Arcoma AB · OMXSTO:ARCOMA
Arcoma AB — Technical Presentation (October 2020)
36 slides from Arcoma AB's technical presentation, Oct 23, 2020

Slide 1: Welcome to today's webinar from Arcoma
WELCOME TO TODAYS WEBINAR FROM WORLD CLASS DIGITAL X-RAY SYSTEMS The presentation will soon begin. Please prepare by disabling your camera and microphone: VIDEO MUTE OFF

Slide 2: The practical stuff
THE PRACTICAL STUFF: • VIDEO – OFF • MICROPHONE – OFF • QUESTIONS – USE CHAT VIDEO MUTE CHAT OFF

Slide 3: Agenda
ARCOMA WEBINAR PREMIUM DIGITAL X-RAY SYSTEMS AGENDA • What differs in Aceso and Aceso+ and what do they have in common? • Overview of system components and how they communicate • How to navigate our documentation and schematics • Q&A

Slide 4: Who am I?
ARCOMA WEBINAR PREMIUM DIGITAL X-RAY SYSTEMS WHO AM I? ERIK SÄLL ARCOMA ACADEMY MANAGER • 15 years of experience with medical x-ray equipment • Hands on technical background (service/installation/troubleshooting) • Focusing majority of time on training and knowledge transferring activities

Slide 5: What differs between Aceso+ and Aceso?
ARCOMA WEBINAR PREMIUM DIGITAL X-RAY SYSTEMS WHAT DIFFERS BETWEEN ACESO+ AND ACESO?

Slide 6: Aceso+
ARCOMA WEBINAR PREMIUM DIGITAL X-RAY SYSTEMS Aceso+ Arcoma product name = Precision Most advanced positioning system provided by Arcoma: • Fully automated system with motorized movements in X, Y, Z, and A, B directions • Configurable autopositions with various mode functionalities • Stitching availa

Slide 7: Aceso
ARCOMA WEBINAR PREMIUM DIGITAL X-RAY SYSTEMS Aceso Arcoma product name = Intuition Basic positioning system provided by Arcoma: • Motorized movements in Z directions only • Autotracking functionality • Stitching available towards Wallstand • Two table types available • Integrated service software fo

Slide 8: System components: generator, control software, positioning system
CMP200 Generator CXDI Control Software NE Positioning system with tube screen (Intuition or Precision) CXDI detectors Overwrap software

Slide 9

Slide 10: Receptor Setup software screen
M\ ARCOMA Receptor Setup ; Receptor = Last Image Hold (ms): Auto Focus Receptor Properties | AEC | Receptor Defaults | Inputs | Ouip ‘Tomo Back-up Time (ms): | 2500

Slide 11: Protocol Editor screen
M\ ARCOMA 83 Protocel Editor 9 Protocol * || exposure mode APRID “@ ib Test Wall Stand =? Wall sim @ Radiography @ Aid Test Wall Stand Fault = A3 Test Wall Stand @ A4 Test Wall Stand Coll Fail ‘@ B1 Test Table Flexible Copy2 ‘@ Bic Test Chest AP Table ‘@ Bic Test Chest AP Table COF= ‘@ Bid Test Tabl

Slide 12: Which system do you work with? Aceso+ or Aceso?
ARCOMA WEBINAR PREMIUM DIGITAL X-RAY SYSTEMS Which system do you work with? Aceso+ or Aceso? PLEASE TYPE SYSTEM NAME IN CHAT Aceso+ (Precision) Aceso (Intuition)

Slide 13: System components Aceso+: CAN bus communication
SYSTEM COMPONENTS Aceso+ CAN BUS COMMUNICATION System Master (CB800) Z/Master Signal WS Bucky interface Table Master Guard Table Guard X Y = isolated A/B independent UI CAN bus Collimator system

Slide 14: System components Aceso+: Ethernet communication
SYSTEM COMPONENTS Aceso+ ETHERNET COMMUNICATION CPI Overwrap Software System Master (CB800) Z/Master Signal WS Bucky interface Guard X Table Master Y Table Guard A/B UI = isolated Collimator independent CAN bus system Tube display

Slide 15: System components Aceso+: Serial communication
SYSTEM COMPONENTS Aceso+ SERIAL COMMUNICATION CPI Overwrap Software System Master (CB800) Z/Master Signal WS Bucky interface Guard X Table Master Y Table Guard A/B UI = isolated Collimator CPI Generator independent CAN bus system Tube display

Slide 16: Service tool connections Aceso+
SERVICE TOOL CONNECTIONS SERIAL COMMUNICATION Aceso+ CPI Overwrap Software Service PC System Master (CB800) Z/Master Signal WS Bucky interface Guard Programming X node Table Master Y Table Guard A/B UI = isolated Collimator CPI Generator independent CAN bus system Tube display

Slide 17: System components Aceso+: physical locations
CANON PC SYSTEM COMPONENTS Aceso+ CPI Overwrap Software PHYSICAL LOCATIONS WALL STAND TABLE OTC SYSTEM System Master CABINET (CB800) Z/Master Signal WS Bucky interface Guard X Table Master Y Table Guard A/B UI = isolated Collimator CPI Generator independent CAN bus system Tube display

Slide 18: Overview of system components: clarifications
OVERVIEW OF SYSTEM COMPONENTS CLARIFICATIONS THE TUBE DISPLAY ASSEMBLY CONTAINS TWO COMMUNICATION MODULES: - ETHERNET NODE COMMUNICATING WITH CB800 - CAN BUS NODE COMMUNICATING WITH COLLIMATOR AND Z/MASTER

Slide 19: Overview of system components: clarifications
OVERVIEW OF SYSTEM COMPONENTS CLARIFICATIONS THE SYSTEM MASTER NODE (CB800) COMMUNICATES BOTH BY ETHERNET AND CAN BUS: - ETHERNET COMMUNICATION WITH TUBE DISPLAY AND CANON PC - CAN BUS COMMUNICATION WITH Z/MASTER IN OTC

Slide 20: Overview of system components: communication with Canon NE
OVERVIEW OF SYSTEM COMPONENTS COMMUNICATION WITH CANON NE Canon NE CPI AND CANON • CPI Overwrap Software (and also Genware MP – the generator service CPI Overwrap Software software) is stored on the Canon PC and communicates with CPI generator • The CPI Overwrap Software communicates with CANON NE S

Slide 21: Service software connection Aceso+ Serial (RS 232)
SERVICE SOFTWARE CONNECTION Aceso+ Serial (RS 232) CONNECTING SERVICE SOFTWARE: RS 232 COMMUNICATION When connecting the Arcoma Service Software we use the already wired RS 232 connection on the Z/Master node CABLE ROUTED FROM Z/MASTER TO SYSTEM CABINET FOR EASIER CONNECTION

Slide 22: CAN bus / Serial RS 232 ports on CBDC-nodes
CAN bus / Serial RS 232 ports on CBDC-nodes CONNECTING SERVICE SOFTWARE: RS 232 COMMUNCATION We must always use RS 232 connection port CAN BUS PORT for serial communication with any node NOTE! Connecting a serial cable to any CAN bus port will cause SERIAL RS 232 PORT permanent damages to node

Slide 23: CAN bus nodes
CAN bus nodes CONSISTENT ELECTRIC DESIGN: - DIGITAL I/O = J0, J1, J2, J3 - ANALOGUE I/O = J4 - CAN BUS PORT = J6 - SERIAL PORT = J7 - 24 VDC POWER = J8

Slide 24: CAN bus systems in general: closed circuit
CAN BUS SYSTEMS IN GENERAL CLOSED CIRCUIT ALL NODES ON Node 1 Node 2 CAN BUS CIRCUIT NEEDS TO STAY CAN_H CONNECTED 120Ohm 120Ohm CAN_L

Slide 25: Software updating documentation
SOFTWARE UPDATING DOCUMENTATION SOFTWARE PACKAGES AND DOCUMENTATION • All software packages have a bundle version number SOFTWARE DOCUMENTATION: RLN AND UDI

Slide 26: Software updating documentation: RLN and UDI
SOFTWARE UPDATING DOCUMENTATION RLN AND UDI All software packages always include two types of critical documents: • The first one is called RLN RLN = RELEASE NOTES • The second one is called UDI UDI = UPDATE INSTRUCTION

Slide 27: RLN - Release notes
SOFTWARE UPDATING DOCUMENTATION RLN - Release notes Release notes is used to determine which version of each node belonging to a specific software package. One for OTC-level (CAN bus positioning system)

Slide 28: RLN - Release notes
SOFTWARE UPDATING DOCUMENTATION RLN – Release notes And another one for system level (CB800, CPI overwrap)

Slide 29: UDI - Update Instruction
SOFTWARE UPDATING DOCUMENTATION UDI – Update Instruction Same thing with the Update Instructions – there is one for OTC level and another one for system level.

Slide 30: Software updating documentation - how to load SW
SOFTWARE UPDATING DOCUMENTATION – HOW TO LOAD SW HOW TO LOAD SOFTWARE We have created a guidance document explaining the following: • How to use the RLN and UDI documents STEP BY STEP GUIDE • In which order to read and determine versions AVAILABLE ON • When and how to save a backup parameter (system

Slide 31: Software updating: how to load software
SOFTWARE UPDATING HOW TO LOAD SOFTWARE Let’s have a look at the document together: DOWNLOAD How to load software HOW TO GUIDE

Slide 32: Schematics: how to navigate
SCHEMATICS HOW TO NAVIGATE There are 3 different types of schematics: BLOCK DIAGRAMS • SBD = System Block Diagram AND WIRING • UBD = Unit Block Diagram DIAGRAMS • WRD = Wiring Diagram

Slide 33: Schematics: how to navigate
SCHEMATICS HOW TO NAVIGATE All cables and wires are categorized with consistent names: • 1.x = Tube stand • 2.x = Table • 3.x = Wallstand • 4.x = System cabinet • 5.x = Image system

Slide 34: Schematics: how to navigate easily
SCHEMATICS HOW TO NAVIGATE EASILY Different schematic types and consistent labeling - Let’s have a look at them! SCHEMATICS AND Installation and Service manual LABELINGS

Slide 35: Questions?
Questions? Feel free to send us questions later: erik.sall@arcoma.se or service@arcoma.se

Slide 36: Thank you
THANK YOU Arcoma AB | Annavägen 1 | 352 46 Växjö | Sweden PHONE EMAIL WEBSITE +46 470 70 69 00 office@arcoma.se www.arcoma.se
Source: Internet Archive copy of arcoma.se
More decks from Arcoma AB
Investor Presentation (July 2024)13 slides · July 2024
Investor Presentation (July 2024)10 slides · July 2024
Investor Presentation (2024)25 slides · Jun 30, 2024
Investor Presentation (June 2024)34 slides · June 2024
Investor Presentation (June 2024)37 slides · June 2024
Investor Presentation (May 2024)12 slides · May 2024