Horiba, Ltd. · TSE:6856

Horiba — Innopharma Technology — Technical Presentation (March 2020)

42 slides from Horiba's technical presentation, Mar 24, 2020

  1. Slide 1: The Value of Real-time Imaging: Integrating Particle Size Analysis

    The Value of Real-time Imaging: Integrating Particle Size Analysis onto Fluid Beds, Twin Screw Granulators and Roller Compactors Darren McHugh & Chris O’Callaghan – Innopharma Technology 2020-03-24 ©2020 Innopharma Technology Ltd. 1

  2. Slide 2: Overview

    Overview • Innopharma Introduction • Eyecon2 – Direct Imaging System for In-line Particle Size Measurement • Practical Considerations for Implementation • or Interface with Process Equipment • Application • TSG, Milling, FBG • Deep Dive – Wuster Coating the Real-Time Prediction of Polymer-Coated Mul

  3. Slide 3: Innopharma Company Background

    Innopharma Company Background • Founded in 2009 • Three divisions: • Education & Upskilling • Technology to Enable Advanced Manufacturing/Process Analytical Technology • Technical Services • Currently ~60 employees experienced in STEM, Pharma development and manufacturing operations, IT & Software D

  4. Slide 4: Innopharma Technology - Our Products

    Innopharma Technology - Our Products Direct Imaging Particle Analyser Multi-point NIR Spectrometer Vertically integrated platform for Smart Process development and Manufacture • Particle analyser for powders and bulk solids • Near infrared spectrophotometer for measuring • Detect Fluid bed Pellet (W

  5. Slide 5: Journey of PAT, Sensors & Advanced Manufacturing Platforms

    Journey of PAT, Sensors & Advanced Manufacturing Platforms PAT, Sensors and Platforms for Advanced Manufacturing 2009 2010 2011 2012 2013 2014 2015 2016 2017 2018 2019 2020 Sensors Development of sensors for solids processing Eyecon in-line real time PSD Eyecon2 second generation PSD Multieye in-lin

  6. Slide 6: Particle Size Analyser: Eyecon2

    Particle Size Analyser: Eyecon2 • Real-Time particle Size Distribution and shape • Use in: • Research & development (QbD/DoE/CPP/CQA) • Scale up • Tech transfer • Manufacturing • Batch • Continuous • Use on: • Fluidised Bed Coating, Granulation, Drying • Twin Screw Granulation • Roller Compaction/Mi

  7. Slide 7: Eyecon2 Technical Specifications

    Eyecon2 Technical Specifications Size Range 50 to 5500 µm Casing materials 304 Stainless Steel, Glass, Silicon (gaskets) Imaging Area 11.25 x 11.25 mm Output PDF session report. CSV, full PSD from D5-D95. JPEG (images) Instrument Ratings GMP Compliant Design EyePASS is both 21 CFR part 11 & GAMP5 Co

  8. Slide 8: Device Overview Video

    Device Overview Video ©2020, Innopharma Technology Ltd 8

  9. Slide 9: Benchtop

    Benchtop ©2020, Innopharma Technology Ltd 9

  10. Slide 10: Inline

    Inline ©2020, Innopharma Technology Ltd 10

  11. Slide 11: Method of Operation: Image Capture

    Method of Operation: Image Capture • A flash-imaging technique is used with an extremely short light-pulse to illuminate moving particles for image capture • Red, Green and Blue LEDs illuminating the sample from different angles for accurate detection of particle boundaries ©2020, Innopharma Technol

  12. Slide 12: Method of Operation: Image Analysis

    Method of Operation: Image Analysis • Each particle initially identified • Best-fit ellipse calculated • Major & minor diameters computed • PSD/D-values determined ©2020, Innopharma Technology Ltd 12

  13. Slide 13: Particle Size

    Particle Size • The D-values are computed from the group of ellipses estimated from the particles • D50 value, also known as mass-median- D10 diameter (MMD) is the diameter which divides the particles into two groups with 10 % Weight equivalent weight / mass. 90 % Weight D50 • Similarly, the mass of

  14. Slide 14: Eyecon2 Data Output

    Eyecon2 Data Output ©2020, Innopharma Technology Ltd 14

  15. Slide 15: Geometry of Illumination

    Geometry of Illumination ©2020, Innopharma Technology Ltd 15

  16. Slide 16: Geometry of Illumination

    Geometry of Illumination ©2020, Innopharma Technology Ltd 16

  17. Slide 17: Illumination Calculator

    Illumination Calculator ©2020, Innopharma Technology Ltd 17

  18. Slide 18: Presenting Particles for Imaging: 1

    Presenting Particles for Imaging: 1 • Particles imaged directly behind surface of window • Applicable with quasi-static bodies of material e.g. fluid bed, and with flowing materials • Window helps to ensure particles are optimally presented within the depth of field Challenges • Wide PSD – smaller p

  19. Slide 19: Presenting Particles for Imaging: 2

    Presenting Particles for Imaging: 2 • Particles imaged flowing between window & chute (typically stainless steel) • Applicable only to flowing material • Useful with wetter / stickier materials as backing surface can have greater resistance to fouling than window Challenges • Constraining flow withi

  20. Slide 20: Equipment Integration Solutions

    Equipment Integration Solutions • What to consider • Presentation • Representation • Maintain consistent presentation of material • For FB position below material bed level • For TSG image onto backing surface • Maximise number of particles captured per image • Optimise positioning in focal plane •

  21. Slide 21: Fluid Bed Interfaces

    Fluid Bed Interfaces Lab Pilot Manufacturing ©2020, Innopharma Technology Ltd 21

  22. Slide 22: Interface Examples

    Interface Examples Develop Scale Up Manufacturing Dev. Scale Conti. Twin-Screw Drier Line Granulation Outlet Roller Filling Milling Compaction ©2020, Innopharma Technology Ltd 22

  23. Slide 23: Fouling

    Fouling • Fouling Control • Prevent the ingress of the fouling material • Low-fouling surfaces (for example, very smooth, implanted with ions, or of low surface energy like Teflon) are an option for some applications. • Anti-static • Sapphire (Low coefficient of friction) • Purge valves • The condit

  24. Slide 24: Fouling Control

    Fouling Control Issue Solution Issue Solution ©2020, Innopharma Technology Ltd 24

  25. Slide 25: Eyecon2 In-Line Application Examples

    Eyecon2 In-Line Application Examples Chris O’Callaghan, Head of Engineering ocallaghanc@innopharmalabs.com ©2020 Innopharma Technology Ltd. Confidential 25

  26. Slide 26: Twin-Screw Wet Granulation

    Twin-Screw Wet Granulation 2000 1800 1600 • Continuous granulation – measurement of particle size Dv50 (µm) throughput 1400 (kg/h) wet particles directly at outlet 1200 2 1000 • Polished, heated chute to reduce sticking 800 3 • Quick DoEs: 5~7 minutes per experiment 600 400 u = 500 rpm xL/S = 20 % •

  27. Slide 27: Milling

    Milling • HME, pelletisation and milling DoE (Hosokawa Alpine 100) • Varied mesh size and RPM • Aim to determine optimum parameter setpoints to minimise risk of O.O.S. material • Eyecon integrated directly after Mill outlet • Measured impact of parameter changes and process fluctuations in real time

  28. Slide 28: Fluid Bed Granulation

    Fluid Bed Granulation • Automation of a fluid bed wet granulation process using Innopharma’s SmartX advanced control platform • Eyecon2 provided real-time particle measurement used for phase end-point determination – greater control of end-product quality • Subsequent study in progress on linking in

  29. Slide 29: Application Deep Dive: Wurster Coating

    Application Deep Dive: Wurster Coating • Dissolution Prediction Example published in Pharmaceutical Technology April 2017 issue, Pharma Focus Asia Issue 33, presented at IFPAC 2017 ©2020, Innopharma Technology Ltd 29

  30. Slide 30: Modified Release Products

    Modified Release Products • Formulations where the in-vitro release time or location of the drug are engineered to meet therapeutic objectives • Release location • Patient convenience • Controlled release / sustained release / delayed release… • In oral solid dosage forms typically accomplished with

  31. Slide 31: The Wurster Coating Process

    The Wurster Coating Process ©2020, Innopharma Technology Ltd 31

  32. Slide 32: Control of Coating Processes

    Control of Coating Processes • Current methods use little to no inline CQA monitoring • Typically controlled by spraying a fixed quantity of coating factor • Coating is an additive process - as coating is applied a particle size increase is expected • Directly related to weight gain • Size increase

  33. Slide 33: Presenting Particles for Imaging

    Presenting Particles for Imaging: 1 • Particles imaged directly behind surface of window • Applicable with quasi-static bodies of material e.g. fluid bed, and with flowing materials • Window helps to ensure particles are optimally presented within the depth of field • Challenges • Wide PSD – smaller

  34. Slide 34: Dissolution Prediction Study: Equipment & Formulation

    Dissolution Prediction Study: Equipment & Formulation • Glatt GPCG2 with 7” expansion chamber extension • 6” PAT-compatible Wurster product container Fitted with Eyecon2 Material Amount/batch CPM layered Sugar Spheres (12 mg) - 18/20 mesh 2000 g Surelease – aqueous ethylcellulose dispersion 1408 g O

  35. Slide 35: DOE & Sampling Strategy

    DOE & Sampling Strategy • Duplicate experiments conducted • CPM-SR-1 – develop basic model • CPM-SR-2 – validate basic model & improve • Coated to 20% weight gain • Samples taken at 2.5% w.g. intervals • Additional samples after 30 & 60 minutes curing time • Offline analysis • Camsizer • Dissolution

  36. Slide 36: In-Line PSD during Wurster Coating: Film Thickness

    In-Line PSD during Wurster Coating: Film Thickness • Observable, consistent growth between sample points Film thickness (µm) as a factor of predicted weight gain percentage - Batch 1 40 35 d Film Thickness (µm) 30 25 20 15 10 d+2f 5 0 0.0% 5.0% 10.0% 15.0% 20.0% 25.0% Sample Point / % Weight Gain Pr

  37. Slide 37: At-Line – In-Line PSD Validation

    At-Line – In-Line PSD Validation • R2 = 0.9895 • Strong correlation between Eyecon & Camsizer Eyecon vs. Camsizer - Combined Particle Size Ranges 1100 Particle Size (diameter, µm) - Eyecon 1050 1000 950 y = 0.9404x + 95.628 900 R² = 0.9895 850 800 800 850 900 950 1000 1050 Particle Size (diameter, u

  38. Slide 38: In-Line PSD during Wurster Coating: Dissolution Data

    In-Line PSD during Wurster Coating: Dissolution Data Dissolution Data Grouped by Time Point – Batch 1 120 100 80 % Dissolved 60 40 20 0 0 60 120 180 240 300 360 420 480 540 600 660 720 Time in Dissolution Medium (minutes) CPM-SR-5% CPM-SR-10% CPM-SR-15% CPM-SR-20% CPM-SR-20%-30 min CPM-SR-20%-1 hr ©

  39. Slide 39: Relationship Between Dissolution & Film Thickness

    In-Line PSD during Wurster Coating: Relationship Between Dissolution & Film Thickness • Polynomial fit between PSD & dissolution for varying film thickness • Shows possibility of model-based real-time measurement / prediction of dissolution! Film Thickness vs. Dissolution – Batch 1 120 100 y = -0.02

  40. Slide 40: Predicted Dissolution vs Actual

    In-Line PSD during Wurster Coating: Predicted Dissolution vs Actual Batch 2: Predicted vs Actual Results 120% 5% WG (P) 100% 10% WG (P) 15% WG (P) 80% 20% WG (P) Dissolution % 30 min cured (P) 60% 60 min cured (P) 5% WG (A) 10% WG (A) 40% 15% WG (A) 20% WG (A) 20% 30 min cured (A) 60 min cured (A) 0

  41. Slide 41: Review

    Review • Innopharma Introduction • Eyecon2 – Direct Imaging System for In-line Particle Size Measurement • Practical Considerations for Implementation • Sensor Interface with Process Equipment • Application • TSG, Milling, FBG • Deep Dive – Wuster Coating the Real-Time Prediction of Polymer-Coated M

  42. Slide 42: Thank you!

    Thank you! Thank you for listening! Questions? Darren McHugh Product Manager Innopharma Technology Ltd mchughd@innopharmalabs.com Chris O’Callaghan Head of Engineering Innopharma Technology Ltd ocallaghanc@innopharmalabs.com ©2020, Innopharma Technology Ltd 42

Innopharma Technology — Technical Presentation (March 2020) on 6ix