mechanism should be the main priority. • This can be facilitated by marking the skin above the cam with skin marker Company Confidential – Not for Distribution 31

  32. Slide 32: Users Tips

    Users Tips • Arrow on transmitter head must be in line with axis of flow • Centre rod must be above the CHPV ideally above the moveable cam • Transmitter head should be parallel to the valve- valve not rotated on the axis of flow Company Confidential – Not for Distribution 32

  33. Slide 33: Users Tips

    Users Tips • Properly position the center rod on the valve mechanism – Possibly because • Previous success with 82-3190 with familiar method of positioning transmitter head relative to valve ( same experienced users now get “repeat adjustment” message sometimes because too proximal or distal to cam

  34. Slide 34: Manufacturing

    Manufacturing Le Locle, Switzerland Clean Room Company Confidential – Not for Distribution 34

  35. Slide 35: Indications and contact information

    Indications The Codman® Hakim® Programmable Valves are implantable devices that provide constant intraventricular pressure and drainage of CSF for the management of hydrocephalus. Contraindications The Codman® Hakim® Programmable Unitized Valve Systems are not recommended for atrial placement. Use t