LivaNova PLC · NASDAQ:LIVN

LivaNova — Technical Presentation

138 slides from LivaNova's technical presentation

  1. Slide 1: User Training

    1 User Training 1/2 Installation and operation 2/2 Configuration IM-7300695-CP-EHLM-1.2 _B #EssenzEra #NewEraofPerfusion

  2. Slide 2: Training introduction

    2 ▪ We will be walking through information regarding the operation, safety and set up of the Essenz Heart Lung Machine. ▪ The following presentation will walk through a series of videos and associated knowledge checks. ▪ Please refer to the product Instruction For Use for additional information and

  3. Slide 3: Training topics

    3 CONCEPT: OPERATION: 1. Overall system concept and architecture 9. Graphical user interface 2. Profiles 10. Operation pre-bypass mode INSTALLATION: 11. Operation bypass mode 3. Hardware configuration 12. Operation pumps 4. Connection of pumps and clamps 13. Operation sensors 5. Connection of sensor

  4. Slide 4: Concept: Overall system concept and architecture

    4 0% CONCEPT: 1. Overall system concept and architecture IM-7300695-CP_C #EssenzEra #NewEraofPerfusion

  5. Slide 5: Video 1

    5 Video 1

  6. Slide 6: Overall system concept and architecture

    6 Overall system concept and architecture True or false? Essenz is a modular HLM. The oxygenator can be mounted on the right or left side of the console. LivaNova provides optimized configurations. You can create your own configuration.

  7. Slide 7: Overall system concept and architecture

    7 Overall system concept and architecture Name the items for monitoring and controlling Cockpit Back-up Control Panel Control Unit or CU Essenz Patient Monitor

  8. Slide 8: Concept: Profiles

    8 4% CONCEPT: 2. Profiles IM-7300695-CP_C #EssenzEra #NewEraofPerfusion

  9. Slide 9: Video 2

    9 Video 2

  10. Slide 10: Profiles

    10 Profiles True or false? The name of a pump is important as it indicates its function. Sensors are automatically linked to a pump by its identified function. Minimum number of profiles: 1 Maximum number of profiles: 12

  11. Slide 11: Profiles

    11 Profiles True or false? The profile concept means that all settings are maintained as you defined them, when selecting a profile and starting operation. Changes of a profile during operation are possible. Changes during operation will be saved in the profile and are available for future cases. Ch

  12. Slide 12: Installation: Hardware configuration

    12 9% Installation: 3. Hardware configuration IM-7300695-CP_C #EssenzEra #NewEraofPerfusion

  13. Slide 13: Video 3

    13 Video 3

  14. Slide 14: Hardware configuration

    14 Hardware configuration True or false? 1 knob will always control 1 pump or the venous clamp on the Essenz Perfusion System The 4 knobs of the back-up control panel always control the same pumps or venous clamp as the 4 knobs of the cockpit.

  15. Slide 15: Hardware configuration

    15 Hardware configuration Where can you easily connect an additional pump? At the back-up control panel. An additional CU is needed to operate the pump! The pump can not be directly connected to the connector.

  16. Slide 16: Hardware configuration

    16 Hardware configuration Where can you find the 3 hand cranks for roller pumps? Inside the lower door of the cabinet.

  17. Slide 17: Installation: Connection of pumps and clamps

    17 14% Installation: 4. Connection of pumps and clamps IM-7300695-CP_C #EssenzEra #NewEraofPerfusion

  18. Slide 18: Video 4

    18 Video 4

  19. Slide 19: Connection of pumps and clamps

    19 Connection of pumps and clamps These icons represent which devices? Small & Large Roller Pump Centrifugal Pump Venous Clamp VC

  20. Slide 20: Connection of pumps and clamps

    20 Connection of pumps and clamps Which connectors need to be used? Devices controlled by the 4 knobs of the cockpit 1A, 1B, 2, 3, 4 Control Unit CU (pumps & venous clamp) 5 - 13 Arterial Clamp AC 9 - 13 Cockpit Control Units

  21. Slide 21: Connection of pumps and clamps

    21 Connection of pumps and clamps Devices controlled by cockpit grey black

  22. Slide 22: Connection of pumps and clamps

    22 Connection of pumps and clamps Example of devices controlled by cockpit grey black

  23. Slide 23: Connection of pumps and clamps

    23 Connection of pumps and clamps An Example of devices controlled by CUs grey black

  24. Slide 24: Connection of pumps and clamps

    24 Connection of pumps and clamps An Example of devices controlled by CUs Grey Black grey black

  25. Slide 25: Connection of pumps and clamps

    25 Connection of pumps and clamps grey black

  26. Slide 26: Connection of pumps and clamps

    26 Connection of pumps and clamps Cockpit itself connects to 5 - 13 grey black

  27. Slide 27: Connection of pumps and clamps

    27 Connection of pumps and clamps Arterial clamp connects to 9 - 13 ≠ Venous Clamp grey black

  28. Slide 28: Installation: Connection of sensors and devices

    28 21% Installation: 5. Connection of sensors and devices IM-7300695-CP_C #EssenzEra #NewEraofPerfusion

  29. Slide 29: Video 5

    29 Video 5

  30. Slide 30: Overall system concept and architecture

    30 Overall system concept and architecture Which icons represent which devices? Pressure Level Bubble Temperature Flow Blood Gas

  31. Slide 31: Overall system concept and architecture

    31 Overall system concept and architecture Maximum number of sensor modules is 10 Pressure Flow Level Bubble Temperature Blood Gas

  32. Slide 32: Connection of pumps and clamps

    32 Connection of pumps and clamps Gas blender connects to connectors 9 - 13 grey black

  33. Slide 33: Mounting: Mounting pumps and clamps

    33 26% Mounting: 6. Mounting pumps and clamps IM-7300695-CP_C #EssenzEra #NewEraofPerfusion

  34. Slide 34: Video 6

    34 Video 6

  35. Slide 35: Mounting roller pumps and venous clamp

    35 Mounting roller pumps and venous clamp What do you need to obey when mounting the arterial pump with the arterial extension arm at a very low position? Ensure to mount the pump in a straight line with the arm to avoid the collision with the wheel. Use cable guides at the bottom of the pump housin

  36. Slide 36: Mounting roller pumps and venous clamp

    36 Mounting roller pumps and venous clamp left side right side

  37. Slide 37: Mounting: Centrifugal pump and arterial clamp

    37 34% Mounting: 7. Mounting centrifugal pump and arterial clamp IM-7300695-CP_C #EssenzEra #NewEraofPerfusion

  38. Slide 38: Video 7

    38 Video 7

  39. Slide 39: Mounting centrifugal pump and arterial clamp

    39 Mounting centrifugal pump and arterial clamp True or false? The arterial clamp is operated by a control knob or a control unit. The arterial clamp is triggered by alarms, negative flow and setting the rpm of the arterial pump to zero.

  40. Slide 40: Mounting: Mounting other devices

    40 37% Mounting: 8. Mounting other devices IM-7300695-CP_C #EssenzEra #NewEraofPerfusion

  41. Slide 41: Video 8

    41 Video 8

  42. Slide 42: Mounting other devices

    42 Mounting other devices True or false? The cockpit can be mounted without any tool.

  43. Slide 43: Mounting other devices

    43 Mounting other devices ▪ Tip! (Connection of a second, spare cockpit) Behavior description Best practices If the Cockpit should fail, you can connect a If the second Cockpit has no profile loaded, you second, spare cockpit to the HLM. The original must operate the pumps with caution as no cockpit

  44. Slide 44: Mounting other devices

    44 Mounting other devices Control Units (CUs) can be mounted at… Single holder Double holder possible to act as a double pump Double arm Housing of the large roller pump not possible at small roller pump

  45. Slide 45: Operation: Graphical user interface

    45 44% Operation: 9. Graphical user interface IM-7300695-CP_C #EssenzEra #NewEraofPerfusion

  46. Slide 46: Video 9

    46 Video 9

  47. Slide 47: Graphical user interface

    47 Graphical user interface Name the 4 areas of the graphical user interface. Monitoring Dynamic Navigation Control

  48. Slide 48: Graphical user interface

    48 Graphical user interface Identify the values found in the 4 rows in the Monitoring area? A. Additional Flows & Pressures B. Venous Clamp C. Bubble & Level D. Gas Blender E. Blood Gas F. Arterial Flow G.Temperatures H. 3T F. Arterial flow 1 C. Bubble & level A. Additional flows & pressures 2 B. Ve

  49. Slide 49: Graphical user interface

    49 Graphical user interface What are the 4 tabs in the navigation area? Overview Cardioplegia Gas Blender Heater Cooler

  50. Slide 50: Graphical user interface

    50 Graphical user interface True or false? There are 3 timers in the navigation area. There is touch functionality in the control area and on the display of the seperate control unit. There is NO touch functionality in the control area and on the display of the separate control unit.

  51. Slide 51: Graphical user interface

    51 Graphical user interface Which flow values can you see when you use a Roller Pump? Measured flow When you use a Centrifugal Pump, both values are the measured flow. Calculated flow

  52. Slide 52: Graphical user interface

    52 Graphical user interface What does it mean when you see a black and white screen? All values are within their predefined ranges and no alarms are present.

  53. Slide 53: Operation: Operation pre-bypass mode

    53 50% Operation: 10. Operation pre-bypass mode IM-7300695-CP_C #EssenzEra #NewEraofPerfusion

  54. Slide 54: Video 10

    54 Video 10

  55. Slide 55: Operation pre-bypass mode

    55 Operation pre-bypass mode True or false? You can only start a pump when a profile is selected. Pumps can already be operated a few seconds after switch on of the HLM. Pay attention as no controls are active!

  56. Slide 56: Operation pre-bypass mode

    56 Operation pre-bypass mode How do you recognize you are in the pre-bypass mode? Blue frame with blue tag pre-bypass mode Start case button Sensor safety checks No timers available All of the above

  57. Slide 57: Operation pre-bypass mode

    57 Operation pre-bypass mode How do you recognize you are in the pre-bypass mode? Blue frame Start case button with blue tag pre-bypass mode No timers available Sensor safety checks IM-7300695-CP_C

  58. Slide 58: Operation pre-bypass mode

    58 Operation pre-bypass mode How do you recognize that the flow sensor is mounted in the wrong direction? No flow value displayed Negative flow value displayed Zero flow value displayed

  59. Slide 59: Operation pre-bypass mode

    59 Operation pre-bypass mode How do you recognize that the flow sensor is mounted in the wrong direction? The flow value is negative.

  60. Slide 60: Operation pre-bypass mode

    60 Operation pre-bypass mode How can you reognize that the venous clamp is calibrated or not? Not calibrated Calibrated

  61. Slide 61: Operation pre-bypass mode

    61 Operation pre-bypass mode Which sensors have to be switched on while going through the checklist? Bubble Pressures are already Level switched on by default. Gas blender

  62. Slide 62: Operation pre-bypass mode

    62 Operation pre-bypass mode What can the checklist help with? Identify defective sensors Identify sensors not mounted Identify sensors wrongly mounted Identify a mix-up of pressure sensors Exception: pre- and post oxygenator pressures are checked together in one line

  63. Slide 63: Operation pre-bypass mode

    63 Operation pre-bypass mode How does the Start Case-button change once all checks are successfully completed?

  64. Slide 64: Operation pre-bypass mode

    64 Operation pre-bypass mode Is it possible to tap the grey Start Case-button anytime and move into bypass mode even if not all checks have been successfully completed? Yes, if you confirm the warning message. Note, that this is not recommended and intended for emergency situations only.

  65. Slide 65: Operation pre-bypass mode

    65 Operation pre-bypass mode What happens if you tap the grey or blue start case-button? ▪ You move into bypass mode. ▪ All safety sensors and the electronic gas blender are switched on automatically. ▪ The tab safety checks is replaced by overview. ▪ Timers are available.

  66. Slide 66: Operation pre-bypass mode

    66 Operation pre-bypass mode ▪ Tip! (Switch on bubble and level sensor in pre-bypass mode) Behavior description Best practices When preparing for a case in pre-bypass mode, Use the On/Off switches inside the checklist to the Cockpit could lose its touch functionality switch on the bubble and level s

  67. Slide 67: Operation: Operation bypass mode

    67 65% Operation: 11. Operation bypass mode IM-7300695-CP_C #EssenzEra #NewEraofPerfusion

  68. Slide 68: Video 11

    68 Video 11

  69. Slide 69: Operation bypass mode

    69 Operation bypass mode True or false? When you change to bypass mode timers appear in the navigation area. When you change to bypass mode the safety check dialog is replaced by overview. If you select overview, patient data and Timers are visible in the dynamic area.

  70. Slide 70: Operation bypass mode

    70 Operation bypass mode True or false? The venous clamp is linked to the arterial flow and closes automatically when the flow is zero. You can’t override the venous clamp closure. You can override the closure by turning the knob.

  71. Slide 71: Operation bypass mode

    71 Operation bypass mode When does the arterial clamp close? In the event of a bubble alarm, level alarm or negative flow. When does the arterial clamp open? When the centrifugal pump reaches 1500 rpm or more.

  72. Slide 72: Operation bypass mode

    72 Operation bypass mode What is the keep-lower link? The sum of all pumps delivering blood to the patient (cardioplegia, cerebral…) is never higher than the main arterial flow. How is a link visualized? By the link icon and a colored line below the value (red, yellow, blue). (If the link icon is no

  73. Slide 73: Operation bypass mode

    73 Operation bypass mode When is the keep-lower link active? When the arterial pump is a roller pump. It is not active if the arterial pump is a centrifugal pump.

  74. Slide 74: Operation: Operation pumps

    74 72% Operation: 12. Operation pumps IM-7300695-CP_C #EssenzEra #NewEraofPerfusion

  75. Slide 75: Video 12

    75 Video 12

  76. Slide 76: Operation pumps

    76 Operation pumps How many timers are available in total? 6 How many and which timers are always available? 3

  77. Slide 77: Operation pumps

    77 Operation pumps What is a reminder? A count-down timer How many reminders are available to run at a given time? 1 How many default times are possible for this reminder? 3

  78. Slide 78: Operation pumps

    78 Operation pumps True or false? You can open the dialog of a value by tapping the value. The values are always visible in the monitoring area, even while changing the value.

  79. Slide 79: Operation pumps

    79 Operation pumps What parameters can you see/select for a centrifugal pump? Tube size Linked sensors Switch between flow- and rpm-mode Manually open the arterial clamp (Only possible if the centrifugal pump is at 1500 rpm or below.)

  80. Slide 80: Operation pumps

    80 Operation pumps Flow-mode and RPM-mode: Identify the icon and state how the mode operates? Flow-Mode Flow is controlled by the knob. Pump is adjusting the rpm to maintain the required flow. RPM-Mode Revolutions per minute are controlled by the knob. Flow is established due to pressure load situat

  81. Slide 81: Operation pumps

    81 Operation pumps What parameters can you see/select for a roller pump? Tube size Linked sensors Change direction of rotation (Only possible if the roller pump is stopped.)

  82. Slide 82: Operation pumps

    82 Operation pumps How does the icon of the roller pump appear if a change of direction vs the profile is established during a case? The arrows will change direction and the icon turns yellow. Art. Art.

  83. Slide 83: Operation: Operation sensors

    83 79% Operation: 13. Operation sensors IM-7300695-CP_C #EssenzEra #NewEraofPerfusion

  84. Slide 84: Video 13

    84 Video 13

  85. Slide 85: Operation sensors

    85 Operation sensors How can you switch off sensors? 1. Tap on the Sensor in the monitoring area 2. Switch it OFF 3. Confirm

  86. Slide 86: Operation sensors

    86 Operation sensors You can change thresholds by using the... +/- Signs Numerical keyboard Slider

  87. Slide 87: Operation sensors

    87 Operation sensors The distance Stop limit between regulation and stop limit is fixed and can only be changed Regulation limit while creating a profile.

  88. Slide 88: Operation sensors

    88 Operation sensors How do you open the Gas Blender values in the dynamic area? Click on the gas flow, FiO2 or CO2 flow value Click on gas blender in the navigation area

  89. Slide 89: Operation sensors

    89 Operation sensors True or false? You can open the Heater Cooler dialog by tapping on any Heater Cooler temperature or tapping Heater Cooler in the Navigation Area. Tapping the warm or cold icon selects the warm or cold temperature of the cardioplegia circuit. The water flow is stopped when neithe

  90. Slide 90: Operation sensors

    90 Operation sensors Where can you modify a pump (tube size, direction of rotation, name)?

  91. Slide 91: Operation sensors

    91 Operation sensors True or false? You can change the name of a pump and with it the role of the pump. After changing a pump name, linked sensors also change accordingly. When selecting a new pump name, you must also select the new tube size and direction of rotation.

  92. Slide 92: Operation sensors

    92 Operation sensors ▪ Tip! (Cardioplegia dose counting after a cardioplegia-related role change) Behavior description Best practices After a cardioplegia role change is performed, Only if you perform a role change involving a the cardioplegia dose volume (and in some cardioplegia pump, note down th

  93. Slide 93: Operation sensors

    93 Operation sensors ▪ Tip! (Connection of an additional pump) Behavior description Best practices If you connect an additional pump and its Try to determine the name/role of the new control unit to the HLM, a dialog pops up to pump by tapping the gear icon followed by the determine the name/role of

  94. Slide 94: Operation sensors

    94 Operation sensors ▪ Tip! (Exchange of arterial pump) Behavior description Best practices If you have problems with your arterial pump Once you have switched to a new arterial and you need to connect an additional pump, pump, stay with it until the end of the case. Do you can name it Arterial and

  95. Slide 95: Operation sensors

    95 Operation sensors ▪ Tip! (Connection of a second, arterial centrifugal pump) Behavior description Best practices If your arterial pump is a centrifugal pump and Observe the flow value in the monitoring area. if it should fail, you can connect a second Pay attention to negative flow. Use the pump

  96. Slide 96: Operation sensors

    96 Operation sensors How to lock and unlock the screen to disable the touch functionality? Tap for 2 seconds Lock Unlock

  97. Slide 97: Operation sensors

    97 Operation sensors What happens when you tap this icon? The last case is saved even Exit the profile when you switch ESSENZ off and on again. Open the last case by tapping the Last Case button. All pumps and timers have to be stopped to be able to exit the case.

  98. Slide 98: Operation: Cardioplegia

    98 99% Operation: 14. Cardioplegia IM-7300695-CP_C #EssenzEra #NewEraofPerfusion

  99. Slide 99: Video 14

    99 Video 14

  100. Slide 100: Cardioplegia

    100 Cardioplegia What information is displayed in the Cardioplegia dialog? Ratio of the pumps Selection of pressure limits CPL blood pump Ante-/retrograde pressure CPL crystalloid pump sensor Bubble sensor Warm/Cold circuit Temp. sensor Volumes Timers Delivery button (Start/Stop)

  101. Slide 101: Cardioplegia

    101 Cardioplegia What information is displayed in the Cardioplegia pump data boxes? Direction and type Flow Tube size

  102. Slide 102: Cardioplegia

    102 Cardioplegia True or false? You can select the ratio of the Cardioplegia pumps. It is not possible to operate the pumps separately. It is possible to operate the pumps separately by selecting Manual. You can change the ratio while the pumps are turning. The pumps adapt to the new ratio without s

  103. Slide 103: Cardioplegia

    103 Cardioplegia True or false? When you start the Cardioplegia delivery and start the pumps, the dose volume and delivery time starts to count. When you stop the pumps the counting pauses automatically. The Ischemia timer starts. The Ischemia timer is also visible it the navigation area. When you s

  104. Slide 104: Profile management: Profile mapping

    104 0% Profile management: 15. Profile mapping IM-7300695-CP_C #EssenzEra #NewEraofPerfusion

  105. Slide 105: Video 15

    105 Video 15

  106. Slide 106: Profile mapping

    106 Profile mapping What components are displayed for mapping a profile? Cockpit knobs CU knobs Sensors

  107. Slide 107: Profile mapping

    107 Profile mapping What components are displayed for mapping a profile? Name → Function Color Direction of rotation Linked sensors are indicated Tube size by the name of the pump IM-7300695-CP_C

  108. Slide 108: Profile mapping

    108 Profile mapping What is the difference between a disabled pump and a spare pump? Not enabled in the profile and Disabled can not be used during an operation (at least not with a name/role) Name/Function can be selected Spare during an operation

  109. Slide 109: Profile mapping

    109 Profile mapping What is the difference of these Cardioplegia pump names? 1 Cardioplegia pump CPL with 2 tubes 2 Cardioplegia pumps CPL - B with 1 tube each → Blood pump 2 Cardioplegia pumps CPL - C with 1 tube each → Crystalloid pump The name of the sensors must precisely match the cardioplegia

  110. Slide 110: Profile mapping

    110 Profile mapping What does the blue frame and blue settings icon indicate? One or more rules of the system mapping are not met. The profile is not completely mapped and can not be used. Open the notification list to receive hints on how to solve the violation.

  111. Slide 111: Profile management: Profile preferences

    111 44% Profile management: 16. Profile preferences IM-7300695-CP_C #EssenzEra #NewEraofPerfusion

  112. Slide 112: Video 16

    112 Video 16

  113. Slide 113: Profile preferences

    113 Profile preferences What settings can you define in the profile preferences? Direction of pump rotation Timers Reminder Max/min limits Defalut settings for starting bypass All of the above What does the blue dot indicate? You have not yet changed the settings for this field. (And you do not have

  114. Slide 114: Profile preferences

    114 Profile preferences Where can you change the formula for calculating the body surface area? Tap overview Tap gear icon Select a formula

  115. Slide 115: Profile preferences

    115 Profile preferences Where can you change the cardiac index for calculating the body surface area? Tap overview Tap Patient Change index

  116. Slide 116: General: Alarms

    116 70% General: 17. Alarms IM-7300695-CP_C #EssenzEra #NewEraofPerfusion

  117. Slide 117: Video 17

    117 Video 17

  118. Slide 118: Alarms

    118 Alarms What are the 3 different alarm priorities and their respective color? High → red Medium → yellow Low → blue They are always listed in the order high, medium and low.

  119. Slide 119: Alarms

    119 Alarms Depending on the different criticality, the same type of alarm can have different priorities. High Example: depending on the Medium role of a pump, the cover alarm can have different priorities. Low

  120. Slide 120: Alarms

    120 Alarms Another example Bubble alarm for arterial line High Bubble alarm for CPL Medium The criticality depends on the level of risk for the patient.

  121. Slide 121: Alarms

    121 Alarms True or false? Information messages appear in grey and always at the bottom of the list. Every alarm has a timer that displays the duration of the alarm situation. You can mute the alarms for 5 minutes. Resolved alarms appear in the alarm history.

  122. Slide 122: Alarms

    122 Alarms ▪ Tip! (Low priority alarm tone) Behavior description Best practices For some seconds after the clearance of a high After clearing a high priority alarm, check the priority alarm the alarm tone of a low priority notification area for possible low priority alarm is not played if a low prio

  123. Slide 123: Alarms

    123 Alarms ▪ Tip! (Unjustified technical messages after a role change) Behavior description Best practices After performing a role change by connecting Check, acknowledge and ignore unjustified an additional pump or modifying the role of an technical messages. existing pump, unjustified technical me

  124. Slide 124: General: General settings

    124 80% General: 18. General settings IM-7300695-CP_C #EssenzEra #NewEraofPerfusion

  125. Slide 125: Video 18

    125 Video 18

  126. Slide 126: General settings

    126 General settings True or false? You can change the language, pressure unit, time, time format, date and date format in the general settings. You can adjust the volume of tones and the brightness. ESSENZ switches automatically to battery operation without stopping the pumps.

  127. Slide 127: General settings

    127 General settings Why do you need to perform a battery test? To extend the life of the battery, and to enable the calculation of the estimated, remaining run time of the machine during battery operation.

  128. Slide 128: General settings

    128 General settings ▪ Tip! (Reconfiguration after a software version update) Behavior description Best practices You can store general settings and customized Note your settings and profiles before the user configurations in profiles. If a software update. Reconfigure the HLM after the version upda

  129. Slide 129: General: Cleaning & disinfection

    129 88% General: 19. Cleaning & disinfection IM-7300695-CP_C #EssenzEra #NewEraofPerfusion

  130. Slide 130: Video 19

    130 Video 19

  131. Slide 131: Cleaning & disinfection

    131 Cleaning & disinfection What ingredients in the wipes are important when cleaning the HLM? Bleach Quarternary ammonium Soap Alcohol > 70%

  132. Slide 132: Cleaning & disinfection

    132 Cleaning & disinfection Put the 4 steps of cleaning the HLM in the correct order? Disconnect the HLM from the Power Supply and turn Disinfect all surfaces that have been touched with 1. it off. pre-soaked disinfectant wipes and dispose the wipes. Close the front and rear hatches, as well as the

  133. Slide 133: Cleaning & disinfection

    133 Cleaning & disinfection Never use water or pre-soaked wipes to clean electrical components such as connectors, plugs, electrical connection points, and all components inside the cabinet or the console.

  134. Slide 134: General: Troubleshooting

    134 95% General: 20. Troubleshooting IM-7300695-CP_C #EssenzEra #NewEraofPerfusion

  135. Slide 135: Video 20

    135 Video 20

  136. Slide 136: Congratulations!

    136 100% Congratulations! Please refer to the product Instruction For Use for additional information and instructions on the use of this product IM-7300695-CP_C #EssenzEra #NewEraofPerfusion

  137. Slide 137: Question & Answer

    137 Essenz @ Perfusion System Question&Answer

  138. Slide 138: Important Safety Information

    138 Important Safety Information Heart-Lung Machine Devices are intended to perform, control, monitor and support extracorporeal blood circulation replacing the mechanical pumping function of the heart, monitoring and regulating physiologic parameters during procedures requiring extracorporeal circu