Slide 1: Better Diagnostics for Life – Interim Clinical Presentation sugarBEAT
NASDAQ: NMRD Better Diagnostics for Life Interim Clinical Presentation sugarBEAT® 12th September 2018
Slide 2: A Prospective Single Centre Evaluation of the Accuracy and Safety of sugarBEAT
A Prospective Single Centre Evaluation of the Accuracy and safety of the sugarBEAT® Non-invasive Continuous Glucose Monitor (CGM) System: An interim Evaluation Dr T Rahman, MBBS, PhD Clinical research Manager, Nemaura Medical Inc., 2
Slide 3: Study Objective
Study Objective A EU based study to determine the safety and accuracy of the sugarBEAT® CGM to support FDA submission 3
Slide 4: Study Design
Study Design ❖Prospective single arm, single centre study ❖Accuracy and safety assessed over 7 consecutive wear days, consisting of 3 non-consecutive in-clinic visits, and 4 home wear days. ❖Venous blood samples used as reference for in-clinic portion of study using Architect c8000 ❖Abbott Freestyle
Slide 5: Study Methods
Study Methods ❖>25 subjects enrolled for screening at each of 3 stages ❖25 subjects selected after screening in each stage, with approximately equal split between Type 1 and Type 2 ❖No subjects lost due to drop out ❖All 25 subjects completed 2 days home study and 3 days in- clinic study ❖12 subjects
Slide 6: Baseline Characteristics
Baseline Characteristics Variable Age - Range 27-67 years Age - Mean 52.8 years Age - Median 54.5 years Type 1 (n) 12 Type 2 (n) 13 6
Slide 7: Accuracy vs Glucose Range – 1-point Calibration
Accuracy vs Glucose Range In-clinic portion of study 1-point calibration Glucose Number of Within Within Within Within Within Outside ±40% range paired ±10% or ±10 mg/dl ±15% or ±15 mg/dl ±20% or ±20 mg/dl ±30% or ±30 mg/dl ±40% or ±40 mg/dl or ±40 mg/dl (mg/dl) points (MARD) (MARD) (MARD) (MARD) (M
Slide 8: Accuracy vs Glucose Range – 2-point Calibration
Accuracy vs Glucose Range In-clinic portion of study 2-point Calibration Glucose Number of Within Within Within Within Within Outside ±40% range paired ±10% or ±10 mg/dl ±15% or ±15 mg/dl ±20% or ±20 mg/dl ±30% or ±30 mg/dl ±40% or ±40 mg/dl or ±40 mg/dl (mg/dl) points % (MARD) % (MARD) % (MARD) % (
Slide 9: Interim Comparative MARD for In-Clinic phase as Primary Accuracy Metric
Interim Comparative MARD for In-Clinic phase as Primary Accuracy Metric Dexcom G5* no. of Overall MARD 9.00 calibrations (20%/20mg/dL) not known % Data 94.00 1-point Overall MARD sugarBEAT calibraiton (20%/20mg/dL) 8.77 % Data 62.61 2-point Overall MARD calibraiton (20%/20mg/dL) 7.97 % Data 74.00 1-
Slide 10: Interim Comparative MARD for home-study phase as Primary Accuracy Metric
Interim Comparative MARD for home-study phase as Primary Accuracy Metric Note 1: Senseonics, Dexcom G5 and sugarBEAT = 2 point finger prick calibration *http://www.diabetesincontrol.com/accuracy-comparison-of-the-dexcom-g5-abbott-freestyle-libre-pro-and- senseonics-eversense/ 10
Slide 11: Safety Evaluation
Safety Evaluation Previous studies: >100 Patient days wear time – No major adverse events or skin irritation reported. >525 Patient days wear time – No major adverse events or skin irritation reported. Current study: >250 Patient days to date with no major adverse events or skin irritation reported.
Slide 12: Conclusion
Conclusion The interim analysis indicates the 3 day in-clinic portion of the study shows accuracy levels, as measured using overall nominal MARD, as comparable to previously reported data and within the targets for the intended indications. 12