GE HealthCare Technologies Inc. · NASDAQ:GEHC

GE HealthCare Technologies — Technical Presentation (April 2020)

29 slides from GE HealthCare Technologies's technical presentation, Apr 21, 2020

  1. Slide 1: Weaning from Mechanical Ventilation using SBT

    Weaning from Mechanical Ventilation using SBT April 21, 2020

  2. Slide 2: Objectives

    Objectives By the end of this course, you should be able to: • Define SBT and weaning • Describe the factors to assess weaning readiness • Understand the types of Spontaneous Breathing Trials (SBT) • Understand SBT Success/failure criteria • Define weaning parameters used to predict weaning and extu

  3. Slide 3: The amount of time spent on weaning a patient from mechanical ventilation

    40% This Photo by Unknown Author is licensed under CC BY-SA-NC The amount of time spent on weaning a patient from mechanical ventilation Zein H, Baratloo A, Negida A, Safari S. Ventilator Weaning and Spontaneous Breathing Trials; an Educational Review. Emerg (Tehran). 2016;4(2):65–71 JB76030XX

  4. Slide 4: What is SBT and Weaning?

    What is SBT and Weaning? JB76030XX

  5. Slide 5: What is Spontaneous Breathing Trial and Weaning?

    What is Spontaneous Breathing Trial and Weaning? Weaning: • The process used to describe the gradual decrease in ventilatory support Spontaneous Breathing Trial (SBT): • The mechanism used to assess the patient’s ability to breath with minimal or no ventilatory support Zein H, Baratloo A, Negida A,

  6. Slide 6: Assessing the Readiness to Wean

    Assessing the Readiness to Wean JB76030XX

  7. Slide 7: Assessing the Readiness to Wean

    2 Assessing the Readiness to Wean 1 4 Factor to assess weaning readiness: 1. Reversal of the indication for ventilatory support 2. Adequate gas exchange  PaO2 > 60mmHg,  FiO2 ≤ 50%  PEEP ≤ 8 cmH20  pH > 7.25  Vd/Vt < 60%  MV < 12 L/min 3. Ability to initiate a breath 4. Hemodynamic stability 

  8. Slide 8: Types of SBT

    Types of SBT JB76030XX

  9. Slide 9: Types of Spontaneous Breathing Trials

    1 Types of Spontaneous Breathing Trials 1. T-Piece Trial https://link.springer.com/chapter/10.100  Humidified oxygen connected directly to the ET tube with a T-piece 7/978-3-319-89981-7_9 2. CPAP Trial*  CPAP level equal to current PEEP level 3. Low level of Pressure Support or Automatic Tube Comp

  10. Slide 10: SBT Duration and Success/Failure Criteria

    SBT Duration and Success/Failure Criteria JB76030XX

  11. Slide 11: SBT Duration and Success/Failure Criteria

    SBT Duration and Success/Failure Criteria SBT Duration: 30-120 minutes, once per day. Criteria of Successful Spontaneous Breathing Trials1 Criteria for Failure of a Spontaneous Breathing Trial2 • Respiratory rate < 35 breaths/minute • Respiratory rate > 35 breaths/minute • Heart Rate < 140/minute or

  12. Slide 12: Weaning Parameters

    Weaning Parameters JB76030XX

  13. Slide 13: Weaning Parameters

    Weaning Parameters • Used to predict weaning readiness Predictor Value • Weaning parameters are categorized Evaluation of ventilatory drive  P0.1 > ‐4cmH20 into 3 groups – Evaluation of ventilatory drive Ventilatory muscle capabilities  Vital Capacity > 10 mL/kg – Ventilatory muscle capability  M

  14. Slide 14: Weaning Parameter Definitions

    Weaning Parameter Definitions P0.1: is a measurement of the airway occlusion pressure during the first 0.1 seconds after beginning an inspiratory effort against an occluded airway • Helps to assess respiratory drive and inspiratory effort • Reference value: 1 • Healthy adults breathing spontaneously

  15. Slide 15: Is using weaning protocols better than no protocol at all?

    Is using weaning protocols better than no protocol at all? JB76030XX

  16. Slide 16: Weaning time reduction when standardized weaning protocols were used

    78% This Photo by Unknown Author is licensed under CC BY-SA-NC Weaning time reduction when standardized weaning protocols were used Blackwood B, Alderdice F, Burns K, Cardwell C, Lavery G, O'Halloran P. Use of weaning protocols for reducing duration of mechanical ventilation in critically ill adult

  17. Slide 17: Weaning Protocols

    Weaning Protocols • Consist of 3 parts1: 1. Objective criteria to judge weaning readiness 2. Guidelines to decrease support gradually 3. Criteria to assess readiness for extubation • Protocols implemented by Respiratory Therapists or Nurses result in shorter weaning times and shorter lengths of 2,3

  18. Slide 18: CARESCAPE R860 SBT

    CARESCAPE R860 SBT JB76030XX

  19. Slide 19: CARESCAPE R860 SBT

    CARESCAPE R860 SBT • A procedure intended to evaluate a patient’s ability to breathe spontaneously during a specified duration of time • Not intended to be used for long term ventilation beyond the set duration time • During SBT mode, the patient initiates spontaneous breaths and the ventilator main

  20. Slide 20: SBT Ends

    SBT Ends Reaching Stop Criteria Successful Completion of SBT • Once the SBT is complete, an alarm • If stop criteria are reached, the alarm message, message stating, “ SBT Completed “SBT ended” displays and the previous Successfully” will be displayed ventilation mode resumes • The previous ventilat

  21. Slide 21: SBT View

    SBT View • Shows graphical and numerical data and trends related to the ongoing trial • Data may be used to see the patient’s progress during the SBT or following the trials completion • Previous SBT data may be reviewed if it was obtained within the last 12 hours • If desired, select Guide and sele

  22. Slide 22: Lung Mechanics on the CARESCAPE R860

    Lung Mechanics on the CARESCAPE R860 JB76030XX

  23. Slide 23: P0.1

    P0.1 P 0.1 is a respiratory measurement used to evaluate the patient’s readiness to be weaned from the ventilator. P 0.1 is a measurement of the airway occlusion pressure 0.1 second after beginning an inspiratory effort against an occluded airway. P0.1 helps to assess respiratory drive and inspirato

  24. Slide 24: Negative Inspiratory Force (NIF)

    Negative Inspiratory Force (NIF) Negative Inspiratory Force is a weaning measurement that is used to evaluate the patient’s readiness to be weaned from the ventilator. NIF is used to determine the patient’s ability to take a deep breath and to generate a cough strong enough to clear secretions. To m

  25. Slide 25: Vital Capacity

    Vital Capacity Vital Capacity is the measurement of a patient’s largest (VTexp) expired Tidal Volume over a 30 second period. During a Vital Capacity measurement, Pinsp and PS are set to zero. When the Vital Capacity measurement is complete Pinsp and PS return to the previous setting. During the Vit

  26. Slide 26: Conclusion

    Conclusion This concludes the Weaning from mechanical ventilation using Spontaneous Breathing Trials course. In this course, you learned about: • SBT and weaning definitions • The factors to assess weaning readiness • The types of Spontaneous Breathing Trials (SBT) • SBT Success/failure criteria • W

  27. Slide 27: Imagination at work

    Imagination at work

  28. Slide 28: Slide 28

  29. Slide 29: Slide 29