  1. Slide 1: Chromatography in the Fast Lane

    Chromatography in the Fast Lane Fast GC Mark Sinnott Application Engineer December 13, 2018 1

  2. Slide 2: Questions to Ask

    Questions to Ask What information do you need from your analysis? Do you have more baseline than you need between your peaks? Do you need to resolve all of the components? Page 2

  3. Slide 3: Variables for Shortening Run Times

    Variables for Shortening Run Times •Stationary Phase •Temperature Programming •Carrier Gas: type and linear velocity •Shorten Column Length •Decrease Film Thickness •Decrease Internal Diameter Page 3

  4. Slide 4: Resolution

    Resolution N  k    − 1 Rs =    4  k + 1    Factors effecting Rs Variables Efficiency N =  (gas, L, rc) L Length rc radius Retention k =  (T, df, rc) df film thickness Selectivity  =  (T, phase) T Temperature Page 4

  5. Slide 5: Resolution

    Resolution N  k    − 1 Rs =    4  k + 1    Factors effecting Rs Variables Efficiency N =  (gas, L, rc) L Length rc radius Retention k =  (T, df, rc) df film thickness Selectivity  =  (T, phase) T Temperature Page 5

  6. Slide 6: Stationary Phase - Common Types

    Stationary Phase - Common Types Siloxane polymers Poly(ethylene) glycols Porous polymers Page 6

  7. Slide 7: Selectivity

    Selectivity Relative spacing of the chromatographic peaks The result of all non-polar, polarizable and polar interactions that cause a stationary phase to be more or less retentive to one analyte than another Page 7

  8. Slide 8: Optimizing Selectivity

    Optimizing Selectivity Match analyte polarity to stationary phase polarity -like dissolves like(oil and water don’t mix) Take advantage of unique interactions between analyte and stationary phase functional groups Page 8

  9. Slide 9: Start with the Right Phase

    Start with the Right Phase DB-1 15m x 0.32mm, 0.25µm Oven: 40°C for 2 min 40-120°C at 5°C/min 0 2 4 6 8 10 12 14 Time (min.) DB-Wax 15m, 0.32mm, 0.25µm Oven: 80-190°C at 20°C/min 0 1 2 3 4 5 Time (min.) Page 9

  10. Slide 10: Resolution

    Resolution N  k    − 1 Rs =    4  k + 1    Factors effecting Rs Variables Efficiency N =  (gas, L, rc) L Length rc radius Retention k =  (T, df, rc) df film thickness Selectivity  =  (T, phase) T Temperature Page 10

  11. Slide 11: Column Length and Efficiency (Theoretical Plates)

    Column Length and Efficiency (Theoretical Plates) Length (m) n 15 69,450 30 138,900 60 277,800 0.25 mm ID n/m = 4630 (for k = 5) Page 11

  12. Slide 12: Column Length and Resolution

    Column Length and Resolution R  n  L Length X 4 = Resolution X 2 t  L Page 12

  13. Slide 13: Column Length VS Resolution and Retention: Isothermal

    Column Length VS Resolution and Retention: Isothermal R=0.84 R=1.16 R=1.68 2.29 min 4.82 min 8.73 min 15 m 30 m 60 m Double the plates, double the time but not double the the resolution Page 13

  14. Slide 14: Decrease the Length

    DECREASE THE LENGTH DB-5 1 4 30 m 2 0.53 mm I.D., 0.5 µm 3 5 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00 7.50 8.00 8.50 9.00 .5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 DB-5 1. Benzene 15 m 2. Toluene 0.53 mm I.D., 0.5 µm 3. Ethylbenzene 4 4. m,p-

  15. Slide 15: Column Length and Cost

    Column Length and Cost 15m 30m 60m $$$$ Page 15

  16. Slide 16: Length Summary

    Length Summary If you Decrease Length: Efficiency Decrease Resolution Decrease Analysis Time Decrease Pressure Decrease Cost Decrease Page 16

  17. Slide 17: Resolution

    Resolution N  k    − 1 Rs =    4  k + 1    Factors effecting Rs Variables Efficiency N =  (gas, L, rc) L Length rc radius Retention k =  (T, df, rc) df film thickness Selectivity  =  (T, phase) T Temperature Page 17

  18. Slide 18: Column Diameter - Theoretical Efficiency

    Column Diameter - Theoretical Efficiency Total Plates I.D. (mm) n/m 5m N ~ 112,000 0.05 23,160 10 m N ~ 112,000 0.10 11,580 0.18 6,660 20 m N ~ 112,000 0.20 5830 0.25 4630 30 m N ~ 112,000 0.32 3660 0.45 2840 0.53 2060 k=5 Page 18

  19. Slide 19: Different Column I. D.

    Different Column I. D. Equal Phase Ratios Column : DB-624 30 m, 0.53 mm, 3 m Carrier: Helium, 40(cm/sec) Oven: 65°C Injection: Split Detector: FID Column : DB-624 30 m, 0.32 mm, 1.8 m 0 5 10 15 20 Time (min) Page 19

  20. Slide 20: Phase Ratio

    PHASE RATIO () Film Thickness Column Dimensions Phase Ratio β 30 m x .53 mm x 3.0 m 44 30 m x .32 mm x 1.8 m 44 KC = k β β = r 2df Page 20

  21. Slide 21: Column Diameter and Capacity

    Column Diameter and Capacity I.D. (mm) Capacity (ng) 0.05 1-2 0.10 6-13 0.18 25-55 0.20 35-70 Like Polarity Phase/Solute 0.25 80-160 0.25 µm film thickness 0.32 110-220 0.45 600-800 0.53 1000-2000 Page 21

  22. Slide 22: Column Diameter - Inlet Head Pressures (Helium)

    Column Diameter - Inlet Head Pressures (Helium) I.D (mm) Pressure (psig) 0.05 275-400 0.10 90-130 0.18 30-45 0.20 25-40 0.25 15-25 30 meters 0.32 10-20 Hydrogen pressures x 1/2 0.45 3-7 0.53 2-4 Page 22

  23. Slide 23: Column Diameter and Carrier Gas Flow

    Column Diameter and Carrier Gas Flow Lower flow rates: Smaller diameter columns Higher flow rates: Larger diameter columns Low flow rates : GC/MS High flow rates: Headspace, purge & trap Page 23

  24. Slide 24: Diameter Summary

    Diameter Summary If you decrease the inside diameter: Efficiency Increase Resolution Increase Pressure Increase Capacity Decrease Flow rate Decrease Page 24

  25. Slide 25: Combining a Change in Length with a Change in Diameter

    Combining a change in Length with a change in Diameter Decrease Length Decrease Diameter Efficiency Analysis Time Page 25

  26. Slide 26: Carrier Gas Considerations

    Carrier Gas Considerations Best velocity? Optimal range of velocities Too low or high results in loss of resolution Balance resolution and analysis time BUT... there is no LAW against going FASTER Page 26

  27. Slide 27: Carrier Gas Considerations: Van Deemter Curve

    Carrier Gas Considerations Van Deemter Curve N2 Excessive Diffusion 1.00 Poor Mass Transfer 0.75 H (min) 0.50 He H2 0.25 uopt 60 10 20 30 40 50 60 u (cm/sec) Page 27

  28. Slide 28: Utilizing Computer Controlled Flow Ramping (EPC)

    Carrier Gas Considerations Utilizing Computer Controlled Flow Ramping (EPC) Lonely Late Eluters? = P Decrease retention and overall run time by increasing pressure (speed up the gas!) Page 28

  29. Slide 29: Fast Detector Electronics – 6890 FID

    Fast Detector Electronics – 6890 FID 200 Hz 100 Hz 50 Hz 20 Hz 10 Hz Proper peak quantitation typically requires a minimum of 10 data points per peak 10 Hz data rate: (peaks less than one second wide will have less than 10 data points) 200 Hz data rate: (peaks as narrow as 50 msec can be handled rel

  30. Slide 30: Easy Options with Method Translation Software

    Easy Options with Method Translation Software Different Column Dimensions Switch He to H2 Carrier Gas and Try Faster Velocities Same Column & Gas Type but Faster Velocities Combination of all of the above Page 30

  31. Slide 31: Temperature Programming

    Temperature Programming 30m, 0.25mm ID Temperature 10m, 0.1mm ID Temperature Temperature program must be modified to give same temperature of elution (i.e. faster ramps, shorter hold times)

  32. Slide 32: CLP-Pesticides - Original Improved Method

    CLP-Pesticides - Original “Improved” Method 0.32mm I.D., Helium Carrier Gas Column: DB-XLB 5,6 30m x 0.32mm i.d., 0.25µm G C 3 - 6 7 0 7 . D \ E C Carrier: D 2 B He, constant flow, 38 cm/s at 120°C 2 Injector: Pulsed Splittless, 220 °C 3 Pulse pressure & time: 35psi for 1.15min 2µL, 50ppb Oven: 120°

  33. Slide 33: Input Original Method Parameters

    Input Original Method Parameters Page 33

  34. Slide 34: Same Column and Gas Type Fast Analysis (OPGV)

    Same Column and Gas Type Fast Analysis (OPGV) Transfer same dimensions New Velocity New Temp. Program Page 34

  35. Slide 35: Same Column and Gas Type Fast Analysis (OPGV)

    1 . 9 e + 0 8 1 . 8 e + 0 8 1 . 7 e + 0 8 Same Column and Gas Type Before 1 . 6 e + 0 8 1 . 5 e + 0 8 1 . 4 e + 0 8 Fast Analysis (OPGV) 1 . 3 e + 0 8 1 . 2 e + 0 8 1 . 1 e + 0 8 1 e + 0 8 9 e + 0 7 8 e + 0 7 7 e + 0 7 6 e + 0 7 5 e + 0 7 4 e + 0 7 3 e + 0 7 No Real 2 e + 0 7 1 e + 0 7 T i m e Chang

  36. Slide 36: Same Column, Hydrogen Carrier Gas Translate Only

    Same Column, Hydrogen Carrier Gas Translate Only Page 38

  37. Slide 37: Same Column, H2 Carrier Gas Translate Only

    1 . 9 e + 0 8 1 . 8 e + 0 8 1 . 7 e + 0 8 Same Column, H2 Carrier Gas Before 1 . 6 e + 0 8 1 . 5 e + 0 8 1 . 4 e + 0 8 Translate Only 1 . 3 e + 0 8 1 . 2 e + 0 8 1 . 1 e + 0 8 1 e + 0 8 9 e + 0 7 8 e + 0 7 7 e + 0 7 6 e + 0 7 5 e + 0 7 4 e + 0 7 3 e + 0 7 2 e + 0 7 1 e + No Real 0 7 2 . 0 0 4 . 0 0

  38. Slide 38: New Column Dimensions, H2 Gas, Translate Only

    New Column Dimensions, H2 Gas, Translate Only Input NEW dimensions Close enough (got lucky) New Velocity New Temp. Program Page 42

  39. Slide 39: New Column Dimensions, H2 Gas, Translate Only

    New Column Dimensions, H2 Gas, Translate Only Page 43

  40. Slide 40: New Column Dimensions, H2 Gas, Translate Only

    New Column Dimensions, H2 Gas, Translate Only 5 10 11 6 Page 44

  41. Slide 41: New Column Dimensions, H2 Gas, Fast Analysis

    New Column Dimensions, H2 Gas, Fast Analysis New Velocity New Temp. Program Page 45

  42. Slide 42: New Column Dimensions, H2 Gas, Fast Analysis

    New Column Dimensions, H2 Gas, Fast Analysis G C 8 - 3 3 4 5 . D \ E C D 2 B Better Resolution and Faster Analysis! 2 . 0 0 2 . 5 0 3 . 0 0 3 . 5 0 4 . 0 0 4 . 5 0 5 . 0 0 5 . 5 0 6 . 0 0 6 . 5 0 7 . 0 0 7 . 5 0 Page 46

  43. Slide 43: New Column Dimensions, H2 Gas, Higher Velocities

    New Column Dimensions, H2 Gas, Higher Velocities Page 47

  44. Slide 44: New Column Dimensions, H2 Gas, Higher Velocities

    New Column Dimensions, H2 Gas, Higher Velocities 105 cm/sec ? 5.5 minutes! 95 cm/sec <6 minutes 85 cm/sec Page 48

  45. Slide 45: Final Method Used at EPA

    Column: DB-XLB Final Method Used at EPA Carrier: 20m x 0.18mm i.d., 0.18µm H2, constant flow, 77.3cm/s at 120°C Injector: Pulsed Splittless, 220 °C Pulse pressure & time: 35psi for 0.5min Flow ramp at 6.25min of 99mL/min2 to 3mL/min 2- 3 2mm i.d. liner G C 8 3 6 7 . D \ E C D 2 B 0.5µL, 50ppb Oven:

  46. Slide 46: Final Method Used at EPA

    Final Method 2 Used at EPA G C 8 - 3 3 6 7 . D \ E C D 2 B 5 3 7 12 10 DB-XLB 1 8 9 13 11 21 6 19 22 14 15 16 18 4 17 20 G C 8 - 3 3 6 7 . D \ E C D 1 A 1 . 5 0 2 . 0 0 2 . 5 0 3 . 0 0 3 . 5 0 4 . 0 0 4 . 5 0 5 . 0 0 5 . 5 0 6 . 0 0 6 . 5 0 7 . 0 0 2 3 5 7 V 9 10 12 i e w M o d e : I n t e g r a 21

  47. Slide 47: Food/Fragrance – Method translation

    Food/Fragrance – Method translation Page 51

  48. Slide 48: Food/Fragrance – Method translation

    Food/Fragrance – Method translation Page 52

  49. Slide 49: Spearmint Oil

    Spearmint Oil 10.6 min Δ -16.8 min (0.18 mm, H2 Carrier) 17.7 min Δ -9.7 min (0.18 mm, He Carrier) (0.25 mm, He Carrier) 27.4 min 3 8 13 18 23 28 Time (min) Page 53

  50. Slide 50: Spearmint Oil – Resolution Check

    Spearmint Oil – Resolution Check Δ -16.8 min (0.18 mm, H2 Carrier) 10.6 min Δ -9.7 min (0.18 mm, He carrier) 17.7 min (0.25 mm, He Carrier) 27.4 min 3 8 13 18 23 28 Page 54

  51. Slide 51: Resolution Maintained

    Resolution Maintained Compound Resolution Compounds 0.25 mm 0.18 mm 0.18 mm Helium Helium Hydrogen Sabinene 1.52 1.59 1.56 β-Pinene α-Terpinene 1.61 1.73 1.86 p-Cymene Speed Gain N/A 35% 61% Practical Fast GC Applications with Capillary GC Columns May 19, 2009 Page 55

  52. Slide 52: Conclusions

    CONCLUSIONS Stationary Phases – Chosen for optimized selectivity Diameter – Smaller allows shorter length but has less capacity Make Small Changes – Again think capacity Carrier Gas – Hydrogen, high velocity, but can still go fast with He Temperature Program – Scale properly to preserve elution patt

  53. Slide 53: The New Agilent Intuvo 9000 GC System

    The New Agilent Intuvo 9000 GC System

  54. Slide 54: Common frustrations with GC

    Common frustrations with GC • Measuring column length correctly • Cutting your column correctly • How tight it too tight? • Clipping columns to deal with active sites, then updating retention times

  55. Slide 55: Intuvo…

    Common Care and Maintenance Scheme for GC Columns 1. Cut off 6”-1ft of the inlet end of the column. 2. Bake out the column for no more than 2 hours. 3. Cut off more column. (repeat as necessary) Intuvo… Change the guard chip Guard Column

  56. Slide 56: Innovating the GC Flow Path

    Innovating the GC Flow Path Conventional Intuvo flow path flow path Click-and-run Inlet Inlet direct connections Guard chip Gold seal Flow chip Nuts and ferrules To detector Classic capillary Intuvo planar column column To detector

  57. Slide 57: Easier and faster maintenance with Intuvo

    Easier and faster maintenance with Intuvo • No more ferrules • Direct face seal connections • Audible and tactile click lets you know connection is made • Less unplanned downtime • Fewer batch reruns and precious samples lost

  58. Slide 58: No More - measuring - over-tightening

    No More - measuring - over-tightening - trimming

  59. Slide 59: A New Portfolio of GC Consumables

    A New Portfolio of GC Consumables UI Inlet Liners Guard Chip Smart Keys Intuvo Flow Chips No-trim column Tools & Accessories

  60. Slide 60: No More: - Measuring - Over-tightening

    No More: - Measuring - Over-tightening - Trimming

  61. Slide 61: Tips to Assure a Good Column Installation

    Tips to Assure a Good Column Installation 1. Finger tighten until only 1 thread on each of the two nuts is showing. If more than 1 thread is showing, wiggle or re-position the column into place to further finger tighten the nuts to 1 thread.

  62. Slide 62: Tips to Assure a Good Column Installation

    Tips to Assure a Good Column Installation 2. Check that the small integrated column nuts on the column are in form fitted place on the heater in the instrument. 3. Click and Run!!

  63. Slide 63: Smart Key Technology

    Smart Key Technology • Smart chip tells your Intuvo what you have • Sets temperature limits for you • Keep track of performance with read/writeable Smart Key

  64. Slide 64: Where Does it Get Dirty?

    Where Does it Get Dirty? Here Here Here Here Here Here Not the Column! Page 78

  65. Slide 65: New Autonomous Leak Checking

    New Autonomous Leak Checking