Vaisala Oyj Class A — Technical Presentation (May 2018)

151 slides from Vaisala Oyj Class A's technical presentation, May 2018

  1. Slide 1: Humidity Theory: Understanding Humidity

    Humidity Theory: Understanding Humidity Jarkko Ruonala Product Manager, Industrial Instruments © Vaisala

  2. Slide 2: Humidity calculator link

    http://go.vaisala.com/humiditycalculator/5.0 © Vaisala

  3. Slide 3: Content

    Content  Theory, terms and definitions  Evaporation and condensation, equilibrium  Vapor pressure, saturation vapor pressure  relative humidity, dew point temperature, frost point temperature  Other humidity variables: Wet-bulb temperature, mixing ratio  Choosing the correct instrument © Vaisa

  4. Slide 4: How does humidity affect your day?

    How does humidity affect your day? © Vaisala

  5. Slide 5: Theory, terms and definitions

    Theory, terms and definitions

  6. Slide 6: Concepts

    Concepts Humidity Vapor pressure, partial pressure Evaporation, condensation, equilibrium Saturation vapor pressure Relative humidity Dew point temperature © Vaisala

  7. Slide 7: What is humidity

    What is humidity  Humidity is dissolved water  Humidity is in gas phase  Moisture - wetness of the solid material If you can see it, it is not humidity © Vaisala

  8. Slide 8: The water molecule

    The water molecule Polarity – uneven distribution of electrical charge, causing attraction between molecules © Vaisala

  9. Slide 9: Vapor pressure

    (0 PSIA) Vapor pressure  Molecules in gas phase are in constant random motion  This motion exerts pressure  The more molecules, higher the vapor pressure  The energy of the molecules depends not only on the amount, but also temperature © Vaisala

  10. Slide 10: Partial pressure of water

    Partial pressure of water  The total pressure is the sum of the partial pressures of the components in the gas (Dalton’s law)  The partial pressure exerted by water vapor is called the vapor pressure Pw 500 hPa Air + 500 hPa Water = 1000 hPa (14.5 PSIA) total Ptot = P1 + P2 + P3.. = Pair + Pwater

  11. Slide 11: Example: Water in a Closed Container

    Example: Water in a Closed Container  Evaporation: As a result of the random movement of molecules, some molecules have enough energy to escape through the water film  Condensation: As a result of random movement of molecules, some move so slow they get trapped back in water.  Equilibrium: With t

  12. Slide 12: The Effect of Temperature

    The Effect of Temperature Equilibrium Liquid phase Gas phase Notice the impact of temperature on equilibrium vapor pressure as it goes from 50 to 104 oF! © Vaisala

  13. Slide 13: Saturation vapor pressure

    Saturation vapor pressure Saturation vapor pressure is highly temperature dependent  The equilibrium of water vapor in a closed container is the saturation vapor pressure in that particular temperature  Condensation rate equals evaporation rate  Air cannot hold any more water vapor © Vaisala

  14. Slide 14: The humidity variables I know or use

    The humidity variables I know or use: A. Relative humidity B. Vapor pressure C. Dew-point/Frost-point temperature D. Mixing ratio E. Absolute humidity F. Wet-bulb temperature G. Water content [lb/MMscf, g/Nm3] H. PPM (parts per million) © Vaisala

  15. Slide 15: Relative Humidity

    Relative Humidity The proportion of water vapor pressure (Pw) to the maximum vapor pressure in the given temperature [Pws(t)] 𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉 𝑝𝑝𝑝𝑝𝑝𝑝𝑝𝑝𝑝𝑝𝑝𝑝𝑝𝑝𝑝𝑝 %RH = 𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆 𝑣𝑣𝑣𝑣𝑣𝑣𝑣𝑣𝑣𝑣 𝑝𝑝𝑝𝑝𝑝𝑝𝑝𝑝𝑝𝑝𝑝𝑝𝑝𝑝𝑝𝑝 𝑎𝑎𝑎𝑎…

  16. Slide 16: Example

    1013 Example  Let’s assume we have vapor pressure of 100 hPa ..And temperature 100 °C 100 RH = = ~10 % 1013 ..And temperature 60 °C 100 200 RH = = ~50 % 200 100 ..And temperature 45,8 °C 100 RH = = ~100 % 100

  17. Slide 17: Relative humidity is by definition relative to

    Relative humidity is by definition relative to A. Absolute amount of water in the air B. Temperature C. Total pressure of the system © Vaisala

  18. Slide 18: When temperature of a gas sample is increased

    When temperature of a gas sample is increased A. Relative humidity increases B. Dew point temperature decreases C. Relative humidity decreases © Vaisala

  19. Slide 19: Dew Point Temperature

    Dew Point Temperature  Dew point temperature is the temperature, where the vapor pressure would equal to saturation vapor pressure  This is the temperature, where the given humidity level condenses by effect of temperature  Proportional to amount of water vapor © Vaisala

  20. Slide 20: Dew point temperature only depends on

    Dew point temperature only depends on A. Amount of water vapor in the air B. Vapor pressure C. Temperature © Vaisala

  21. Slide 21: Impact of pressure

    Impact of pressure  Let’s assume a closed container with water vapor pressure of 10 hPa (0.145 PSIA) dew point temperature of 7 °C (44.6 °F) and total pressure of 1 bar (14.5 PSIA)  When we compress the gas isothermally to 4 bars (58 PSIA), the vapor pressure increases linearly  As a result, the

  22. Slide 22: Frost Point Temperature

    Frost Point Temperature  Frost point temperature is the temperature, where the water vapor condenses directly to solid ice, or frost  Relevant in dry conditions where saturation temperatures are well below freezing point © Vaisala

  23. Slide 23: Frost point temperature

    Frost point temperature Temperature -2°C (28.4°F)  Saturation vapor pressure over ice is slighlty lower than over supercooled water Pw = 5.3 hPa Pw = 5.1 hPa = 0.08 PSIA = 0.07 PSIA  Water molecules in ice form a rigid lattice  Lower rate of evaporation  Lower saturation vapor pressure  In most

  24. Slide 24: How much do frost point and dew point differ?

    How much do frost point and dew point differ?  The difference is ~10% of indication  Vaisala products output Tdf, which is dew point when saturation temperature is above freezing point, and frost point when saturation temperature is below freezing point © Vaisala

  25. Slide 25: When measuring dew point, what does not induce a change

    When measuring dew point, what does not induce a change in the reading: A. Changing pressure of the gas sample B. Changing temperature of the gas sample C. Changing amount of water vapor in the gas sample © Vaisala

  26. Slide 26: Wet-bulb temperature

    Wet-bulb temperature  Evaporation involves change in energy  Evaporation process involves cooling effect  Rate of evaporation is governed by relative humidity  Wet-bulb represents the lowest achievable temperature by means of evaporation of water  Dry- and wet-bulb temperatures are tabulated fo

  27. Slide 27: Mixing ratio

    Mixing ratio  Gives the mass of water vapor divided by total gas mass  Independent on temperature or pressure  Typical applications:  process control, drying  Mass balance calculations  Processes where pressures and temperatures vary from one stage to another Mixing ratio represents the mass f

  28. Slide 28: Ppm (Parts Per Million)

    Ppm (Parts Per Million)  Vaisala products output ppm –value per dry bases  Amount of water divided by amount of dry air  Volumetric (ppmv) and by-weight (ppmm) numbers available  Mostly used in trace humidity measurements  Note – exponential relation to wet basis ppm number (amount of water div

  29. Slide 29: Theory in practice

    Theory in practice

  30. Slide 30: Measurements under high humidity - Ceramics drying kiln

    Measurements under high humidity- Case example: Ceramics drying kiln © Vaisala

  31. Slide 31: Normal probe in high humidity application

    Normal probe in high humidity application (ceramics dryer) 100.0 Probe in saturation RH [%] without warmed probe 90.0 80.0 70.0 Relative humidity [%] 60.0 50.0 40.0 30.0 20.0 10.0 0.0 0.0 5.0 10.0 15.0 20.0 25.0 30.0 35.0 40.0 45.0 50.0 Time [h] © Vaisala

  32. Slide 32: Warmed Probe Technology

    Warmed Probe Technology  With additional T –probe; temperature elevated from ambient  Without additional T-probe (heated dew point measurement) seeks to 70..80 %RH Whole probe heated above ambient  Constant heating; affects the filter a must in a fuel-cell applications dew-point and T back-  XHE

  33. Slide 33: How does probe warming work?

    How does probe warming work? Tambient = 14 °C (57.2°F) RHambient = 97% RH Tambient = 14 °C (57.2°F) Tsensor = 16 °C (60.8°F) RHsensor = 83 % RH at 14°C = 97 %RH Probe warming resistor Page 34 / 5/7/2018 / name / Internal use / © Vaisala

  34. Slide 34: HMT337 warmed probe

    HMT337 warmed probe 100.0 RH [%] with warmed probe 90.0 80.0 70.0 60.0 Relative humidity [%] 50.0 40.0 - Condensation problem solved 30.0 - Better quality 20.0 - Better yield 10.0 0.0 0.0 5.0 10.0 15.0 20.0 25.0 30.0 35.0 40.0 45.0 50.0 Time [h] © Vaisala

  35. Slide 35: Practical control application in a ceramics dryer

    Practical control application in a ceramics dryer © Vaisala

  36. Slide 36: A customer case

    A customer case  Ceramics dryer Control variables: temperature and  Control options humidity mixing ratio  No control  Indirect control by feed t = 130..140 °C rate x = 100 g/kg (td = 53 °C)  Humidity based control RH = 5 % © Vaisala

  37. Slide 37: Another supplier said..

    Another supplier said..  Brand X distributors proposed solution: ”calibration at process temperature Another brand will provide the necessary accuracy”4 Calibration at 70 °C reference [%RH] Deviation from Brand X hysteresis 2 1.50 1.00 0.50 0 d %RH 1,5 %RH 0.00 0 50 100 0 20 40 60 80 100 -0.50 -2 -

  38. Slide 38: Accuracy of control

    Accuracy of control  T = 130 °C  RH = 5 % +/- 3%  Brand X specified repeatability short-term: 0,5 %RH  Brand X hysteresis: 1,5 %RH RH inaccuracy  Brand X estimated drift in 1 year: 2 %RH multiplies in high x = 100 +/- 80 g/kg Another brand temperatures for calculated variables  DMT345 specifie

  39. Slide 39: DMT345 for high temperatures

    DMT345 for high temperatures  DRYCAP180S sensor  High dew point measurements up to 140 °C dry-bulb  Low dew-point applications (ovens and similar) up to 100 °C dry-bulb temperature  Installation flange and long probe provide easy installation through insulation materials © Vaisala

  40. Slide 40: What should be considered when measuring elevated or variable temperatures?

    What should be considered when measuring elevated or variable temperatures?  What is the maximum temperature at the measurement position?  Is the instrument specified for accuracy in my temperature range?  What is the instrument’s accuracy in terms of the engineering unit I am using? © Vaisala

  41. Slide 41: Cooling towers and wet-bulb temperature

    Cooling towers and wet-bulb temperature © Vaisala

  42. Slide 42: Practical issues

    Practical issues Temperature  Maintenance Cooling water in  Scaling – Reduced performance R – Make-up water A N  Pumps and fans G E  Control  Temperature set-point Cooling water out A – Generally optimal 2..3 °C above P P the wet-bulb temperature R O – Too small approach leads to A C excessive

  43. Slide 43: Unshielded installation with low quality sensor

    Unshielded installation with low quality sensor Radiated heat 75 %RH calibration after outdoor testing +3 °C (5.4 °F) Wet-bulb temperature +2 °C (+3.6 °F) Drift of the Impossible to sensor control accurately +/- 5 %RH for optimizing approach (2..3 °C) © Vaisala

  44. Slide 44: A good installation with a high-quality sensor

    A good installation with a high-quality sensor Radiated heat +0.2 °C Wet-bulb temperature +/- 0.3 °C Drift of the RH sensor +/- 2 %RH White top surface on each plat reflects heat Black bottom surface on each plate effectively emits any absorbed heat © Vaisala

  45. Slide 45: Theoretical low limit of beverage's temperature

    In the set-up below, the theoretical low limit of beverage’s temperature is (use humidity calculator) A. 25.0 °C (77.0 °F) B. 20.1 °C (68.2 °F) C. 14.5 °C (58.1 °F) D. Don’t know, it’s gone already! T = 25 °C RH = 30 % © Vaisala

  46. Slide 46: End of part 1

    End of part 1 © Vaisala

  47. Slide 47: Humidity measurement technologies

    Humidity measurement technologies

  48. Slide 48: History of Humidity Measurement

    History of Humidity Measurement © Vaisala

  49. Slide 49: Comparison of traditional technologies

    Comparison of traditional technologies Mechanical Psychrometer Chilled mirror Resistive Range 10 % 15 % -95 .. -20 °C Tf 5 % minimum Range max 100 % 100 % < 100 °C Td 100 % Accuracy typ. +/- 5% +/-2..5 % ~1 %RH +/- 5 % Hysteresis Very high None None High Stability Very poor Poor Very good Poor Maint

  50. Slide 50: Performance

    Performance  Chilled mirror hygrometer  HMP7 Relative Humidity and Temperature  Laboratory reference etc. probe with capacitive HUMICAP sensor  Field instrument for indudustrial use  Requires maintenance  No regular maintenance need  Accuracy: +/- 0.1 °C in dew point  Accuracy: +/- 0.8 %RH a

  51. Slide 51: Radio sonde

    Radio sonde  Measurements in atmosphere up to 30 km above sea level with 5 m/s vertical speed  RH range 0..100 %RH  Rapid changes up and down  Temperature +60..-90 °C  Cloud layer © Vaisala

  52. Slide 52: Vaisala innovation: Capacitive thin-film polymer sensor

    Vaisala innovation: Capacitive thin-film polymer sensor  Vaisala founded by prof. Vilho Väisälä in 1940’s  Humicap invented in 1973  For radiosonde application  Current de facto technology  Later on productized for industrial applications [Name] 5/7/2018 © Vaisala 6

  53. Slide 53: Vaisala HUMICAP

    Vaisala HUMICAP® The sensor capacitance responds directly to changes in relative humidity Operating range: Relative humidity 0..100 %RH Temperature -40..+180 °C Withstands chemicals Withstands wetting Fast response time Low hysteresis © Vaisala

  54. Slide 54: Theory of operation

    Theory of operation Overlap © Vaisala

  55. Slide 55: HUMICAP sensor is sensitive to changes in

    HUMICAP sensor is sensitive to changes in A. Dew point temperature B. Vapor pressure C. Relative humidity © Vaisala

  56. Slide 56: HUMICAP vs DRYCAP

    HUMICAP® vs DRYCAP ®  Capacitive humidity sensors are sensitive to relative humidity as the primary physical variable  HUMICAP® withstands continuous high humidity and immersion 0..100 %RH  DRYCAP ® is optimized for low humidity measurements, generally for < 10 or 5 %RH  In general, when frost p

  57. Slide 57: HUMICAP sensor types

    HUMICAP ® sensor types  HUMICAP180  HUMICAP180R Composite temperature sensor  HUMICAP180L2 for chemical purge  HUMICAP180V C for Composite Humicap180RC Humicap180C  HUMICAP R2 and R2C Humicap180VC © Vaisala

  58. Slide 58: Which sensor should I use?

    Which sensor should I use?  HUMICAP180R  HUMICAP R2  R for Robust  General purpose  General purpose  Latest generation, like Humicap180R but with improved corrosion  HUMICAP180 resistance  Previous generation  Faster response time under low temperatures  HUMICAP180V  For applications wher

  59. Slide 59: HUMICAP robustness

    HUMICAP® robustness © Vaisala

  60. Slide 60: Sensor purge

    Sensor purge Sensor purge Saturated Ethylene Glycol test 16  Regeneration of the 14 Deviation from reference [%RH] sensing element by 12 brief heating 10 8  Evaporates volatile 6 contaminants from the 4 active polymer 2 0 0 5 10 15 20 25 30 -2 Exposure in days [Name] 5/7/2018 © Vaisala 14

  61. Slide 61: HUMICAP stability

    HUMICAP® stability 2  Three Humicap180R sensors in 1.5 outdoor test 1 At 75 %RH 0.5 Drift [%RH]  Drift less than 1 %RH over full 0 At 0 %RH range -0.5 -1  1000 days test -1.5 -2 0 200 400 600 800 1000 1200 Time [days] Three sensors in outdoor test located in Southern Finland © Vaisala

  62. Slide 62: Full Temperature Range Linearity

    Full Temperature Range Linearity Calibration at 70 °C 3.5 Deviation from reference [%RH] 2.5 1.5 0.5 -0.5 0 10 20 30 40 50 60 70 80 90 -1.5 -2.5 -3.5 Reference %RH © Vaisala

  63. Slide 63: Response time

    Response time  The response time of the sensor is ~15 s in room temperature and still air.  Diffusion – temperature  Gas exchange – flow  Temperature – thermal mass Choice of filter affects to the thermal mass, gas exchange and response time Diffusion time multiplies by factor of ~2 for every 10

  64. Slide 64: End of part 2

    End of part 2 © Vaisala

  65. Slide 65: Calibration and traceability

    Calibration and traceability © Vaisala

  66. Slide 66: Why knowing calibration, uncertainty and traceability is important?

    Why knowing calibration, measurement uncertainty and traceability is important?  Uncertainty is numerical quality of a measurement, with respect to the international measurement system  Traceability means the documented path from the instrument to the international measurement system Measurements

  67. Slide 67: I take care of calibrations by

    I take care of calibrations by A. Not applicable B. My equipment is not calibrated C. Own calibration laboratory D. Third-party calibration laboratory E. Field-checks F. Other © Vaisala

  68. Slide 68: Traceability and calibration

    Traceability and calibration © Vaisala

  69. Slide 69: Traceability

    Traceability “Property of a SI measurement result BIPM Primary whereby the result can be reference National related to a reference Secondary laboratories through a documented reference Accredited laboratories unbroken chain of Reference standards Working standards calibrations, each End users contri

  70. Slide 70: International System of Units: SI

    International System of Units: SI Système international d’unités SI base units Length 1 metre Mass 1 kilogram Time 1 second Electric current 1 Ampere Temperature 1 Kelvin Amount of substance 1 mole Luminous intensity 1 candela Measurement standard, in which all traceable measurements can be traced [

  71. Slide 71: Traceability of Vaisala RH instrument

    Simplified example: Traceability of Vaisala RH instrument SI units National laboratory NIST MIKES MIKES Accredited Accredited laboratory laboratory Primary standard Triple-point cell of Pressure primary Vaisala SPRT reference water standard Measurement Standards Working standard Laboratory PRT refer

  72. Slide 72: What is calibration?

    What is calibration?  Calibration means comparing the instrument reading with a reference (a calibration standard)  Quality of calibration is defined by the following  The competence of the laboratory  The management system of the laboratory By definition, every measurement has  The methods use

  73. Slide 73: What is adjustment?

    What is adjustment? Linear adjustment  Adjustment is correcting the (offset and gain) instrument to minimize deviation from the calibration reference Gain  Normal service procedure is calibration (”as-found”) and adjustment + calibration (”as-left”)  This is sometimes called Offset ”Calibration”

  74. Slide 74: Types of adjustment: Offset

    Types of adjustment: Offset Offset © Vaisala

  75. Slide 75: Types of adjustment: Linear adjustment

    Types of adjustment: Linear adjustment Gain Offset © Vaisala

  76. Slide 76: Multi-point correction

    Multi-point correction Necessary when sensor is non- linear and further linearization is Gain Gain Offset Offset required Gain Offset Gain Offset © Vaisala

  77. Slide 77: Things to ask

    Things to ask What does your quality system require? What does your application require? [Name] 5/7/2018 © Vaisala 13

  78. Slide 78: Quality Management System

    Quality Management System  ISO9001: Quality Management Systems – Requirements  ISO9001 clause 7.6: Control of measurement and monitoring equipment  ISO10012:2003: Measurement management systems - Requirements for measurement processes and measuring equipment  ISO17025:2005: General requirements

  79. Slide 79: How often to calibrate?

    How often to calibrate? Calibration interval is subject to the operating conditions  Humidity/temperature  Chemical contaminants As-found certificates help in determining suitable calibration interval [Name] 5/7/2018 © Vaisala 15

  80. Slide 80: What options do I have for calibration?

    What options do I have for calibration?  Field calibration  Reference instrument  Portable calibrator  Laboratory calibration Quality of  In-house calibration:  Service • Competences • Management system • Methods  Factory service • Equipment and the references [Name] 5/7/2018 © Vaisala 16

  81. Slide 81: Calibration interval is defined by

    Calibration interval is defined by A. The manufacturer of the instrument B. The quality policy C. Calibration laboratory’s recommendation © Vaisala

  82. Slide 82: Options for calibration: Field Calibration

    Options for calibration: Field Calibration © Vaisala

  83. Slide 83: Single point or multi-point

    Single point or multi-point Single-point Multi-point  When stable condition  When relative humidity  When not close to changing on wide range extremes of  When measuring close to measurement range extremes of the measurement range [Name] 5/7/2018 © Vaisala 19

  84. Slide 84: Field calibration references

    Field calibration references Handheld Humidity Single-point HMK15 Salt generator calibrator Multi-point Multi-point [Name] 5/7/2018 © Vaisala 20

  85. Slide 85: HMK15 salt calibrator

    HMK15 salt calibrator  Intrinsic reference based on known property of saturated salt solutions generating known relative humidity levels  Uncertainty 1.2..2 %RH  Batch traceable Salt Equilibrium RH at 20 °C Applicable standards: LiCl 11.3 %RH ASTM E104-85 DIN 5008 MgCl 33.1 %RH JIS Z8806 NaCl 75.

  86. Slide 86: HMK15 salt calibrator

    HMK15 salt calibrator  A reference instrument can be used as a reference for additional accuracy or traceability reasons Common mistakes: Insufficient stabilization time Temperature gradients Adjustment in too narrow RH range [Name] 5/7/2018 © Vaisala 22

  87. Slide 87: Options for calibration: Laboratory

    Options for calibration: Laboratory calibration © Vaisala

  88. Slide 88: Laboratory calibration

    Laboratory calibration  Reduces uncertainty components from calibration  Constant and controlled environment  Multi-point calibration possible  For wide-span applications and validating linearity  Special calibrations  RH calibrations under different temperatures  Accredited calibrations © Va

  89. Slide 89: Accredited calibration?

    Accredited calibration?  Accredited calibration laboratory is running a Quality Measurement System in compliance to ISO/IEC 17025, ANSI/NCSL Z540 or other standard  Management system  Technical requirements • Professionally made and validated uncertainty calculations • Verified traceability to in

  90. Slide 90: What Type of Calibration Do I Need?

    What Type of Calibration Do I Need? Does your company’s Quality Management Standards (QMS) require an 1SO17025 Accredited Calibration? No Yes 1 don’t know ‘Will you need to defend or demonstrate proper calibration procedures to <— Yes assure the quality of test results? Accredited Calibration No Do

  91. Slide 91: Which option works best for you?

    Which option works best for you? • QMS requirements • Number of points needed • Humidity  Field calibration range  Factory calibration  Own staff or sourced • Temperature  Highest quality references  In-house laboratory range  Calibration + Adjustment  Manage your own lab, training, standards

  92. Slide 92: Case example

    Case example © Vaisala

  93. Slide 93: Case example: spot-check calibration: In spec or not?

    Case example: ”spot-check” calibration: In spec or not? Field calibration (spot check) of a wall-mount RH/T transmitter  Device under test (DUT):  Temperature [°C]: 22.1 (71.8°F)  Relative Humidity [%RH]: 57.6  Reference instrument:  Temperature [°C]: 21.87 (71.37°F)  Relative Humidity [%RH]:

  94. Slide 94: Should it be adjusted?

    DUT Reference Temp [°C/°F] 22.1 / 71.8 21.87 / 71.37 RH [%RH] 57.6 55.10 Should it be adjusted? A.Yes B.No © Vaisala

  95. Slide 95: Steps of uncertainty estimation

    Steps of uncertainty estimation  Recognize the relevant uncertainty components  Estimate the standard uncertainty of each component  Express the model equation  Calculate the sensitivity coefficients  Calculate the expanded uncertainty © Vaisala

  96. Slide 96: Uncertainty components of temperature calibration

    Uncertainty components of temperature calibration  What are the uncertainty components?  Related to the reference?  Calibration correction  Calibration uncertainty  Stability  Resolution  Related to the device under test?  Resolution  Related to the set-up?  Humidity spatial distribution 

  97. Slide 97: Source of information: Datasheet

    Source of information: Datasheet © Vaisala

  98. Slide 98: Uncertainty of temperature?

    Uncertainty of temperature?  Reference: Vaisala HM70 handheld  Accuracy: 0.1 °C  Calibration correction: -0.01 °C  Calibration uncertainty: 0.1 °C  Stability: 0.05 °C  Resolution: 0.01 °C  Set-up and DUT? RH/T sensor  Temperature spatial: 0.5 °C location  Temperature temporal: 0.1 °C  Temp

  99. Slide 99: Model equation and uncertainty

    Model equation and uncertainty  Correction = Reference value – Indication of DUT  Uncertainty = Uncertainty of the reference + Uncertainty of the DUT  Uncertainties sum up quadratically  Distribution of the uncertainty varies. Different shapes of distribution are scaled to normal distribution eq

  100. Slide 100: Uncertainty budget for temperature

    Uncertainty budget for temperature © Vaisala

  101. Slide 101: Do we need to adjust for temperature?

    Do we need to adjust for temperature? Influence on total calibration uncertainty 24 0.350 23.5 0.300 0.250 23 0.200 22.5 0.150 0.100 22 0.050 0.000 21.5 21 20.5 20 Ref T [°C] DUT T [°C] © Vaisala

  102. Slide 102: Uncertainty of Relative Humidity calibration

    Uncertainty of Relative Humidity calibration  Reference: Vaisala HM70 handheld  Accuracy: 1.0 %RH  Calibration correction: -0.6 %RH  Calibration uncertainty: 0.6 %RH  Resolution: 0.1 %RH  Long-term stability: 1.0 %RH  Set-up and DUT?  Temperature equilibrium: 0.6 °C  RH difference spatial/t

  103. Slide 103: RH sensitivity coefficient for temperature

    RH sensitivity coefficient for temperature  Relative Humidity has non-linear dependency on temperature  The magnitude of the dependency differs with the RH and temperature condition  Sensitivity coefficient tells how the uncertainty in a variable is reflected to the uncertainty of another variabl

  104. Slide 104: Uncertainty budget for Relative Humidity

    Uncertainty budget for Relative Humidity © Vaisala

  105. Slide 105: Should we adjust for Relative Humidity?

    Should we adjust for Relative Humidity? Chart Title Influence on total calibration uncertainty 60 0.700 0.600 58 0.500 56 0.400 54 0.300 52 0.200 0.100 50 0.000 Reference Uncetainty in Specified Calibration Microclimate RH Ref resolution DUT 48 drift temperatures accuracy uncertainty stabilization r

  106. Slide 106: What uncertainty to expect with RH calibration?

    What uncertainty to expect with RH calibration?  Field check: +/- 2..5 %RH  HMK15: +/- 1.2..2 %RH  NIST: Hybrid humidity generator: +/- 1 %  Vaisala factory calibration: +/- 0.6.. 1.0 % © Vaisala

  107. Slide 107: End of part 3

    End of part 3 © Vaisala

  108. Slide 108: Dew point in compressed air

    Dew point in compressed air © Vaisala

  109. Slide 109: Compressed Air

    Compressed Air Compressed air is often called the fourth utility in industry, right behind water, electricity and natural gas Where is it measured?  Energy storage:  Instrument air for tools, robots, pneumatic actuators..  Trains: brakes, couplings, pneumatic doors..  Vortex cooling  Conveyor m

  110. Slide 110: Why Measure Dew Point?

    Why Measure Dew Point?  Dryer performance monitoring  Condition monitoring  Dryer switching endpoint  Time controlled vs dew point demand controlled  Efficiency  Compressed air quality at point of use  Quality assurrance 5/7/2018 © Vaisala 3

  111. Slide 111: When Dew Point, When Relative Humidity?

    When Dew Point, When Relative Humidity? When RH is ~5 %RH ±2 °C Td = ±1 %RH ±5.6°F Td = ±1 %RH When RH is ~1 %RH ±2 °C Td = ±0.2 %RH ±5.6°F Td = ±0.2 %RH © Vaisala

  112. Slide 112: How dew point sensor works?

    How dew point sensor works?  All capacitive and resistive solid state sensors are based on equilibration of humidity between the surroundings and the sensor  Similar operation with humidity sensors  When relative humidity and temperature are known, the dew point temperature can be calculated © Va

  113. Slide 113: Vaisala dew point technology

    Vaisala dew point technology  DRYCAP sensor is optimized for sensitivity in low humidity ranges by a special engineered polymer  Sensor purge function always included  Unique auto-calibration function  Sensor responds to very small changes in relative humidity  Sensitivity combined with robustn

  114. Slide 114: Vaisala DRYCAP Auto-calibration technology

    Vaisala DRYCAP® Auto-calibration technology 1. Auto-calibration start: Sensor Sensor warmed temperature RH value decreases when T increases Sensor 2. Sensor starts to cool down after humidity short warming. time RH reading starts to increase. 1. 2. 5. 6. 3. 3. RH and T values logged during humidity

  115. Slide 115: DRYCAP sensor for low humidity

    DRYCAP sensor for low humidity  Tolerates wetting, recovers perfectly  Tolerates contaminants: Compressor oils, volatile organics  Fast response time  No drifting – recommended calibration interval 2 years  Small hysteresis  ISO9001 calibrated at factory © Vaisala

  116. Slide 116: Compressed Air Systems

    Compressed Air Systems Basics: The air contains water as water vapor. The water may cause problems in the compressed air system – condensation and corrosion in the distribution system and quality problems at the point of use Challenge: Energy efficient control of dryers requires measurement with com

  117. Slide 117: Compressed Air: Regenerating Drying

    Compressed Air REGENERATING DRYING © Vaisala 10

  118. Slide 118: Compressed Air: Drying Regenerating

    Compressed Air DRYING REGENERATING © Vaisala 11

  119. Slide 119: Analyzing correct operation of the dryer by dew point measurement

    Analyzing correct operation of the dryer by dew point measurement 10 10 Dew point temperature [°C] 0 0 Dew point temperature [°C] -10 -10 Input -20 -20 Input Output Output -30 -30 -40 -40 -50 -50  Dryer operates normally  Dryer is loaded too early  Regeneration cycles  Air leak into the dryer 

  120. Slide 120: Desiccant dryer condition monitoring

    Desiccant dryer condition monitoring 10 10 0 0 Dew point temperature [°C] Dew point temperature [°C] Input Input -10 -10 Output Output -20 -20 -30 -30 -40 -40 -50 -50  Dryer does not work  Regenerated dryer does not work  Regeneration does not work  Insufficient cooling time  Severely contamina

  121. Slide 121: Process Pressure vs. Ambient

    Process Pressure vs. Ambient  Measurement process at pressure  Measurement at ambient pressure  + typically true specified dew point  + line pressure fluctuation has no effect  Value that user wants to know e.g. in dryers  + easy to remove instrument for service  + no sampling system needed 

  122. Slide 122: Sampling Cells

    Sampling Cells DMT242SC • G3/8”, G1/4” ISO (inlet, outlet) • G1/2” for probe DMT242SC2 • Like above, but with 1/4” swagelok fittings DSC74C • For pressurized air, quick connector and leak screw • Pressurized measurement DSC74C • For measurement in atmospheric pressure • Fixed leak screw 3 ..1 0 bar

  123. Slide 123: Portable Dewpoint Vaisala DSS70A

    Portable Dewpoint Vaisala DSS70A 5/7/2018 © Vaisala 16

  124. Slide 124: Requirements for Accurate Switchpoint Control?

    Requirements for Accurate Switchpoint Control? Stability at -10°C / 14°F frost point Stability at -40°C/-40°F frost point © Vaisala

  125. Slide 125: Dew point error at constant temperature

    Dew point error at constant temperature Overshoot 15 10 5 Vaisala DRYCAP180U: DMT152 Dew point error, °C Sensor B Sensor C 0 Sensor D Vaisala DRYCAP180D: DMT143 Sensor F -5 Sensor G Sensor H -10 -15 -76 -66 -56 -46 -36 -26 Reference dew point, °C Undershoot 5/7/2018 © Vaisala 18

  126. Slide 126: Temperature dependency at -50°C dew point level

    Temperature dependency at -50°C / -58 °F dew point level 15 10 Vaisala DRYCAP180U: DMT152 5 Sensor B Dew point error, °C Sensor C 0 Sensor D Sensor E Sensor F -5 Vaisala DRYCAP180D: DMT143 Sensor H -10 Sensor I Sensor J Sensor K -15 -20 -10 0 10 20 30 40 50 Temperature of environment, °C MIKKA 5/7/2

  127. Slide 127: Repeatability

    Repeatability 2.00 Vaisala Vaisala Ultra DryDRYCAP180U: 1.50 DMT152 Sensor B Michell Ultra Dry 1.00 Sensor C GE Pana- metrics Dew point error, °C 0.50 Sensor D Cosa Xentaur 0.00 Sensor E Alpha Moisture -0.50 Sensor F TEKHNE -1.00 Vaisala Vaisala StandardDRYCAP180M: Dry -1.50 DMT143 Michell Standard

  128. Slide 128: Short-term stability

    Short-term stability -44 Vaisala DRYCAP180U: DMT152 -46 Sensor B -48 Sensor C -50 Sensor D -52 Sensor E -54 Vaisala DRYCAP180M: DMT143 -56 Sensor G -58 Sensor H -60 0 12 24 36 48 60 72 84 time, h 5/7/2018 © Vaisala 21

  129. Slide 129: End of part 4

    End of part 4 © Vaisala

  130. Slide 130: Installation and Good Measurement Practices

    Installation and Good Measurement Practices © Vaisala

  131. Slide 131: How do you decide what instrument to purchase?

    How do you decide what instrument to purchase? (choose top 3 in order) A. Referral from a colleague B. Web research C. Magazines D. Vendor websites E. Vendor salesperson © Vaisala

  132. Slide 132: Follow the Manufacturer's Instructions

    Follow the Manufacturer's Instructions  Read the Manual  Use the Technical Support  Use the Application Engineers  Use the Sales Engineers © Vaisala

  133. Slide 133: Proper installation

    Follow the Manufacturer's Instructions Proper installation  Must be mounted in area where air or gas is representative of the process or environment  Air should flow around or past the sensor – good air flow is an advantage; it ensures that the probe and ambient air are at the same temperature & e

  134. Slide 134: Probe installation

    Follow the Manufacturer's Instructions Probe installation  horizontal is preferred; bend in the cable  allow for temperature and pressure fittings  use insulation and sample line heating  ensure representative sample of air with flow preferred to be sealed to be insulated © Vaisala

  135. Slide 135: Ensure a proper calibration

    Ensure a proper calibration  Is the calibration within the recommended interval?  Is it traceable?  Is there a valid certificate available?  Does the calibration comply with your requirements? © Vaisala

  136. Slide 136: Keep records

    Keep records  Record of calibration  Record of adjustments  Record of repairs  Out of tolerance results  Record of any damage to the instrument © Vaisala

  137. Slide 137: Spot-check performance

    Spot-check performance  Check at intervals between calibrations  Check before and after any event that may stress the instrument  Transportation  Packaging or shipping  Check against one or, even better, two other instruments © Vaisala

  138. Slide 138: Be knowledgeable about the parameter

    Be knowledgeable about the parameter a x ppm Td Tw h RH  Know what parameter you are measuring and be aware of the potential effects of the environment around the point of measure.  sensor type  gas effect – CO2, pure hydrogen, other high concentrations of gas  temperature & pressure effect  is

  139. Slide 139: Be knowledgeable about the parameter

    Be knowledgeable about the parameter Polymer & organic sensors measure RH 1200 1000 Saturation Curve Pw 800 Pw (mbar) %RH = 100 x P ws (t) 600 400 200 0 0 10 20 30 40 50 60 70 80 90 100 Temperature (°C) temperature – temperature - temperature - temperature Note: at 20°C a temperature difference of j

  140. Slide 140: Be knowledgeable about the parameter

    Be knowledgeable about the parameter How is the parameter determined?  Is it measured directly or calculated?  Does the calculation require pressure input?  Is the calculation dependent on temperature?  Dew point or Frost point? (for example)  Pressure dew point or atmospheric dewpoint?  Mixin

  141. Slide 141: Be aware of potential external factors

    Be aware of potential external factors • Do not introduce external factors that will affect the measurement – direct sunlight – body heat and humidity – non-representative sources of heat – stagnant and/or non representative air samples – temperature leakage © Vaisala

  142. Slide 142: Be aware of potential external factors

    Be aware of potential external factors © Vaisala

  143. Slide 143: Anomalous Sources of Heat or Cold

    Be aware of potential external factors Anomalous Sources of Heat or Cold 77F 50F © Vaisala

  144. Slide 144: Temperature Leakage

    Be aware of potential external factors Temperature Leakage • If the temperature of the process is considerably higher than that of the environment, the whole probe and preferably a part of the cable must be in the process [adds 3.2% error] 80°C / 176°F 70 %RH 22°C / 72°F 78°/172° © Vaisala

  145. Slide 145: Temperature leakage

    Be aware of potential external factors Temperature leakage – Less critical, but enough difference where you want the whole portion of the metal probe inside the process and the cable sealed. 22°C 28°C 72°F 82°F © Vaisala

  146. Slide 146: Temperature leakage

    Be aware of potential external factors Temperature leakage – Very large differences require heavy insulation 180C/356F 85%RH © Vaisala

  147. Slide 147: Temperature

    Be aware of potential external factors Temperature – From room to high temperature high humidity chamber Testing chamber 70°F ambient temp 104ºF 60% RH Td = 87ºF What is the dew point? You must PRE-HEAT the probe!! © Vaisala

  148. Slide 148: Water Vapor Pressure

    Be aware of potential external factors Water Vapor Pressure Testing chamber 80°F @ 35%RH 120ºF What is Td? 50F Td = -76ºF What is ppmv? 12,000 ppmv ppmv=11 What is the ppmv? © Vaisala

  149. Slide 149: Summary – Good Measurement Practice 7 Points

    Summary – Good Measurement Practice  7 Points 1. Choose the correct instrument 2. Follow the manufacturers instructions 3. Ensure a proper calibration 4. Keep records 5. ‘Spot-check’ performance 6. Be knowledgeable about the parameter 7. Be aware of potential external factors © Vaisala

  150. Slide 150: Was this Measurement Practice section beneficial?

    Was this Measurement Practice section beneficial? A. Very beneficial B. Somewhat beneficial C. Neutral D. Minimally beneficial E. Not at all beneficial © Vaisala

  151. Slide 151: End of part 5

    End of part 5 © Vaisala