  1. Slide 1: FULCRUM-VT Investor Event

    FULCRUM-VT Investor Event Dr. Atul Verma • Dr. William Stevenson • Dr. Matthew Hakimi April 27, 2026 1

  2. Slide 2: Disclaimer: Forward Looking Statements

    Disclaimer Forward Looking Statements This presentation contains forward-looking statements that involve substantial risks and uncertainties. All statements, other than statements of historical facts, contained in this presentation, including statements regarding the potential of our Proprietary Ult

  3. Slide 3: Effectiveness and Safety of Ultra-Low Temperature Ablation of VT: Pivotal FULCRUM-VT Outcomes

    EFFECTIVENESS AND SAFETY OF ULTRA-LOW TEMPERATURE ABLATION OF VENTRICULAR TACHYCARDIA IN PATIENTS WITH STRUCTURAL HEART DISEASE: OUTCOMES OF THE PIVOTAL FULCRUM-VT TRIAL Atul Verma, MD, McGill University Health Centre on behalf of FULCRUM-VT investigators THANK YOU TO ALL CO-INVESTIGATORS! April 26,

  4. Slide 4: Disclosures

    DISCLOSURES Dr. Atul Verma Dr. William Stevenson Research Grants Honoria Bayer, Biotronik, Biosense Webster, Abbott, Biotronik, Boston Scientific, Medtronic, Abbott, Cardiofocus Johnson & Johnson, Medtronic Advisory Board Cardiofocus, Adagio Medical, Lilly, Research Support Biosense Webster, Kardium

  5. Slide 5: Motivation

    MOTIVATION • RF lesions – insufficient depth for effective endocardial- only ablations in the ventricle • Safe application of PFA to achieve deeper lesions in the ventricle is an open question • Lack of universal approach (and approved devices) for ablations in both the ICM and NICM patients RF=Radi

  6. Slide 6: ULTA System Design

    ULTA SYSTEM DESIGN PURPOSE-BUILT FOR ENDOCARDIAL VENTRICULAR ABLATIONS ABLATION PROTOCOL (freeze – thaw - freeze cycle) Freeze Duration Lesion Depth (sec) Range (mm)1 15 mm 30 <4 60 4-6 • Console + 9 Fr bi-deflectable catheter 120 7-9 • 15 mm long ablation element, -120 C 180 ≥ 10 • 8 electrodes for

  7. Slide 7: Ultra-Low Temperature Ablation

    ULTRA-LOW TEMPERATURE ABLATION Ablative power to produce large footprint, depth-controlled endocardial lesions ≥10 mm, unaffected by the presence of the scar PRE-CLINICAL LESIONS Dewland T, et al. Ultra-low temperature cryoablation versus ultra-low temperature cryoablation combined with pulsed field

  8. Slide 8: FULCRUM-VT Trial Design

    FULCRUM-VT TRIAL DESIGN Study Design Patient Population Primary Endpoints • Single arm • 20% ≤ LVEF ≤ 50% • Protocol-defined Major Adverse Events (MAEs) 0-7 days • 209 patients • ICM • Acute non-inducibility of VTs targeted for • 19 sites • NICM, incl. ARVC ablation • Endocardial access only • ≤ 1 p

  9. Slide 9: Patient Demographics

    PATIENT DEMOGRAPHICS All patients ITT Cohort* *Intention to Treat (ITT) Cohort excluded Early Feasibility Study Number of Patients 209 157 (EFS) and Roll-in patients Age 68 ± 11 y.o. 69 ± 10 y.o. Male Sex 92.8% 94.3%) BMI 30 ± 6 30 ± 6 Cardiomyopathy ICM 64.1% 65.6% NICM 30.1% 28.0% both ICM and NIC

  10. Slide 10: Procedural Detail

    PROCEDURAL DETAIL ICM NICM* n=103 n=54 Lesions / patient 12 +/- 6 11 +/- 7 ICM 209 Total freeze time/lesion, min 4 +/- 1 4 +/- 1 250 *includes patients with combined NICM-ICM cardiomyopathy 200 109 NO ABLATION-INDUCED VF 150 OR MEANINGFUL EFFECT ON Minutes 55 THE PERFORMANCE OF ICD 54 NICMCANS, PACI

  11. Slide 11: Primary Safety

    PRIMARY SAFETY OTHER PROCEDURE-RELATED • Primary Safety Endpoint: 2.4% (5/209) SERIOUS ADVERSE EVENTS* EVENT DESCRIPTION # EVENTS • 2 (1.0%) device-related events Access site complications 2 • 8% (17/209) rate of non-device, procedure-related Fluid overload (HF) 1 events PEA arrest 1 MAJOR ADVERSE E

  12. Slide 12: Acute Effectiveness

    ACUTE EFFECTIVENESS 98.0% PATIENT-LEVEL ACUTE SUCCESS • Primary Acute Effectiveness: 2 No VT induction after pre-ablation PES 157 patients (296/302) 5 24 patients Post-ablation PES 133 patients not performed 11 patients 122 patients 115 302 Target VTs 94.3% - Full Success 6 296 98.4% - Clinical Succ

  13. Slide 13: Primary Chronic Effectiveness

    PRIMARY CHRONIC EFFECTIVENESS KAPLAN-MEIER FREEDOM FROM ANY VT RECURRENCE OR AAD ESCALATION Cumulative Freedom From Primary Event, % PRIMARY EVENTS, BY TYPE Number of Events n=58 59.0% ICD Therapy (ATP or Shock) 54 ATP Only 32 ICD Shock 22 By Etiology p-value Monitoring Zone VT > 30 sec 3 ICM 58.0%

  14. Slide 14: Effectiveness by Etiology

    EFFECTIVENESS BY ETIOLOGY IDENTICAL EFFECTIVENESS for both ICM and NICM patients K-M FREEDOM FROM ANY ICD THERAPY: 61.8% K-M FREEDOM FROM ICD SHOCK: 84.3% Cumulative Freedom From ICD Therapy, % Cumulative Freedom From ICD Shock, % NICM*: 84.5% NICM*: 63.2% ICM: 84.0% p=0.970 ICM: 60.7% p=0.893 Days

  15. Slide 15: Reduction in VT Burden

    REDUCTION IN VT BURDEN VT BURDEN PRE- AND POST-ABLATION* 75 -75 50 -50 25 -25 00 25 25 50 50 75 75 PATIENTS WITH ICD SHOCK* POST-ABLATION PRE-ABLATION POST-ABLATION MEDIAN (IQR) MEDIAN (IQR) ICM : 81% 3 (1,7) 0 (0,1) NICM : 82% p<0.001 * 85 patients with > 150 days of ICD data in both pre- and post-

  16. Slide 16: Reduction in AAD Use

    REDUCTION IN AAD USE AMIODARONE USE* 100% AT 6 MO 90% 80% 70% 60% 72% •72% of patients are free or on reduced dose of 50% amiodarone 40% 30% 36% •60% 20% 28% 10% relative reduction 0% in amiodarone use Baseline Post-procedure 6 months New Rx Dose Increase Baseline Dose Dose Decrease No Rx * 137 pati

  17. Slide 17: 30-Day Re-admissions

    30-DAY RE-ADMISSIONS Literature FULCRUM-VT Reported1,2 (2/152) VT 7-8% 1.9% (3/154) HF 2-3% 1.3% (2/154) (5/152) All-cause 14-19% 8.4% (13/154) 1Cheung et al. Circ Arrhythm Electrophysiol. 2018;11:e006754 2Tan MC et al. Journal of Interventional Cardiac Electrophysiology (2024) 67:513–521 HF=heart f

  18. Slide 18: FULCRUM-VT Expansion Trial Now Enrolling

    FULCRUM-VT EXPANSION TRIAL NOW ENROLLING FREEZE TIME PER LESION REDUCTION* 2nd vs 1st GENERATION OF ULTA CATHETER 4 Total Freeze Time / Lesion, min FULCRUM-VT AVERAGE 1st Gen 2nd Gen Sheath compatibility 9 Fr 8.5 Fr 3 Shaft stiffness profile Uniform Variable 75% Bi-directional deflection F/F D/F 2 A

  19. Slide 19: Conclusions

    CONCLUSIONS THE FIRST IDE FOR PURPOSE-BUILT VENTRICULAR CATHETER • Endocardial ablation only, one-third non-ischemic CM • 2.4% primary safety event rate • 59% freedom from ANY VT recurrence or AAD escalation • 84% freedom from ICD shock • Equivalent outcomes in ischemic and non-ischemic CM • Median

  20. Slide 20: Q&A

    Q&A - CONFIDENTIAL - 24)