Boston Scientific Corporation · NYSE:BSX

Boston Scientific — Technical Presentation (October 2024)

30 slides from Boston Scientific's technical presentation, October 2024

  1. Slide 1: One-year Outcomes of ACURATE neo2 vs Approved TAVR Devices: ACURATE IDE Trial

    One-year Outcomes of ACURATE neo2 vs Approved TAVR Devices in All-risk Patients with Severe AS: the ACURATE IDE Trial Michael Reardon, MD Houston Methodist DeBakey Heart & Vascular Center Houston, TX On behalf of the ACURATE IDE Investigators

  2. Slide 2: Disclosure of Relevant Financial Relationships

    Disclosure of Relevant Financial Relationships Within the prior 24 months, I have had a relevant financial relationship with a company producing, marketing, selling, re-selling, or distributing healthcare products used by or on patients: Nature of Financial Relationship Ineligible Company Consultant

  3. Slide 3: Background Clinical Experience

    Background Clinical Experience ACURATE neo2 Aortic Valve System is commercially available in 60+ countries (outside US) with 80,000+ patients treated with the ACURATE Platform globally Post-market ACURATE neo2 PMCF Early neo2 Registry 3 PROVE Registry4 study/registry : N=250 N=554 N=1044 30 days1 1

  4. Slide 4: ACURATE neo2 Valve Design

    ACURATE neo2 Valve Design Unique frame design Self-expanding • Self-expanding Nitinol frame with top-down deployment open-cell frame • Open-cell design with axial stabilization arches • Supports unrestricted coronary access1 & predictable Porcine commissure alignment2 pericardium leaflets Supra-annu

  5. Slide 5: ACURATE IDE Trial Design

    ACURATE IDE Trial Design Prospective, multicenter, randomized study N=1500 patients with symptomatic severe native aortic stenosis indicated for TAVR Operators pre-specify valve type to be used if randomized to Control 1:1 Randomization CoreValve Evolut R Evolut Evolut PRO Evolut PRO+ N=244 ACURATE

  6. Slide 6: ACURATE IDE Trial Administration

    ACURATE IDE Trial Administration Principal Investigators Top Enrolling Sites Michael Reardon, MD Raj Makkar, MD Enrollment rank Investigator(s) Clinical Site Core Laboratories 1 Raj R Makkar Cedars-Sinai Medical Center Baim Institute for Clinical Research Eric Gnall Angiography & CT: 2 Lankenau Medi

  7. Slide 7: Key Inclusion Criteria

    Key Inclusion Criteria Severe Symptomatic Aortic Stenosis • AVA ≤ 1.0 cm2 or AVA index ≤ 0.6 cm2/m2 • AND mean gradient ≥ 40 mmHg or max AV velocity ≥ 4.0 m/s or Doppler velocity index ≤0.25 • NYHA Functional Class ≥ II • Heart Team agreement that subject is indicated for TAVR Documented aortic annu

  8. Slide 8: Key Exclusion Criteria

    Key Exclusion Criteria Anatomic Clinical • Unicuspid or bicuspid valve • Acute MI within 1 month • Pre-existing prosthetic aortic or mitral valve • Stroke or TIA within 6 months • Severe (4+) aortic, tricuspid, or mitral • Renal insufficiency (eGFR < 20 ml/min) and/or regurgitation renal replacement

  9. Slide 9: Statistical Methods

    Statistical Methods Non-inferiority of ACURATE neo2 vs Control for the primary Primary hypothesis endpoint of all-cause mortality, stroke or rehospitalization† at 1y • Non-inferiority test performed using Bayesian method; ITT population Expected rate 22.3% for both arms • Based on a weighted average

  10. Slide 10: Trial Enrollment Timeline

    Trial Enrollment Timeline • 1500 RCT subjects enrolled at 70 centers in United States and Canada • Enrollment occurred over a 47-month period, with higher risk patients enrolled earlier in the trial Total RCT Enrollment by Month & by Risk 80 # of RCT Randomizations (Enrollment) COVID (subsequent wav

  11. Slide 11: ACURATE neo2 Implant Metrics

    ACURATE neo2 Implant Metrics • Sites averaged 2.9 months between ACURATE neo2 implants • Total site volume of ACURATE neo2 cases was low  72% of physicians had 5 or fewer cases over the course of the trial  Only 10% of implanters did more than 10 cases ACURATE neo2 Cases per Site 1% 9% 7.4% ≤ 10 c

  12. Slide 12: Subject Disposition

    Subject Disposition 1500 patients enrolled 1:1 Randomization ACURATE neo2 Control • Evolut (n=244) ITT Population (N=752) (N=748) • SAPIEN (n=504) (primary analysis population) No ACURATE neo2 Implanted (N=25) No Control Implanted (N=12) Withdrawal prior to procedure: 18 Withdrawal prior to procedur

  13. Slide 13: Baseline Characteristics and Medical History

    Baseline Characteristics and Medical History ACURATE neo2 Control Evolut SAPIEN (N=752) (N=748) (N=244) (N=504) Age (yrs) 78.0±6.9 (752) 79.0±6.5 (748) 78.8±6.0 (244) 79.1±6.8 (504) Female (%) 52.5% (395/752) 51.2% (383/748) 57.8% (141/244) 48.0% (242/504) STS Score (%) 2.7±1.8 (752) 2.8±1.9 (748) 2

  14. Slide 14: Procedural Characteristics

    Procedural Characteristics ACURATE neo2 Control Evolut SAPIEN P-value (N=244) (N=504) (N=752) (N=748) Correct positioning of a single valve into the proper anatomical location 98.4% (721/733) 99.1% (730/737) 0.243 97.5% (235/241) 99.8% (495/496) Prosthetic aortic valve malpositioning* 1.2% (9/752) 0

  15. Slide 15: Primary Endpoint – Bayesian Analysis

    Primary Endpoint – Bayesian Analysis Hypothesis Test in ITT population PE_ACURATE – PE_Control < Δ non-inferiority For the primary endpoint of all-cause mortality, stroke or rehospitalization† at 1 year Posterior Median and 95% BCI Non-Inferiority Test* Posterior Median Difference Non-inferiority AC

  16. Slide 16: Kaplan-Meier Analysis through 1 Year

    Kaplan-Meier Analysis through 1 Year ITT Population (N=1500) Death ∆1.1% Death, stroke, or rehospitalization† [-1.0%, 3.2%] 5.0% 3.9% ACURATE neo2 Control Stroke ∆2.3% [0.1%, 4.4%] ∆5.8% 5.7% [2.4%, 9.1%] 3.4% 14.8% 9.1% Rehospitalization† ∆1.8% [-0.4%, 3.9%] No. at risk ACURATE neo2 5.3% 752 733 71

  17. Slide 17: Additional Safety Outcomes

    Additional Safety Outcomes Time-to-Event Analysis through 1 Year Post-Procedure, ITT Population (N=1500) ACURATE neo2 Control Hazard Ratio Evolut SAPIEN (N=752) (N=748) [95% CI] (N=244) (N=504) Death 5.0% (36) 3.9% (28) 1.30 [0.80, 2.14] 3.4% (8) 4.1% (20) Cardiovascular death 3.7% (27) 1.8% (13) 2.

  18. Slide 18: Echocardiography Outcomes

    Echocardiography Outcomes Mean Gradient ACURATE neo2 neo2 (n=752) Paravalvular Regurgitation Severe Evolut (n=244) Moderate 16.0 SAPIEN (n=504) Mild 11.6 None/Trace 12.0 12.0 11.4 mmHg 9.0 8.0 Discharge 1 Year 8.0 8.2 7.7 7.9 7.0 0.1% 0.5% 4.0 1.0% 0.8% 3.2% 4.6% 0.2% 6.7% 23.0% 20.4% Discharge 30 D

  19. Slide 19: Review of ACURATE neo2 Performance

    Review of ACURATE neo2 Performance • Despite limited operator experience with ACURATE neo2, early outcomes were encouraging  Periprocedural outcomes were favorable and comparable to Control  Echo parameters (gradient and leak) were as expected  No acute safety signals • COVID-related factors impa

  20. Slide 20: Retrospective Review

    Retrospective Review • A post-hoc case review of ACURATE neo2 implantation evaluated proper valve frame expansion per angiographic imaging ➢ Under-expansion can be identified by non-parallel commissure posts Valve frame under-expansion was present in ~20% of ACURATE neo2 cases

  21. Slide 21: Bench Testing: Impact of Valve Frame Under-Expansion

    Bench Testing: Impact of Valve Frame Under-Expansion Video Video Expanded ACURATE neo2 Max Velocity 1.6 m/s • Laminar flow • Adequate washout Video Video Under-Expanded ACURATE neo2 Max Velocity 3.1 m/s • Turbulent flow • Reduced washout

  22. Slide 22: Post-Hoc Clinical Analysis: Valve Frame Expansion

    Post-Hoc Clinical Analysis: Valve Frame Expansion Time-to-Event Analysis through 1 Year Post-Procedure (N=703) • Patients with under-expanded valves had increased rates of death (2-fold) and stroke (3-fold) compared to those with well-expanded valves ACURATE neo2 Expanded Under-Expanded Valve Frame

  23. Slide 23: Post-Hoc Clinical Analysis: Valve Frame Expansion

    Post-Hoc Clinical Analysis: Valve Frame Expansion Death, stroke, or rehospitalization† (N=1433) ACURATE neo2 Under-expanded ACURATE neo2 Expanded Control 18.8% 12.4% ∆3.4% [-0.1%, 6.8%] 9.1% No. at risk ACUR. neo2 Under-exp. (N=150) 150 150 144 135 132 121 ACUR. neo2 Expanded (N=553) 553 543 535 514

  24. Slide 24: Post-Hoc Clinical Analysis: Valve Frame Expansion

    Post-Hoc Clinical Analysis: Valve Frame Expansion ACURATE neo2 Under-expanded ACURATE neo2 Expanded Control Death Stroke 11.0% 7.4% 3.7% 3.5% 3.6% 3.4% Days since Procedure Days since Procedure No. at risk No. at risk ACUR. neo2 ACUR. neo2 150 150 149 147 145 135 150 150 146 137 136 123 Under-exp. U

  25. Slide 25: ACURATE neo2 Under-Expansion: Potential Associations

    ACURATE neo2 Under-Expansion: Potential Associations • Valve under-expansion has the potential to impact later clinical outcomes • Procedural technique may play a role  Pre- and post-dilation was inconsistent with commercial practice due to use of smaller-than-recommended balloons, which has an imp

  26. Slide 26: Importance of Post-Dilation: Proof of Concept

    Importance of Post-Dilation: Proof of Concept ACURATE neo2 under-expansion can be easily recognized and is improved by post-dilation Valve under expanded with Post-dil resolves under-expansion Expanded valve with no angulated post evident & angulation of commissural post angulated post evident Video

  27. Slide 27: Importance of Post-Dilation: Case Example

    Importance of Post-Dilation: Case Example Under-expansion is easy to identify and correct during the procedure Video When is post dilation recommended? • Gradient ≥10mmHg ✓ • ≥ mild PVL ✓ • Valve under expansion • Identified by diameter or non-parallel posts (2nd view) • Easy to visualize on ACURATE

  28. Slide 28: Study Limitations

    Study Limitations • Trial-related factors  First RCT experience with ACURATE neo2 enrolled patients of all risks  Control arm included two comparator devices, with operator selection of preferred Control at time of randomization • Enrollment factors  COVID – impact on staffing resources (hospital

  29. Slide 29: ACURATE IDE Trial Key Take-Aways

    ACURATE IDE Trial Key Take-Aways • The ACURATE IDE Trial is the largest head-to-head trial to evaluate TAVR with ACURATE neo2  Ability to compare ACURATE neo2 with commercially available devices was complicated by a challenging trial environment (trial-related and enrollment factors) • Post-hoc ana

  30. Slide 30: Deep-dive session on the ACURATE IDE study

    Deep-dive session on the ACURATE IDE study • Date/Time: Wednesday, October 30th 12:15-1:30 pm • Location: Presentation Theater #5 (FDA Theater) • Moderator: Michael Joseph Rinaldi • Faculty • Michael J. Reardon – A Closer Look at the ACURATE IDE Study • Ole de Backer – ACURATE IDE Insights: Optimizi

Boston Scientific — Technical Presentation (October 2024) on 6ix